Literature DB >> 16480019

Fatal leaflet escape in an Edwards TEKNA aortic prosthesis.

Heidi Pfeiffer1, Julia Bertolini, Hans Heinrich Scheld, Bernd Brinkmann.   

Abstract

The case is reported of a 26-year-old male patient who died eight years after the replacement of an aortic valve with a bileaflet mechanical valve (TEKNA; Edwards, USA). Following prosthesis implantation, the patient had been in a good state of health, and his death occurred unexpectedly. Forensic autopsy revealed a leaflet escape, with two fragments of the leaflet being found bilaterally in the common iliac arteries. Death occurred due to an acute cardiac insufficiency. Immunohistochemical investigations revealed fresh myocardial fiber necroses. Stereomicroscopic and scanning electron microscopic investigations demonstrated surface erosions of the leaflet. Although the valve was withdrawn from the market in June 2000, it had previously been implanted in over 18,000 patients. Thus, from a clinical viewpoint, the question of using a prophylactic replacement in affected patients must be discussed.

Entities:  

Mesh:

Year:  2006        PMID: 16480019

Source DB:  PubMed          Journal:  J Heart Valve Dis        ISSN: 0966-8519


  2 in total

1.  Leaflet escape in an Edwards TEKNA mitral prosthesis.

Authors:  Shinya Higuchi; Iwao Matsunaga; Kiyokazu Koga; Eiichi Teshima; Atsuhiro Nakashima; Ryuji Tominaga; Tsuyoshi Ito
Journal:  Gen Thorac Cardiovasc Surg       Date:  2018-03-13

2.  Spontaneous leaflet fracture resulting in embolization from mechanical valve prostheses.

Authors:  Gijs G J van Steenbergen; Queeny H Y Tsang; Stefan M van der Heide; Michel W A Verkroost; Wilson W L Li; Wim J Morshuis
Journal:  J Card Surg       Date:  2018-12-30       Impact factor: 1.620

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.