Comparison of a novel, single-lead atrial sensing system with a dual-chamber implantable cardioverter-defibrillator system in patients without antibradycardia pacing indications: results of a randomized study.

Christian Sticherling1, Markus Zabel, Sebastian Spencker, Udo Meyerfeldt, Lars Eckardt, Steffen Behrens, Michael Niehaus.   

Abstract

BACKGROUND: Supraventricular tachyarrhythmias are the main cause for inappropriate therapy by implantable cardioverter-defibrillators (ICDs). For better rhythm discrimination, an atrial electrogram is helpful and usually obtained from an additional atrial lead, even in the absence of sinus node or atrioventricular nodal disease. An A+-ICD system with integrated atrial sensing rings mounted 15 to 18 cm from the tip of an ICD lead may obviate the need to implant a separate atrial lead. The aim of the study was to compare the novel A+-ICD and a conventional dual-chamber (DR)-ICD. METHODS AND
RESULTS: Two hundred forty-nine patients with standard ICD indications but no requirement for antibradycardia pacing were randomized to receive an A+-ICD (n=124) or a DR-ICD (n=125). Implantation details, need for ICD system revision, long-term sensing, documented arrhythmia episodes, and the respective rhythm discrimination during follow-up were analyzed. The implantation time was significantly shorter in the A+-ICD group (67±30 vs 79±30 minutes, P=0.003). Mean P-wave amplitudes were 3.5±0.8 mV (A+-ICD) and 3.2±0.6 mV (DR-ICD) and remained stable during the follow-up period of 12 months. Surgical revision was necessary in 13 patients in the DR-ICD and 10 in the A+-ICD group. All 593 ventricular tachyarrhythmia episodes were correctly discriminated. Sensitivity and specificity of supraventricular tachyarrhythmia discrimination were not different between the study groups.
CONCLUSIONS: The novel A+-ICD system can be implanted faster and is equivalent to a standard DR-ICD with regard to the detection of ventricular tachyarrhythmias and supraventricular tachyarrhythmias. It represents a useful alternative to obtain atrial sensing.

Entities:  

Mesh:

Year:  2010        PMID: 21156772     DOI: 10.1161/CIRCEP.110.958397

Source DB:  PubMed          Journal:  Circ Arrhythm Electrophysiol        ISSN: 1941-3084


  12 in total

Review 1.  ICD programming.

Authors:  Mauro Biffi
Journal:  Indian Heart J       Date:  2013-12-20

2.  The Dx-AF study: a prospective, multicenter, randomized controlled trial comparing VDD-ICD to VVI-ICD in detecting sub-clinical atrial fibrillation in defibrillator patients.

Authors:  Mohammed Shurrab; Amir Janmohamed; Jean-François Sarrazin; Felix Ayala-Paredes; Marcio Sturmer; Randall Williams; Satish Toal; Chris Lane; Kevin E Thorpe; Jeff S Healey; Eugene Crystal
Journal:  J Interv Card Electrophysiol       Date:  2017-07-27       Impact factor: 1.900

3.  A new single chamber implantable defibrillator with atrial sensing: a practical demonstration of sensing and ease of implantation.

Authors:  Dietmar Bänsch; Ralph Schneider; Ibrahim Akin; Cristoph A Nienaber
Journal:  J Vis Exp       Date:  2012-02-28       Impact factor: 1.355

Review 4.  "Two for the Price of One": A Single-Lead Implantable Cardioverter-Defibrillator System with a Floating Atrial Dipole.

Authors:  Nicole E Worden; Musab Alqasrawi; Siva M Krothapalli; Alexander Mazur
Journal:  J Atr Fibrillation       Date:  2016-04-30

5.  Cardiac resynchronization therapy in persistent left superior vena cava: Can you do it two-leads-only?

Authors:  Mauro Biffi; Giulia Massaro; Igor Diemberger; Cristian Martignani; Alessandro Corzani; Matteo Ziacchi
Journal:  HeartRhythm Case Rep       Date:  2016-10-19

6.  Atrial electrogram quality in single-pass defibrillator leads with floating atrial bipole in patients with permanent atrial fibrillation and cardiac resynchronization therapy.

Authors:  Christian Sticherling; Dirk Müller; Beat A Schaer; Silke Krüger; Christof Kolb
Journal:  Indian Pacing Electrophysiol J       Date:  2018-03-27

7.  The VDD ICD lead: Friend or Foe?

Authors:  Suneet Mittal
Journal:  Indian Pacing Electrophysiol J       Date:  2020-04-20

8.  Outcomes of single- or dual-chamber implantable cardioverter defibrillator systems in Japanese patients.

Authors:  Akiko Ueda; Yasushi Oginosawa; Kyoko Soejima; Haruhiko Abe; Ritsuko Kohno; Hisaharu Ohe; Yuichi Momose; Mika Nagaoka; Noriko Matsushita; Kyoko Hoshida; Yosuke Miwa; Mutsumi Miyakoshi; Ikuko Togashi; Akiko Maeda; Toshiaki Sato; Hideaki Yoshino
Journal:  J Arrhythm       Date:  2015-12-10

9.  Novel two-lead cardiac resynchronization therapy system provides equivalent CRT responses with less complications than a conventional three-lead system: Results from the QP ExCELs lead registry.

Authors:  Naushad A Shaik; Michael Drucker; Christopher Pierce; Gabor Z Duray; Shane Gillett; Crystal Miller; Camden Harrell; George Thomas
Journal:  J Cardiovasc Electrophysiol       Date:  2020-06-01

10.  Clinical safety of ProMRI implantable cardioverter-defibrillator systems during head and lower lumbar magnetic resonance imaging at 1.5 Tesla.

Authors:  Wolfgang Rudolf Bauer; Dennis H Lau; Christian Wollmann; Andrew McGavigan; Jacques Mansourati; Theresa Reiter; Simone Frömer; Mark E Ladd; Harald H Quick
Journal:  Sci Rep       Date:  2019-12-03       Impact factor: 4.379

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.