Literature DB >> 29487771

Machine-Specific Magnetic Resonance Imaging Quality Control Procedures for Stereotactic Radiosurgery Treatment Planning.

Ali Fatemi1, Somayeh Taghizadeh2, Claus Chunli Yang2, Madhava R Kanakamedala2, Bart Morris2, Srinivasan Vijayakumar2.   

Abstract

Purpose Magnetic resonance (MR) images are necessary for accurate contouring of intracranial targets, determination of gross target volume and evaluation of organs at risk during stereotactic radiosurgery (SRS) treatment planning procedures. Many centers use magnetic resonance imaging (MRI) simulators or regular diagnostic MRI machines for SRS treatment planning; while both types of machine require two stages of quality control (QC), both machine- and patient-specific, before use for SRS, no accepted guidelines for such QC currently exist. This article describes appropriate machine-specific QC procedures for SRS applications. Methods and materials We describe the adaptation of American College of Radiology (ACR)-recommended QC tests using an ACR MRI phantom for SRS treatment planning. In addition, commercial Quasar MRID3D and Quasar GRID3D phantoms were used to evaluate the effects of static magnetic field (B0) inhomogeneity, gradient nonlinearity, and a Leksell G frame (SRS frame) and its accessories on geometrical distortion in MR images. Results QC procedures found in-plane distortions (Maximum = 3.5 mm, Mean = 0.91 mm, Standard deviation = 0.67 mm, >2.5 mm (%) = 2) in X-direction (Maximum = 2.51 mm, Mean = 0.52 mm, Standard deviation = 0.39 mm, > 2.5 mm (%) = 0) and in Y-direction (Maximum = 13. 1 mm , Mean = 2.38 mm, Standard deviation = 2.45 mm, > 2.5 mm (%) = 34) in Z-direction and < 1 mm distortion at a head-sized region of interest. MR images acquired using a Leksell G frame and localization devices showed a mean absolute deviation of 2.3 mm from isocenter. The results of modified ACR tests were all within recommended limits, and baseline measurements have been defined for regular weekly QC tests. Conclusions With appropriate QC procedures in place, it is possible to routinely obtain clinically useful MR images suitable for SRS treatment planning purposes. MRI examination for SRS planning can benefit from the improved localization and planning possible with the superior image quality and soft tissue contrast achieved under optimal conditions.

Entities:  

Keywords:  mri; mri guided radiotherapy; mri qa; qc; srs treatment planning

Year:  2017        PMID: 29487771      PMCID: PMC5815649          DOI: 10.7759/cureus.1957

Source DB:  PubMed          Journal:  Cureus        ISSN: 2168-8184


Introduction

Using magnetic resonance (MR) images for stereotactic radiosurgery (SRS) treatment planning requires careful consideration of a number of factors [1], including choice of the correct magnetic resonance imaging (MRI) pulse sequences (3D, no slice gap, less geometrical distortion, high signal-to-noise ratio, and isotropic spatial resolution), immobilization devices (MRI-compatible SRS frame), customized radiofrequency (RF) coils (proper sensitivity, low RF deposition, and consequently less contour deformation) such as a single channel send-and-receive RF head coil (Rx/Tx RF head coil), and most importantly, confirmation that the MRI images acquired possess high geometrical accuracy and stability. Existing MR quality control (QC) procedures [2-6] are inadequate for assessing MRI scanners for SRS treatment-planning purposes, primarily because existing tests have been developed for machines used in general diagnostic radiology. There, the goal is to maintain image quality rather than spatial fidelity and signal intensity [7]. Several well-established references from the American College of Radiology (ACR) [2,3] and the American Association of Physicists in Medicine (AAPM) [4-6] provide guidance regarding QC procedures for MRI scanners used in diagnostic radiology, but no guidance documents currently describe the unique QC factors that must be considered when using MRI scanners in SRS treatment planning [7-9]. However, existing quality control tests can be modified to provide the necessary information for a given SRS-planning application by testing over appropriate volumes and using SRS-specific MR imaging parameters. Development of our QC process started with an evaluation of gross machine factors, including main static magnetic field (B0) inhomogeneity and gradient non-linearity over a large field of view using the scanner body coil, then narrowed to study the effects on the geometrical stability of MR images due to use of an MRI-compatible SRS frame and its localizer using a Tx/Rx RF head coil [10]. Finally, adapted ACR tests were performed to evaluate the image contrast, spatial resolution, gradient stability for accurate slice selection and thickness, RF coil sensitivity, and acquisition of artifact-free MR images. Testing these factors ensures that acquired images possess the quality and resolution required for precision SRS treatment planning, accurately identifying disease extent and proximity relative to adjacent organs at risk (OAR) [1]. We are establishing a quality assurance (QA) program to continuously and systematically evaluate MRI scanner performance, safety and stability for SRS treatment planning. Our goal in this article is to describe our QC tests and strategy in establishing a QA program for MRI-guided SRS treatment planning. This paper focuses narrowly on MRI machine-specific aspects of the QC procedure and leaves patient-specific QC tests, including patient-specific geometrical distortion evaluation, correction methods, customized RF coils, patient comfort, MRI safety, and MRI pulse sequence optimization, for future reports.

