Ya-Chi Li1, Hui-Ling Lin2,3,4, Fang-Chun Liao1, Sing-Siang Wang1, Hsiu-Chu Chang1, Hung-Fu Hsu5, Sue-Hsien Chen6, Gwo-Hwa Wan2,4,7.
Abstract
BACKGROUND: Few studies have investigated the difference in bacterial contamination between conventional reused ventilator systems and disposable closed ventilator-suction systems. The aim of this study was to investigate the bacterial contamination rates of the reused and disposable ventilator systems, and the association between system disconnection and bacterial contamination of ventilator systems.Entities:
Mesh:
Year: 2018 PMID: 29547638 PMCID: PMC5856346 DOI: 10.1371/journal.pone.0194246
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Fig 1Diagram depicting the enrollment and follow-up of the study participants.
Fig 2Diagram of a ventilator system including the inlet tube of the HH (A), the HH (B), inspiratory limb (C), Y-adapter (D), 15-cm corrugated tube (E), and the expiratory limb (F).
Personal characteristics of study subjects.
| ID | Sex | Age | Ventilation day | Major diagnosis |
|---|---|---|---|---|
| 1 | F | 78 | 17 | Severe sepsis, hypercapnic respiratory failure, pneumonia |
| 2 | F | 81 | 29 | Cardiac arrest, hypoxemic encephalopathy |
| 3 | F | 83 | 29 | Acute respiratory failure, fever |
| 4 | M | 79 | 20 | HCAP, respiratory failure |
| 5 | M | 80 | 26 | HCAP with hypercapnic respiratory failure, COPD with acute exacerbation |
| 6 | M | 81 | 29 | COPD with acute exacerbation, VAP |
| 7 | F | 83 | 24 | CAP with respiratory failure |
| 8 | M | 63 | 20 | Septic shock |
| 9 | F | 71 | 23 | Mixed type respiratory failure, VAP |
| 10 | M | 78 | 21 | Mixed type respiratory failure, cardiac arrest |
| 11 | M | 70 | 21 | Nosocomial pneumonia |
| 12 | M | 69 | 21 | Influenza A pneumonia with acute respiratory distress syndrome |
| 13 | M | 91 | 33 | HAP |
| 14 | M | 49 | 36 | HCAP |
| 15 | M | 20 | 46 | Bilateral lung contusion |
| 16 | F | 42 | 22 | HAP with acute respiratory distress syndrome |
HCAP, healthcare-associated pneumonia; COPD, chronic obstructive pulmonary disease; VAP, ventilator-associated pneumonia; CAP, community-acquired pneumonia; HAP, hospital-acquired pneumonia.
Detection rates and concentrations of bacteria detected at different locations of reused and disposable ventilator systems.
| Location | Detection rate (n, %) | Concentration (CFU/mL) | ||||
|---|---|---|---|---|---|---|
| Reused system | Disposable system | Reused system | Disposable system | |||
| Total | 91 (94.8%) | 59 (81.9%) | < 0.01 | 1.98 × 106 (5.98 × 105–1.43 × 107) | 2.41 × 106 (1.9 × 105–2.41 × 107) | 0.616 |
| IHH | 12 (75%) | 2 (16.7%) | < 0.01 | 20.00 (2.5–75.00) | 0 (0–10) | 0.021 |
| HH | 15 (93.8%) | 9 (75%) | 0.168 | 1.96 × 106 (8.25 × 105–5.13 × 106)*¶ | 2.08 × 105 (2.50 × 103–7.05 × 105) | 0.004 |
| IL | 16 (100%) | 12 (100%) | — | 2.69 × 106 (1.53 × 106–1.48 × 107) | 2.56 × 107 (1.29 × 107–1.42 × 108) | 0.002 |
| Y-adapter | 16 (100%) | 12 (100%) | — | 1.25 × 106 (6.13 × 105–1.72 × 106) | 1.51 × 106 (1.12 × 106–4.08 × 106) | 0.117 |
| 15-cm tube | 16 (100%) | 12 (100%) | — | 7.43 × 106 (1.76 × 106–4.79 × 107) | 1.82 × 107 (5.19 × 106–2.43 × 107) | 0.099 |
| EL | 16 (100%) | 12 (100%) | — | 1.47 × 107 (8.25 × 106–1.93 × 107) | 1.28 × 108 (1.35 × 107–2.55 × 108) | 0.004 |
| 0.02 | < 0.01 | < 0.01 | < 0.01 | |||
IHH: IHH, inlet tube of the HH; HH, heated humidifier; IL, inspiratory limb; EL, expiratory limb.
