| Literature DB >> 29549270 |
S Van Vrekhem1, K Vloebergh2, M Asadian2, C Vercruysse3, H Declercq3, A Van Tongel4, L De Wilde4, N De Geyter2, R Morent2.
Abstract
Insufficient glenoid fixation is one of the main reasons forEntities:
Mesh:
Substances:
Year: 2018 PMID: 29549270 PMCID: PMC5856771 DOI: 10.1038/s41598-018-22921-6
Source DB: PubMed Journal: Sci Rep ISSN: 2045-2322 Impact factor: 4.379
Figure 1Schematic representation of the atmospheric pressure plasma jet configuration.
Figure 2Water contact angle values as a function of treatment time for different jet velocities.
Process parameters of selected conditions with their respective WCA and surface roughness values.
| Treatment time (s) | Jet velocity (mm/min) | Gas flow (slm) | Discharge power (W) | Energy density (J/cm²) | WCA (°) | Rq (nm) | |
|---|---|---|---|---|---|---|---|
| Saturation 1 (S1) | 4.68 | 450 | 2 | 1.60 | 564.14 | 47.1 ± 0.7 | 145 ± 22 |
| Saturation 2 (S2) | 4.68 | 150 | 2 | 1.60 | 564.14 | 46.0 ± 1.1 | 142 ± 34 |
| Saturation 3 (S3) | 4.68 | 50 | 2 | 1.60 | 564.14 | 47.3 ± 1.0 | 140 ± 27 |
| Condition B | 0.35 | 450 | 2 | 1.60 | 42.19 | 72.6 ± 1.0 | 124 ± 19 |
| Condition F | 4.68 | 50 | 0.75 | 2.60 | 916.73 | 25.2 ± 2.2 | 133 ± 27 |
O/C-ratio and relative percentage of carbon-containing functional groups present on the sample’s surface for each condition.
| O/C | C-C/C-H (%) | C-O (%) | C = O (%) | O-C = O (%) | |
|---|---|---|---|---|---|
| S1 | 0.43 ± 0.03 | 63.2 ± 2.4 | 14.2 ± 3.2 | 10.0 ± 2.1 | 12.6 ± 1.2 |
| S2 | 0.41 ± 0.03 | 62.3 ± 4.9 | 17.2 ± 5.8 | 8.0 ± 1.5 | 12.5 ± 1.0 |
| S3 | 0.40 ± 0.02 | 65.9 ± 6.4 | 18.5 ± 7.8 | 6.4 ± 1.6 | 9.1 ± 1.2 |
| B | 0.26 ± 0.03 | 72.6 ± 3.1 | 14.0 ± 3.5 | 8.4 ± 0.9 | 5.0 ± 0.6 |
| F | 0.42 ± 0.02 | 62.0 ± 3.7 | 20.5 ± 5.4 | 8.0 ± 2.2 | 9.5 ± 0.8 |
Figure 3WCA values as a function of storage time after plasma treatment.
Percentage of remaining surface modification R, O/C-ratio and relative percentage of carbon-containing functional groups present on the surface after 7 days of ageing in ambient air.
| R (%) | O/C | C-C/C-H (%) | C-O (%) | C = O (%) | O-C = O (%) | |
|---|---|---|---|---|---|---|
| S1 | 77 | 0.27 ± 0.01 | 73.1 ± 0.2 | 11.9 ± 0.2 | 8.2 ± 0.3 | 6.9 ± 0.4 |
| S2 | 76 | 0.23 ± 0.03 | 72.6 ± 0.5 | 12.7 ± 0.1 | 8.1 ± 0.1 | 6.7 ± 0.5 |
| S3 | 73 | 0.23 ± 0.02 | 74.1 ± 0.8 | 12.0 ± 0.5 | 7.5 ± 0.4 | 6.4 ± 0.5 |
| B | 96 | 0.12 ± 0.02 | 79.4 ± 0.8 | 12.4 ± 0.4 | 5.9 ± 0.5 | 2.3 ± 0.1 |
| F | 80 | 0.23 ± 0.04 | 73.3 ± 0.4 | 11.7 ± 0.1 | 7.6 ± 0.4 | 7.5 ± 0.3 |
Figure 4Pull-out stress for different process conditions.
Figure 5Percentage of viable attached MC3T3 cells for different conditions compared to a positive control culture for a culture time of 1 day and 7 days.
Figure 6Fluorescence images of MC3T3 cells for different conditions after a culture time of 7 days. Upper row: Untreated (left), condition S1 (middle), condition S3 (right). Lower row: condition F (left), condition B (right).
Figure 7FTIR-spectra of the CaP-layer on plasma-treated (S1) and untreated samples after immersion in 2.0 SBF for 14 days.
Figure 8Micrograph of the CaP deposition on untreated (left) and plasma-treated (S1, right) samples.
Figure 9SEM images of plasma-treated samples (condition S1) for x40 (upper left), x250 (upper right), x4000 (lower left) and x15000 (lower right) magnifications.
Figure 10SEM images of untreated samples for x40 (upper left), x250 (upper right), x4000 (lower left) and x15000 (lower right) magnifications.