| Literature DB >> 29487771 |
Ali Fatemi1, Somayeh Taghizadeh2, Claus Chunli Yang2, Madhava R Kanakamedala2, Bart Morris2, Srinivasan Vijayakumar2.
Abstract
Purpose Magnetic resonance (MR) images are necessary for accurate contouring of intracranial targets, determination of gross target volume and evaluation of organs at risk during stereotactic radiosurgery (SRS) treatment planning procedures. Many centers use magnetic resonance imaging (MRI) simulators or regular diagnostic MRI machines for SRS treatment planning; while both types of machine require two stages of quality control (QC), both machine- and patient-specific, before use for SRS, no accepted guidelines for such QC currently exist. This article describes appropriate machine-specific QC procedures for SRS applications. Methods and materials We describe the adaptation of American College of Radiology (ACR)-recommended QC tests using an ACR MRI phantom for SRS treatment planning. In addition, commercial Quasar MRID3D and Quasar GRID3D phantoms were used to evaluate the effects of static magnetic field (B0) inhomogeneity, gradient nonlinearity, and a Leksell G frame (SRS frame) and its accessories on geometrical distortion in MR images. Results QC procedures found in-plane distortions (Maximum = 3.5 mm, Mean = 0.91 mm, Standard deviation = 0.67 mm, >2.5 mm (%) = 2) in X-direction (Maximum = 2.51 mm, Mean = 0.52 mm, Standard deviation = 0.39 mm, > 2.5 mm (%) = 0) and in Y-direction (Maximum = 13. 1 mm , Mean = 2.38 mm, Standard deviation = 2.45 mm, > 2.5 mm (%) = 34) in Z-direction and < 1 mm distortion at a head-sized region of interest. MR images acquired using a Leksell G frame and localization devices showed a mean absolute deviation of 2.3 mm from isocenter. The results of modified ACR tests were all within recommended limits, and baseline measurements have been defined for regular weekly QC tests. Conclusions With appropriate QC procedures in place, it is possible to routinely obtain clinically useful MR images suitable for SRS treatment planning purposes. MRI examination for SRS planning can benefit from the improved localization and planning possible with the superior image quality and soft tissue contrast achieved under optimal conditions.Entities:
Keywords: mri; mri guided radiotherapy; mri qa; qc; srs treatment planning
Year: 2017 PMID: 29487771 PMCID: PMC5815649 DOI: 10.7759/cureus.1957
Source DB: PubMed Journal: Cureus ISSN: 2168-8184
Approved MRI pulse sequences for SRS treatment planning.
CISS: Three-dimensional (3D) constructive interference in steady state; MPRAGE: Magnetization-prepared 180-degree radio-frequency pulses and rapid gradient-echo; MRI: Magnetic resonance imaging; SRS: Stereotactic radiosurgery; TE: Time of Echo; TR: Time of Repetition.
| Sequence/contrast | Parameters | Disease |
| Axial T1-weighted MPRAGE | 1 x 1 x 1 mm3, TR/TE = 2200/2.91 ms, 300 Hz/pixel | Brain metastasis, pituitary/parasellar lesions, acoustic neuroma/schwannoma, trigeminal neuralgia, arteriovenous malformation |
| Axial T2-weighted Space | 0.9 x 0.9 x 1 mm3, TR/TE = 1400/184 ms, 345 Hz/pixel | Pituitary/parasellar lesions, acoustic neuroma/schwannoma, arteriovenous malformation |
| Axial T2 CISS | 0.9 x 0.9 x 1 mm3, TR/TE = 5.48/2.38 ms, 340 Hz/pixel | Pituitary/parasellar lesions, acoustic neuroma/schwannoma, trigeminal neuralgia |
Quality control tests and frequencies for magnetic resonance imaging-guided stereotactic radiosurgery.
ACR: American College of Radiology; Bn: Magnetic field; MRI: Magnetic resonance imaging; PMU: Phasor measurement unit; QA: Quality assurance; RF: Radiofrequency; Rx: Receive; SRS: Stereotactic radiosurgery; Tx: Transmit.
