Literature DB >> 17461884

Runaway implantable defibrillator--a rare complication of radiation therapy.

Jan Nemec1.   

Abstract

A case of a patient with runaway implantable cardioverter defibrillator (ICD) due to radiation therapy of a lung cancer is reported. This manifested as poorly tolerated wide complex tachycardia due to inappropriate rapid ventricular pacing, The event terminated with polymorphic VT, which inhibited pacing and ceased spontaneously before ICD discharge. The likely cause was corruption of device random access memory by ionizing radiation.

Entities:  

Mesh:

Year:  2007        PMID: 17461884     DOI: 10.1111/j.1540-8159.2007.00735.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  8 in total

1.  Experimental study of neutron-induced soft errors in modern cardiac pacemakers.

Authors:  Alexandre Trigano; Guillaume Hubert; Jannie Marfaing; Karine Castellani
Journal:  J Interv Card Electrophysiol       Date:  2011-08-24       Impact factor: 1.900

Review 2.  [Radiation therapy in patients with cardiac pacemakers or implantable cardioverter defibrillators. Interdisciplinary safety recommendations].

Authors:  M Dorenkamp; C Stromberger; C von Heymann; W Haverkamp; P Wust; M Roser
Journal:  Strahlenther Onkol       Date:  2012-11-11       Impact factor: 3.621

3.  DEGRO/DGK guideline for radiotherapy in patients with cardiac implantable electronic devices.

Authors:  Benjamin Gauter-Fleckenstein; Carsten W Israel; Marc Dorenkamp; Jürgen Dunst; Mattias Roser; Rainer Schimpf; Volker Steil; Jörg Schäfer; Ulrike Höller; Frederik Wenz
Journal:  Strahlenther Onkol       Date:  2015-03-05       Impact factor: 3.621

Review 4.  Management of radiation therapy patients with cardiac defibrillator or pacemaker.

Authors:  Francesca Salerno; Sara Gomellini; Cristina Caruso; Raffaele Barbara; Daniela Musio; Tamara Coppi; Mario Cardinale; Vincenzo Tombolini; Ugo de Paula
Journal:  Radiol Med       Date:  2015-12-26       Impact factor: 3.469

5.  Implantable cardiac pacemaker failure by cumulative dose effects of flattening filter free beams.

Authors:  Kazuhiko Nakamura; Takahiro Aoyama; Naoki Kaneda; Masashi Otsuji; Yoshitaka Minami; Ami Sakuragi; Masaru Nakamura
Journal:  J Radiat Res       Date:  2021-07-10       Impact factor: 2.724

6.  Pacemaker malfunction associated with proton beam therapy: a report of two cases and review of literature-does field-to-generator distance matter?

Authors:  Tomoko Ueyama; Takeshi Arimura; Takashi Ogino; Naoaki Kondo; Ryutaro Higashi; Fumihiko Nakamura; Soichiro Ito; Takashi Yoshiura; Yoshio Hishikawa
Journal:  Oxf Med Case Reports       Date:  2016-08-29

7.  Pacemaker and radiotherapy in breast cancer: is targeted intraoperative radiotherapy the answer in this setting?

Authors:  Mohammed Rs Keshtgar; David J Eaton; Claire Reynolds; Katharine Pigott; Tim Davidson; Benjamin Gauter-Fleckenstein; Frederik Wenz
Journal:  Radiat Oncol       Date:  2012-08-01       Impact factor: 3.481

8.  Management of radiation oncology patients with a pacemaker or ICD: a new comprehensive practical guideline in The Netherlands. Dutch Society of Radiotherapy and Oncology (NVRO).

Authors:  Coen W Hurkmans; Joost L Knegjens; Bing S Oei; Ad J J Maas; G J Uiterwaal; Arnoud J van der Borden; Marleen M J Ploegmakers; Lieselot van Erven
Journal:  Radiat Oncol       Date:  2012-11-24       Impact factor: 3.481

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.