Literature DB >> 29489589

Quantification of Iodine Concentration Using Single-Source Dual-Energy Computed Tomography in a Calf Liver.

Andrea Agostini, Usman Mahmood, Yusuf Erdi, Alessandra Borgheresi, Monica Ragucci, Peter Sawan, Davinia Ryan, Maria Elena Laino, Giuseppe Corrias, Lorenzo Mannelli.   

Abstract

OBJECTIVE: To evaluate the accuracy of single-source dual-energy computed tomography (ssDECT) in iodine quantification using various segmentation methods in an ex vivo model.
METHODS: Ten sausages, injected with variable quantities of iodinated contrast, were inserted into 2 livers and scanned with ssDECT. Material density iodine images were reconstructed. Three radiologists segmented each sausage. Iodine concentration, volume, and absolute quantity were measured. Agreement between the measured and injected iodine was assessed with the concordance correlation coefficient (CCC). Intrareader agreement was assessed using the intraclass correlation coefficient (ICC).
RESULTS: Air bubbles were observed in sausage (IX). Sausage (X) was within the same view as hyper-attenuating markers used for localization. With IX and X excluded, CCC and ICC were greater than 0.98 and greater than 0.88. When included, CCC and ICC were greater than 0.94 and greater than 0.79.
CONCLUSIONS: Iodine quantification was reproducible and precise. However, accuracy reduced in sausages consisting of air filled cavities and within the same view as hyperattenuating markers.

Entities:  

Mesh:

Substances:

Year:  2018        PMID: 29489589      PMCID: PMC5847415          DOI: 10.1097/RCT.0000000000000685

Source DB:  PubMed          Journal:  J Comput Assist Tomogr        ISSN: 0363-8715            Impact factor:   1.826


  22 in total

1.  Iodine quantification with dual-energy CT: phantom study and preliminary experience with VX2 residual tumour in rabbits after radiofrequency ablation.

Authors:  Y Li; G Shi; S Wang; S Wang; R Wu
Journal:  Br J Radiol       Date:  2013-07-24       Impact factor: 3.039

2.  Quantitative imaging of element composition and mass fraction using dual-energy CT: three-material decomposition.

Authors:  Xin Liu; Lifeng Yu; Andrew N Primak; Cynthia H McCollough
Journal:  Med Phys       Date:  2009-05       Impact factor: 4.071

3.  Objective characterization of GE discovery CT750 HD scanner: gemstone spectral imaging mode.

Authors:  Da Zhang; Xinhua Li; Bob Liu
Journal:  Med Phys       Date:  2011-03       Impact factor: 4.071

4.  Quantitative therapy response assessment by volumetric iodine-uptake measurement: initial experience in patients with advanced hepatocellular carcinoma treated with sorafenib.

Authors:  Xu Dai; Heinz-Peter Schlemmer; Bernhard Schmidt; Karolin Höh; Ke Xu; Tom M Ganten; Maria-Katharina Ganten
Journal:  Eur J Radiol       Date:  2012-12-12       Impact factor: 3.528

5.  Iodine quantification with dual-energy CT: phantom study and preliminary experience with renal masses.

Authors:  Hersh Chandarana; Alec J Megibow; Benjamin A Cohen; Ramya Srinivasan; Danny Kim; Christianne Leidecker; Michael Macari
Journal:  AJR Am J Roentgenol       Date:  2011-06       Impact factor: 3.959

6.  Iodine quantification using dual-energy multidetector computed tomography imaging: phantom study assessing the impact of iterative reconstruction schemes and patient habitus on accuracy.

Authors:  Sebastian Feuerlein; Tobias J Heye; Mustafa R Bashir; Daniel T Boll
Journal:  Invest Radiol       Date:  2012-11       Impact factor: 6.016

7.  Novel oncologic drugs: what they do and how they affect images.

Authors:  Roberto García Figueiras; Anwar R Padhani; Vicky J Goh; Joan C Vilanova; Sandra Baleato González; Carmen Villalba Martín; Antonio Gómez Caamaño; Anaberta Bermúdez Naveira; Peter L Choyke
Journal:  Radiographics       Date:  2011 Nov-Dec       Impact factor: 5.333

Review 8.  Dual- and Multi-Energy CT: Principles, Technical Approaches, and Clinical Applications.

Authors:  Cynthia H McCollough; Shuai Leng; Lifeng Yu; Joel G Fletcher
Journal:  Radiology       Date:  2015-09       Impact factor: 11.105

9.  Appearance and Frequency of Gas Interface Artifacts Involving Small Bowel on Rapid-Voltage-Switching Dual-Energy CT Iodine-Density Images.

Authors:  En-Haw Wu; So Yeon Kim; Z Jane Wang; Wei-Chou Chang; Li-Qin Zhao; Benjamin M Yeh
Journal:  AJR Am J Roentgenol       Date:  2016-02       Impact factor: 3.959

10.  Monitoring targeted therapy using dual-energy CT: semi-automatic RECIST plus supplementary functional information by quantifying iodine uptake of melanoma metastases.

Authors:  M Uhrig; M Sedlmair; H P Schlemmer; J C Hassel; M Ganten
Journal:  Cancer Imaging       Date:  2013-07-22       Impact factor: 3.909

View more