Literature DB >> 11297209

Fracture embolization of a Tekna mitral prosthesis: case report.

S Jazayeri1, J P Meunier, E Tatou, I L'Huillier, O Bouchot, C Toulouse, R Brenot, M David.   

Abstract

The modified Duromedics-Tekna bileaflet pyrolitic carbon mechanical prosthesis was reintroduced by Baxter in 1990. This report details the first case of sudden leaflet fracture of a Tekna mitral valve five years after implantation, which was managed successfully by replacement with a St. Jude Medical mechanical prosthesis. The fracture had occurred transversely, with the fragments embolizing to the terminal aortic bifurcation and the left common femoral artery. These were localized by computed tomography and removed two days after valve replacement.

Entities:  

Mesh:

Year:  2001        PMID: 11297209

Source DB:  PubMed          Journal:  J Heart Valve Dis        ISSN: 0966-8519


  2 in total

1.  Leaflet escape in an Edwards TEKNA mitral prosthesis.

Authors:  Shinya Higuchi; Iwao Matsunaga; Kiyokazu Koga; Eiichi Teshima; Atsuhiro Nakashima; Ryuji Tominaga; Tsuyoshi Ito
Journal:  Gen Thorac Cardiovasc Surg       Date:  2018-03-13

2.  Spontaneous leaflet fracture resulting in embolization from mechanical valve prostheses.

Authors:  Gijs G J van Steenbergen; Queeny H Y Tsang; Stefan M van der Heide; Michel W A Verkroost; Wilson W L Li; Wim J Morshuis
Journal:  J Card Surg       Date:  2018-12-30       Impact factor: 1.620

  2 in total