| Literature DB >> 29564407 |
Mohamad Reza Sedaghat1, Hadi Ostadi-Moghadam2,3, Mahmoud Jabbarvand4, Farshad Askarizadeh2,3, Hamed Momeni-Moghaddam2,3, Foroozan Narooie-Noori2,3.
Abstract
PURPOSE: To evaluate and compare corneal hysteresis (CH) and corneal resistance factor (CRF) in pellucid marginal degeneration (PMD), keratoconus (KCN), and normal eyes using the Ocular Response Analyzer (ORA).Entities:
Keywords: Corneal biomechanics; Corneal hysteresis; Corneal resistance factor; Keratoconus; Pellucid marginal degeneration
Year: 2017 PMID: 29564407 PMCID: PMC5859559 DOI: 10.1016/j.joco.2017.08.002
Source DB: PubMed Journal: J Curr Ophthalmol ISSN: 2452-2325
Baseline data of all subjects by groups and outcomes of ocular response analyzer (ORA).
| Variables | Groups | |||
|---|---|---|---|---|
| Normal (Mean ± SD) ( | KCN (Mean ± SD) ( | PMD (Mean ± SD) ( | ||
| Age (years) | 30.29 ± 5.33 | 28.62 ± 3.88 | 31.48 ± 6.75 | 0.78 |
| SE (D) | −2.50 ± 0.43 | −2.81 ± 2.18 | −2.24 ± 2.51 | |
| KMAX (D) | 45.07 ± 1.53 | 51.59 ± 5.19 | 51.32 ± 4.44 | |
| CCT (μm) | 541.45 ± 29.25 | 477.68 ± 37.33 | 492.09 ± 38.72 | |
| TPP (μm) | 536.48 ± 29.64 | 469.18 ± 42.16 | 471.89 ± 59.76 | |
| CH (mmHg) | 10.89 ± 1.08 | 8.43 ± 0.78 | 8.91 ± 1.05 | |
| CRF (mmHg) | 10.69 ± 1.41 | 7.19 ± 1.11 | 8.21 ± 1.35 | |
| CH–CRF (mmHg) | 0.22 ± 0.87 | 1.24 ± 0.86 | 0.69 ± 1.06 | |
| IOPCC (mmHg) | 15.03 ± 1.62 | 15.57 ± 1.42 | 15.86 ± 1.55 | |
| IOPG (mmHg) | 14.84 ± 2.02 | 12.86 ± 1.07 | 13.64 ± 1.53 | |
| IOPCC–IOPG (mmHg) | 0.19 ± 1.53 | 2.80 ± 1.55 | 2.20 ± 1.70 | |
KCN: Keratoconus, PMD: Pellucid marginal degeneration, SD: Standard deviation, SE: Spherical equivalent, KMAX: Maximum keratometry, CCT: Central corneal thickness, TPP: Thinnest point pachymetry, CH: Corneal hysteresis, CRF: Corneal resistance factor, IOPCC: Corneal compensated intraocular pressure, IOPG: Goldmann correlated intraocular pressure.
One-way ANOVA test. Bold values are significant. No missing data are available. A P-value < 0.05 was considered statistically significant.
Fig. 1The receiver operating characteristic (ROC) curve analysis for determination of the cut-off point of corneal biomechanical parameters include: corneal hysteresis (CH), corneal resistance factor (CRF), and difference between CH and CRF (CH–CRF) in pellucid marginal degeneration using the Ocular Response Analyzer (ORA).
Fig. 2The receiver operating characteristic (ROC) curve analysis for determination of the cut-off point of corneal-compensated intraocular pressure (IOPCC), Goldmann-correlated intraocular pressure (IOPG), and difference between IOPCC and IOPG (IOPCC–IOPG) in pellucid marginal degeneration using the Ocular Response Analyzer (ORA).
Results of the receiver operating characteristic (ROC) curve analysis for outcomes of ocular response analyzer (ORA) in eyes with pellucid marginal degeneration.
| Variables | AUC value (at 95% CI) | Cut points (sensitivity–specificity) | |
|---|---|---|---|
| CH | 0.903 | <0.0001 | ≤9.5 (71.57%–91.83%) |
| CRF | 0.896 | <0.0001 | ≤9.0 (75.49%–84.62%) |
| CH–CRF | 0.627 | <0.0003 | >1.3 (32.35%–93.27%) |
| IOPCC | 0.596 | 0.0030 | >14.3 (81.40%–40.40%) |
| IOPG | 0.649 | <0.0001 | ≤15.0 (83.30%–47.10%) |
| IOPCC–IOPG | 0.816 | <0.0001 | >1.8 (63.73%–88.46%) |
AUC: Area under the ROC curve, CI: Confidence interval, CH: Corneal hysteresis, CRF: Corneal resistance factor, CH–CRF: Difference between CH and CRF, IOPCC: Corneal-compensated intraocular pressure, IOPG: Goldmann correlated intraocular pressure, IOPCC–IOPG: Difference between and IOPCC and IOPG.
Unit of all variables: mmHg.