Abstract

Image guidance with cone beam computed tomography in radiotherapy can guarantee the precision and accuracy of patient positioning prior to treatment delivery. During the image guidance process, operators need to take great effort to evaluate the image guidance quality before correcting a patient's position. This work proposes an image registration assessment method based on control chart monitoring to reduce the effort taken by the operator. According to the control chart plotted by daily registration scores of each patient, the proposed method can quickly detect both alignment errors and image quality inconsistency. Therefore, the proposed method can provide a clear guideline for the operators to identify unacceptable image quality and unacceptable image registration with minimal effort. Experimental results demonstrate that by using control charts from a clinical database of 10 patients undergoing prostate radiotherapy, the proposed method can quickly identify out-of-control signals and find special cause of out-of-control registration events.

Entities:  

Keywords:  control chart monitoring; image registration assessment; radiotherapy image guidance

Year:  2018        PMID: 29564368      PMCID: PMC5849944          DOI: 10.1117/1.JMI.5.2.021221

Source DB:  PubMed          Journal:  J Med Imaging (Bellingham)        ISSN: 2329-4302


  16 in total

1.  Statistical process control for radiotherapy quality assurance.

Authors:  Todd Pawlicki; Matthew Whitaker; Arthur L Boyer
Journal:  Med Phys       Date:  2005-09       Impact factor: 4.071

Review 2.  Image-guided radiotherapy: from current concept to future perspectives.

Authors:  David A Jaffray
Journal:  Nat Rev Clin Oncol       Date:  2012-11-20       Impact factor: 66.675

3.  Impact of planning target volume margins and rectal distention on biochemical failure in image-guided radiotherapy of prostate cancer.

Authors:  Benedikt Engels; Guy Soete; Thierry Gevaert; Guy Storme; Dirk Michielsen; Mark De Ridder
Journal:  Radiother Oncol       Date:  2014-03-13       Impact factor: 6.280

4.  Quality control of high-dose-rate brachytherapy: treatment delivery analysis using statistical process control.

Authors:  Charles M Able; Megan Bright; Bart Frizzell
Journal:  Int J Radiat Oncol Biol Phys       Date:  2012-06-30       Impact factor: 7.038

5.  Subject-specific four-dimensional liver motion modeling based on registration of dynamic MRI.

Authors:  Yolanda H Noorda; Lambertus W Bartels; Max A Viergever; Josien P W Pluim
Journal:  J Med Imaging (Bellingham)       Date:  2016-02-19

6.  Evaluation of model-based deformation correction in image-guided liver surgery via tracked intraoperative ultrasound.

Authors:  Logan W Clements; Jarrod A Collins; Jared A Weis; Amber L Simpson; Lauryn B Adams; William R Jarnagin; Michael I Miga
Journal:  J Med Imaging (Bellingham)       Date:  2016-03-23

7.  Fiducial registration error as a statistical process control metric in image-guidance radiotherapy with fiducial markers.

Authors:  N M Ung; L Wee
Journal:  Phys Med Biol       Date:  2011-11-11       Impact factor: 3.609

8.  A study on the theoretical and practical accuracy of conoscopic holography-based surface measurements: toward image registration in minimally invasive surgery.

Authors:  J Burgner; A L Simpson; J M Fitzpatrick; R A Lathrop; S D Herrell; M I Miga; R J Webster
Journal:  Int J Med Robot       Date:  2012-07-04       Impact factor: 2.547

9.  Patient-reported complications from fiducial marker implantation for prostate image-guided radiotherapy.

Authors:  S Gill; J Li; J Thomas; M Bressel; K Thursky; C Styles; K H Tai; G M Duchesne; F Foroudi
Journal:  Br J Radiol       Date:  2012-01-17       Impact factor: 3.039

10.  Fabrication and control of CT number through polymeric composites based on coronary plaque CT phantom applications.

Authors:  Carlton F O Hoy; Hani E Naguib; Narinder Paul
Journal:  J Med Imaging (Bellingham)       Date:  2016-02-18
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.