| Daily QA (MRI technologists) using ACR phantom | Monthly QA (Therapy physicist/MRI physicist) using MRID3D, GRID3D and ACR phantoms | Annual QA (MRI physicist) using MRID3D, GRID3D and ACR phantoms |
| Inspect bore for loose metal (bobby pins, earrings, etc.) | Patient safety (monitors, intercom, panic ball, emergency buttons, and signage) | 20-channels RF coil integrity check |
| Tx/Rx and 20-channel RF coil SRS check using uniform phantom | Patient comfort (bore light and fan) | B0 constancy |
| Patient safety (intercom, panic ball, detector) | Percent signal ghosting | B1+ constancy |
| Geometry accuracy and B0 check using ACR phantom | Percent image uniformity | Gradient linearity constancy |
| High/low contrast accuracy | Slice thickness accuracy | |
| Weekly QA (MRI technologists) Using ACR phantom | Coach position accuracy | Slice position accuracy |
| Transmitter gain constancy | Image artifact | Geometrical accuracy |
| Center frequency constancy | Geometrical accuracy (large field of view) | Rx/Tx RF head coil check |
| 20-channel RF head coil SNR | Geometrical accuracy (small field of view) with and without frame | 20-channel RF head coil check |
| Rx/Tx RF head coil SNR | Dynamic field map | |
| Slice thickness accuracy | Eddy current compensation | |
| Slice position accuracy | Gradient delay | |
| Geometric accuracy and B0 check using ACR phantom | Gradient sensitivity | |
| Body coil image brightness | ||
| Magnet shim | ||
| Rx gain calibration | ||
| Body coil tuning | ||
| Spike | ||
| PMU transmit | ||
| Rx stability | ||
| Tx stability |
Summary of magnetic resonance imaging quality control tests for stereotactic radiosurgery treatment planning.
DSV: Diameter of spherical volume; MRI: Magnetic resonance imaging; PIU: Percent integral uniformity; PMU: Phasor measurement unit; PSG: Percentage signal ghosting; RF: Radio frequency; Rx/Tx: Receive/transmit; SNR: Signal-to-noise ratio.
| Test | MRI machine tolerance |
| MRI geometrical distortion | |
| Evaluate distortion vector, combined effect (B0 inhomogeneity and gradient nonlinearity) over big field of view (37 cm) | <1 mm over 20 cm DSV and <2 mm over 37 cm DSV |
| Evaluate B0 inhomogeneity over large field of view (37 cm) | 2 ppm |
| Evaluate the geometrical distortion vector with stereotactic frame (small field of view, 20 cm) | <1 mm |
| Adapted ACR QC tests | |
| Setup and table position accuracy | <1 mm |
| Center frequency | Pass/Fail |
| Signal ghosting | ≤2.5% |
| Transmitter gain or attenuation | Pass/Fail |
| High contrast spatial resolution | Row and column resolution ≤ 1 mm |
| Low contrast detectability | Nine rows total for up to 1.5 T |
| Magnetic field homogeneity | Action limit ± 2 ppm |
| Artifact evaluation | Pass/Fail |
| Magnetic field homogeneity | Action limit ± 2 ppm |
| Geometrical accuracy | Within < 1.5 mm of actual length |
| Visual checklist | Pass/Fail |
| Slice position accuracy | Difference from actual position ≤ 3 mm |
| Slice thickness accuracy | Action limit is 5 ± 0.7 mm |
| 20-channel RF head coil evaluation | Signal-to-noise ratio; PIU ≥ 87.5% (<3 T), PSG ≤ 2.5% |
| Rx/Tx RF head coil evaluation | Signal-to-noise ratio; PIU ≥ 87.5% (<3T), PSG ≤ 2.5% |
| Rx/Tx RF head coil check | Pass/Fail |
| 20-channel RF head coil check | Pass/Fail |
| Dynamic field map | Pass/Fail |
| Eddy current compensation | Pass/Fail |
| Gradient delay | Pass/Fail |
| Gradient sensitivity | Pass/Fail |
| Body coil image brightness | Pass/Fail |
| Magnet shim | Pass/Fail |
| Rx gain calibration | Pass/Fail |
| Body coil tuning | Pass/Fail |
| Spike | Pass/Fail |
| PMU transmit | Pass/Fail |
| Rx stability | Pass/Fail |
| Tx stability | Pass/Fail |
| Assessment of MRI safety program | Pass/Fail |
Figure 1Quasar MRID3D geometrical distortion phantom (Modus Medical).
Figure 2Phantoms used in the study.
Left: Quasar GRID3D image distortion phantom (Modus Medical, Canada); Right: Standard American College of Radiology magnetic resonance imaging phantom.
MRI pulse sequences and parameters for adapted magnetic resonance imaging ACR quality control tests.
ACR: American College of Radiology; BW: Bandwidth; FOV: Field of view; MRI: Magnetic resonance imaging; NEX: Number of excitation; TE: Time of Echo; TR: Time of Repetition.