Materials and methods

We recently installed a Leksell Gamma Knife® Icon™ SRS treatment unit (Elekta AB, Stockholm, Sweden) and a MAGNETOM Aera 1.5 T (Siemens Healthcare, Germany) Radiotherapy Edition MRI machine at our institute. The SRS committee consists of three physicists, a radiologist, a radiation oncologist, and a neurosurgeon, who work together to develop guidelines for MRI-guided SRS treatment planning. MR images are used to assess cases of brain metastasis, pituitary/parasellar lesions, acoustic neuroma, trigeminal neuralgia and arteriovenous malformation (AVM). The MRI SRS QC procedure has been developed based on factors including imaging site, MRI pulse sequence(s), adapted or standard RF coils, and any immobilization devices required. SRS patients are scanned on a regular diagnostic MRI table, using a Leksell G frame with immobilization and localization devices, or frameless, as appropriate. In our institute, we use a Tx/Rx CP head coil (Siemens Healthcare, Germany) to fit the Leksell G frame, plus an MRI indicator box with an adaptor to the coil, and are still able to keep the specific absorption rate (SAR) under 3 W/kg. The downside of using such an RF coil is a less-than-ideal signal-to-noise ratio (SNR) and long scanning time; thus, we use a regular 20-channel RF head coil for frameless cases. MRI pulse sequences have been evaluated by the SRS committee based on disease site and treatment planning criteria detailed in Table 1.
Table 1

Approved MRI pulse sequences for SRS treatment planning.

CISS: Three-dimensional (3D) constructive interference in steady state; MPRAGE: Magnetization-prepared 180-degree radio-frequency pulses and rapid gradient-echo; MRI: Magnetic resonance imaging; SRS: Stereotactic radiosurgery; TE: Time of Echo; TR: Time of Repetition.

Sequence/contrast Parameters Disease
Axial T1-weighted MPRAGE 1 x 1 x 1 mm3, TR/TE = 2200/2.91 ms, 300 Hz/pixel Brain metastasis, pituitary/parasellar lesions, acoustic neuroma/schwannoma, trigeminal neuralgia, arteriovenous malformation
Axial T2-weighted Space 0.9 x 0.9 x 1 mm3, TR/TE = 1400/184 ms, 345 Hz/pixel Pituitary/parasellar lesions, acoustic neuroma/schwannoma, arteriovenous malformation
Axial T2 CISS 0.9 x 0.9 x 1 mm3, TR/TE = 5.48/2.38 ms, 340 Hz/pixel Pituitary/parasellar lesions, acoustic neuroma/schwannoma, trigeminal neuralgia

Approved MRI pulse sequences for SRS treatment planning.

CISS: Three-dimensional (3D) constructive interference in steady state; MPRAGE: Magnetization-prepared 180-degree radio-frequency pulses and rapid gradient-echo; MRI: Magnetic resonance imaging; SRS: Stereotactic radiosurgery; TE: Time of Echo; TR: Time of Repetition. We summarize the commissioning and QC tests (test, frequency, and machine tolerance) in Table 2 and Table 3.
Table 2

Quality control tests and frequencies for magnetic resonance imaging-guided stereotactic radiosurgery.

ACR: American College of Radiology; Bn: Magnetic field; MRI: Magnetic resonance imaging; PMU: Phasor measurement unit; QA: Quality assurance; RF: Radiofrequency; Rx: Receive; SRS: Stereotactic radiosurgery; Tx: Transmit.

