Navin Mukundu Nagesh1, Nimesh Patel1, Jonathan Howell1.
Abstract
Total hip replacement is a successful operation for the management of hip pain but there are potential complications, of which dislocation is one of the most common. The management of recurrent dislocation is a challenging problem that requires a multimodal approach and the use of dual-mobility implants is one option. We present a patient who was previously revised with a dual-mobility implant for recurrent dislocation, who had a complication after closed reduction of a subsequent intraprosthetic dislocation. Following a missed radiographical diagnosis, the patient experienced mechanical symptoms on hip flexion caused by a disassociated dual-mobility implant. Subsequent surgical removal of the failed implant and revision was required. Careful study of radiographs revealed an eccentric femoral head and evidence of the disassociated implant within the surrounding soft tissues. Radiographs following closed reduction of intraprosthetic dislocations should be scrutinised closely to detect implant failure to prevent further complications. © BMJ Publishing Group Ltd (unless otherwise stated in the text of the article) 2018. All rights reserved. No commercial use is permitted unless otherwise expressly granted.Entities:
Keywords: dual mobility implant; hip implants; hip prosthesis implantation; intraprosthetic dislocation; prosthesis failure; total hip replacement
Mesh:
Year: 2018 PMID: 29572368 PMCID: PMC5878339 DOI: 10.1136/bcr-2017-223553
Source DB: PubMed Journal: BMJ Case Rep ISSN: 1757-790X