*Acquired three separate series, each consisting of a single image through the center of phantom with minimum and highest bandwidths.
| Protocol | Phantom used | TR (ms) | TE (ms) | FOV (mm) | # Slices | Slice thickness (mm) | GAP | NEX | Matrix | BW | |
| ACR T1 Localizer | ACR | 200 | 20 | 25 | 1 | 20 | N/A | 1 | 256 x 256 | Routine (15.6 kHz) | |
| ACR T1 Axial | ACR | 500 | 20 | 25 | 11 | 5 | 5 | 1 | 256 x 256 | Routine (15.6 kHz) | |
| ACR T2 Axial | ACR | 2000 | 20/80 | 25 | 11 | 5 | 5 | 1 | 256 x 256 | Routine (15.6 kHz) | |
| Site T1 Axial | ACR | Site protocol | Site protocol | 25 | 11 | 5 | 5 | Site protocol | 256 x 256 | Site protocol BW | |
| Site T2 Axial | ACR | Site protocol | Site protocol | 25 | 11 | 5 | 5 | Site protocol | 256 x 256 | Site protocol BW | |
| Low BW Axial * | ACR | 500 | 20 | 25 | 1 | 5 | N/A | 1 | 256 x 256 | Minimum BW @ 256 x 256 matrix | |
| Low BW Coronal * | ACR | 500 | 20 | 25 | 1 | 5 | N/A | 1 | 256 x 256 | Minimum BW @ 256 x 256 matrix | |
| Low BW Sagittal* | ACR | 500 | 20 | 25 | 1 | 5 | N/A | 1 | 256 x 256 | Minimum BW @ 256 x 256 matrix | |
| High BW Axial* | ACR | 500 | 20 | 25 | 1 | 5 | N/A | 1 | 256 x 256 | Highest BW @ 256 x 256 matrix | |
| High BW Coronal* | ACR | 500 | 20 | 25 | 1 | 5 | N/A | 1 | 256 x 256 | Highest BW @ 256 x 256 matrix | |
| High BW Sagittal* | ACR | 500 | 20 | 25 | 1 | 5 | N/A | 1 | 256 x 256 | Highest BW @ 256 x 256 matrix | |
Summary statistics for MRID3D geometrical distortion measurements.
STD: Standard deviation.
| Mean (mm) | STD (mm) | Max (mm) | >2.5 mm (%) | |
| dx | 0.91 | 0.67 | 3.5 | 2 |
| dy | 0.52 | 0.39 | 2.51 | 0 |
| dz | 2.38 | 2.45 | 13.1 | 34 |
| dr | 2.79 | 2.36 | 13.19 | 40 |
Figure 3Static magnetic field (B0) inhomogeneity and distortion measurement results using an MRID3D phantom.
Magnetic resonance imaging stereotactic radiosurgery planning quality control results.
BW: Bandwidth; MRI: Magnetic resonance imaging; PMU: Phasor measurement unit; RF: Radiofrequency; Rx/Tx: Receiver/transmitter; PIU: Percent image uniformity; SNR: Signal-to-noise ratio.
| MRI equipment evaluation summary | |||
| Setup and table position accuracy | Pass | ||
| Center frequency | Pass | ||
| Transmitter gain or attenuation | Pass | ||
| Geometric accuracy measurements* | Pass | ||
| High-contrast spatial resolution* | Pass | ||
| Low-contrast detectability* | Pass | ||
| Artifact evaluation | Pass | ||
| Visual checklist | Pass | ||
| Magnetic field homogeneity | Pass | ||
| Method of Testing | BW diff | ||
| Slice-position accuracy* | Pass | ||
| Slice-thickness accuracy* | Pass | ||
| Radiofrequency coil checks (20-channel RF head coil) | |||
| a. SNR | Pass | ||
| b. Volume coil percent image uniformity | Pass | ||
| c. Percent signal ghosting | Pass | ||
| Radiofrequency coil checks (Rx/Tx RF head coil) | |||
| a. SNR | Pass | ||
| b. Volume coil PIU | Pass | ||
| c. Percent signal ghosting | Pass | ||
| Rx/Tx RF head coil check | Pass | ||
| 20-channel RF head coil check | Pass | ||
| Dynamic field map | Pass | ||
| Eddy current compensation | Pass | ||
| Gradient delay | Pass | ||
| Gradient sensitivity | Pass | ||
| Body coil image brightness | Pass | ||
| Magnet shim | Pass | ||
| 23 | RX gain calibration | Pass | |
| 24 | Body coil tuning | Pass | |
| 25 | Spike | Pass | |
| 26 | PMU transmit | Pass | |
| 27 | Rx stability | Pass | |
| 28 | Tx stability | Pass | |