Daily QA (MRI technologists) using ACR phantom Monthly QA (Therapy physicist/MRI physicist) using MRID3D, GRID3D and ACR phantoms Annual QA (MRI physicist) using MRID3D, GRID3D and ACR phantoms
Inspect bore for loose metal (bobby pins, earrings, etc.) Patient safety (monitors, intercom, panic ball, emergency buttons, and signage) 20-channels RF coil integrity check
Tx/Rx and 20-channel RF coil SRS check using uniform phantom Patient comfort (bore light and fan) B0 constancy
Patient safety (intercom, panic ball, detector) Percent signal ghosting B1+ constancy
Geometry accuracy and B0 check using ACR phantom Percent image uniformity Gradient linearity constancy
  High/low contrast accuracy Slice thickness accuracy
Weekly QA (MRI technologists) Using ACR phantom Coach position accuracy Slice position accuracy
Transmitter gain constancy Image artifact Geometrical accuracy
Center frequency constancy Geometrical accuracy (large field of view) Rx/Tx RF head coil check
20-channel RF head coil SNR Geometrical accuracy (small field of view) with and without frame 20-channel RF head coil check
Rx/Tx RF head coil SNR   Dynamic field map
Slice thickness accuracy   Eddy current compensation
Slice position accuracy   Gradient delay
Geometric accuracy and B0 check using ACR phantom   Gradient sensitivity
   Body coil image brightness
   Magnet shim
   Rx gain calibration
    Body coil tuning
    Spike
    PMU transmit
    Rx stability
    Tx stability
Table 3

Summary of magnetic resonance imaging quality control tests for stereotactic radiosurgery treatment planning.

DSV: Diameter of spherical volume; MRI: Magnetic resonance imaging; PIU: Percent integral uniformity; PMU: Phasor measurement unit; PSG: Percentage signal ghosting; RF: Radio frequency; Rx/Tx: Receive/transmit; SNR: Signal-to-noise ratio.

Test MRI machine tolerance
MRI geometrical distortion  
Evaluate distortion vector, combined effect (B0 inhomogeneity and gradient nonlinearity) over big field of view (37 cm) <1 mm over 20 cm DSV and <2 mm over 37 cm DSV
Evaluate B0 inhomogeneity over large field of view (37 cm) 2 ppm
Evaluate the geometrical distortion vector with stereotactic frame (small field of view, 20 cm) <1 mm
Adapted ACR QC tests  
Setup and table position accuracy <1 mm
Center frequency Pass/Fail
Signal ghosting ≤2.5%
Transmitter gain or attenuation Pass/Fail
High contrast spatial resolution Row and column resolution ≤ 1 mm
Low contrast detectability Nine rows total for up to 1.5 T
Magnetic field homogeneity Action limit ± 2 ppm
Artifact evaluation Pass/Fail
Magnetic field homogeneity Action limit ± 2 ppm
Geometrical accuracy Within < 1.5 mm of actual length
Visual checklist Pass/Fail
Slice position accuracy Difference from actual position ≤ 3 mm
Slice thickness accuracy Action limit is 5 ± 0.7 mm
20-channel RF head coil evaluation Signal-to-noise ratio; PIU ≥ 87.5% (<3 T), PSG ≤ 2.5%
Rx/Tx RF head coil evaluation Signal-to-noise ratio; PIU ≥ 87.5% (<3T),  PSG ≤ 2.5%
Rx/Tx RF head coil check Pass/Fail
20-channel RF head coil check Pass/Fail
Dynamic field map Pass/Fail
Eddy current compensation Pass/Fail
Gradient delay Pass/Fail
Gradient sensitivity Pass/Fail
Body coil image brightness Pass/Fail
Magnet shim Pass/Fail
Rx gain calibration Pass/Fail
Body coil tuning Pass/Fail
Spike Pass/Fail
PMU transmit Pass/Fail
Rx stability Pass/Fail
Tx stability Pass/Fail
Assessment of MRI safety program Pass/Fail

Quality control tests and frequencies for magnetic resonance imaging-guided stereotactic radiosurgery.

ACR: American College of Radiology; Bn: Magnetic field; MRI: Magnetic resonance imaging; PMU: Phasor measurement unit; QA: Quality assurance; RF: Radiofrequency; Rx: Receive; SRS: Stereotactic radiosurgery; Tx: Transmit.

