| Literature DB >> 29536235 |
Shinya Higuchi1, Iwao Matsunaga2, Kiyokazu Koga2, Eiichi Teshima2, Atsuhiro Nakashima2, Ryuji Tominaga2, Tsuyoshi Ito2.
Abstract
In almost every type of artificial valve, structural failure has been described. We are reporting on a case of a sudden leaflet escape of an Edwards TEKNA mitral valve prosthesis 12 years after implantation. The patient had a sudden onset of dyspnea and severe pulmonary edema with subsequent cardiogenic shock. An emergency mitral replacement was successfully performed. A multi-detector computed tomography scanning and three-dimensional imaging showed two fragments that had embolized in the terminal aorta and the left common iliac artery. The patient presented visual field abnormality, and postoperative head computed tomography showed watershed cerebral infarction. The escaped leaflet that fractured transversely was removed, following the patient's recovery, during cardiac surgery.Entities:
Keywords: Leaflet escape; Mitral valve surgery; Multi-detector computed tomography; Prosthesis failure
Mesh:
Year: 2018 PMID: 29536235 DOI: 10.1007/s11748-018-0891-4
Source DB: PubMed Journal: Gen Thorac Cardiovasc Surg ISSN: 1863-6705