#: Data was presented as median (min–max).
*: compared to IHH;
†: compared to HH;
‡: compared to IL;
§: compared to Y-adapter;
@: compared to 15-cm tube;
¶: compared to EL.
Associations of disconnection frequency and bacterial concentrations at different locations of reused and disposable ventilator systems.
| (1) | (2) | (3) | (4) | (5) | (6) | (7) | |
|---|---|---|---|---|---|---|---|
| (1) Disconnection frequency | 1 | ||||||
| (2) IHH | 0.182 | 1 | |||||
| (3) HH | 0.185 | 0.35 | 1 | ||||
| (4) IL | 0.243 | 0.251 | 0.603 | 1 | |||
| (5) Y-adapter | 0.027 | -0.159 | 0.193 | 0.353 | 1 | ||
| (6) 15-cm tube | 0.094 | -0.028 | 0.374 | 0.456 | 0.677 | 1 | |
| (7) EL | -0.223 | -0.01 | 0.096 | 0.318 | 0.549 | 0.571 | 1 |
| (1) Disconnection frequency | 1 | ||||||
| (2) IHH | -0.3 | 1 | |||||
| (3) HH | 0.106 | -0.406 | 1 | ||||
| (4) IL | -0.347 | -0.275 | 0.346 | 1 | |||
| (5) Y-adapter | -0.098 | -0.134 | 0.592 | 0.445 | 1 | ||
| (6) 15-cm tube | 0.306 | -0.005 | 0.373 | 0.336 | 0.797 | 1 | |
| (7) EL | -0.137 | -0.027 | 0.233 | 0.572 | 0.767 | 0.697 | 1 |
IHH, inlet tube of the HH; HH, heated humidifier; IL, inspiratory limb; EL, expiratory limb.
*: P < 0.05
**: P < 0.01.
Bacterial species distributions of sputum specimens, reused and disposable ventilator systems.
| Species | Sputum specimen | Reused ventilator system | Disposable ventilator system | |||
|---|---|---|---|---|---|---|
| (n = 16) | (n = 16) | (n = 16) | ||||
| | 0 | (0%) | 2 | (3.77%) | 4 | (10%) |
| | 0 | (0%) | 3 | (5.66%) | 1 | (2.5%) |
| | 1 | (5.55%) | 1 | (1.89%) | 1 | (2.5%) |
| | 1 | (5.55%) | 1 | (1.89%) | 1 | (2.5%) |
| | 0 | (0%) | 0 | (0%) | 1 | (2.5%) |
| | 3 | (16.67%) | 14 | (26.42%) | 6 | (15%) |
| | 0 | (0%) | 1 | (1.89%) | 0 | (0%) |
| | 0 | (0%) | 3 | (5.66%) | 1 | (2.5%) |
| | 0 | (0%) | 1 | (1.89%) | 1 | (2.5%) |
| | 1 | (5.55%) | 0 | (0%) | 0 | (0%) |
| | 0 | (0%) | 1 | (1.89%) | 1 | (2.5%) |
| | 1 | (5.55%) | 5 | (9.43%) | 3 | (7.5%) |
| | 1 | (5.55%) | 0 | (0%) | 0 | (0%) |
| | 0 | (0%) | 0 | (0%) | 1 | (2.5%) |
| | 1 | (5.55%) | 0 | (0%) | 0 | (0%) |
| | 3 | (16.67%) | 0 | (0%) | 1 | (2.5%) |
| | 1 | (5.55%) | 0 | (0%) | 0 | (0%) |
| | 4 | (22.22%) | 6 | (11.32%) | 4 | (10%) |
| | 0 | (0%) | 2 | (3.77%) | 1 | (2.5%) |
| | 0 | (0%) | 1 | (1.89%) | 1 | (2.5%) |
| | 0 | (0%) | 2 | (3.77%) | 1 | (2.5%) |
| | 1 | (5.55%) | 10 | (18.88%) | 11 | (27.5%) |
Data were presented as n (%).