Summary of magnetic resonance imaging quality control tests for stereotactic radiosurgery treatment planning.

DSV: Diameter of spherical volume; MRI: Magnetic resonance imaging; PIU: Percent integral uniformity; PMU: Phasor measurement unit; PSG: Percentage signal ghosting; RF: Radio frequency; Rx/Tx: Receive/transmit; SNR: Signal-to-noise ratio. Evaluation of geometrical distortion over a large field of view We used a QUASAR™ MRID3D geometrical distortion phantom (Figure 1) (Modus Medical, Canada) to measure B0 inhomogeneity and gradient non-linearity using a reverse gradient technique over a 37 cm x 32 cm (W x L) phantom area. The phantom was scanned with a 3D VIBE T1-weighted sequence: 1 mm3 isotropic voxels, NEX of two, TE of 4 ms, TR of 9 ms, a flip angle ~10°, and a bandwidth of 120 Hz/pixel. QUASAR™ MRID3D comes with easy-to-use image analysis software for calculation of the phantom boundary distortion vector field, volumetric 3D distortion vector field, and B0 distortion vs gradient distortion, using 3D spherical harmonic analysis.
Figure 1

Quasar MRID3D geometrical distortion phantom (Modus Medical).

Evaluation of the effect of an SRS frame and localizer on geometrical stability of MR images We used a QUASAR™ GRID3D Image Distortion Phantom and analysis system (Modus Medical, Canada) to evaluate MR image distortion due to the introduction of an SRS frame and localizers. The system is comprised of a phantom and analysis software which work together to produce a 3D map of spatial distortion with submillimeter accuracy throughout a volume of interest. The phantom (Figure 2, left) is an acrylic cube containing a 1-cm 3D grid of channels filled with copper sulfate solution. The region of interest is a 14 x 13 x 11 cm3 volume containing 2002 vertex locations, the positions of which are known to within 0.1 mm. The phantom accurately and reproducibly mounts securely to the SRS Leksell Frame G at a known position. It fits within both the Leksell® MR Indicator and Leksell® CT Indicator. We scanned our phantom using a 3D MPRAGE pulse sequence: T1-weighted, 1 x 1 x 1 mm3, TR/TE of 2200/3.74 ms, and 350 Hz/pixel. The MPRAGE is the only MRI pulse sequence currently being used for treatment planning, and the rest of the sequences will be registered rigidly.
Figure 2

Phantoms used in the study.

Left: Quasar GRID3D image distortion phantom (Modus Medical, Canada); Right: Standard American College of Radiology magnetic resonance imaging phantom.

Phantoms used in the study.

Left: Quasar GRID3D image distortion phantom (Modus Medical, Canada); Right: Standard American College of Radiology magnetic resonance imaging phantom. ACR MRI tests adapted for SRS treatment planning We used a standard MRI ACR phantom (Figure 2, right) to evaluate the rest of our adapted QC MRI tests. The ACR phantom has been scanned based on MRI pulse sequences and parameters summarized in Table 4. All QC tests with their tolerances are summarized in Table 3.
Table 4

MRI pulse sequences and parameters for adapted magnetic resonance imaging ACR quality control tests.

ACR: American College of Radiology; BW: Bandwidth; FOV: Field of view; MRI: Magnetic resonance imaging; NEX: Number of excitation; TE: Time of Echo; TR: Time of Repetition.

*Acquired three separate series, each consisting of a single image through the center of phantom with minimum and highest bandwidths.

Protocol Phantom used TR (ms) TE (ms) FOV (mm) # Slices Slice thickness (mm) GAP NEX Matrix BW
ACR T1 Localizer ACR 200 20 25 1 20 N/A 1 256 x 256 Routine (15.6 kHz)
ACR T1 Axial ACR 500 20 25 11 5 5 1 256 x 256 Routine (15.6 kHz)
ACR T2 Axial ACR 2000 20/80 25 11 5 5 1 256 x 256 Routine (15.6 kHz)
Site T1 Axial ACR Site protocol Site protocol 25 11 5 5 Site protocol 256 x 256 Site protocol BW
Site T2 Axial ACR Site protocol Site protocol 25 11 5 5 Site protocol 256 x 256 Site protocol BW
Low BW Axial * ACR 500 20 25 1 5 N/A 1 256 x 256 Minimum BW @ 256 x 256 matrix
Low BW Coronal * ACR 500 20 25 1 5 N/A 1 256 x 256 Minimum BW @ 256 x 256 matrix
Low BW Sagittal* ACR 500 20 25 1 5 N/A 1 256 x 256 Minimum BW @ 256 x 256 matrix
High BW Axial* ACR 500 20 25 1 5 N/A 1 256 x 256 Highest BW @ 256 x 256 matrix
High BW Coronal* ACR 500 20 25 1 5 N/A 1 256 x 256 Highest BW @ 256 x 256 matrix
High BW Sagittal* ACR 500 20 25 1 5 N/A 1 256 x 256 Highest BW @ 256 x 256 matrix

MRI pulse sequences and parameters for adapted magnetic resonance imaging ACR quality control tests.

ACR: American College of Radiology; BW: Bandwidth; FOV: Field of view; MRI: Magnetic resonance imaging; NEX: Number of excitation; TE: Time of Echo; TR: Time of Repetition. *Acquired three separate series, each consisting of a single image through the center of phantom with minimum and highest bandwidths. Evaluating MRI safety As part of our acceptance tests, we used a gauss meter to carefully map and post the five-gauss line with proper signage. We monitor all patients through both a questionnaire and in-person consultation to ensure that any person with a cardiac pacemaker or neurosimulators does not cross the five-G line. Our MRI room is also equipped with a Ferroguard® wall-mounted system deployed in an entryway mode on both sides of the doorway. This system provides real-time monitoring of the local ferromagnetic environment with an audible alert system. We also check the patient/console intercom system, table-top button (magnet housing and console), emergency stop buttons, emergency rundown unit, and door switches on a regular basis. Establishing the MRI quality assurance program We summarize our proposed QC tests and their frequencies in Table 2. What follows is a formulaic approach to monitor B0 inhomogeneity and geometrical distortion with weekly and daily QC tests using an ACR phantom. We found that incorporating these tests with the recommended weekly ACR tests run by a technologist and using only an MRI ACR phantom makes the process faster and more efficient in our busy clinic. First, we defined our reference B0 inhomogeneity and geometrical distortion during monthly and commissioning processes using Quasar MRID3D. The MRI image geometrical distortion and machine B0 inhomogeneity were defined over a 37 cm x 32 cm (W x L) area in three dimensions (Dx, Dy, Dz) and absolute value from MRI isocenter. Next, we scanned the regular ACR MRI phantom and defined B0 inhomogeneity using a bandwidth difference technique, and defined geometrical distortion using sagittal slices one and five for all three dimensions. We used geometrical distortion measurements for slice 5 (Dx, Dy) and the sagittal plane (Dz) for baseline calculation, assuming that slice five is at or very close to the MRI isocenter. Finally, the average baseline was defined based on equations 1 and 2 for the same slices and diameter on both MRID3D and ACR phantoms. The baseline measure is used for weekly checks, and we define our tolerance as 2% changes, and action level as a measured 4% difference. Equation 1: Base B0 = ACRB0 − MRID3DB0 Base B0 is an averaged B0 inhomogeneity at the same slice at MRID3D and ACR phantoms; MRID3DB0 is the measured average B0 inhomogeneity in ppm; and ACRB0 is averaged B0 inhomogeneity (ppm) using the bandwidth difference technique, and Equation 2: Basegeometrical distortion = ACRgeometrical distortion – MRID3Dgeometrical distortion Where Basegeometrical distortion is the geometrical distortion at the slice and orientation at MRID3D and ACR phantoms; ACRgeometrical distortion is the measured geometrical distortion at slice five and sagittal plane on all three directions (Dx, Dy, Dz); and MRID3Dgeometrical distortion is the measured geometrical distortion at the same ACR slice and orientation.

Results

The geometrical distortion over a 37 cm x 32 cm (W x L) area was evaluated in all three dimensions (Dx, Dy, Dz), absolute distance from MRI isocenter. Table 5 contains summary statistics; the maximum distortion in the x and y plane (axial plane) was 3.5 mm and 2.5 mm at the boundaries.
Table 5

Summary statistics for MRID3D geometrical distortion measurements.

STD: Standard deviation.

  Mean (mm) STD (mm) Max (mm) >2.5 mm (%)
dx 0.91 0.67 3.5 2
dy 0.52 0.39 2.51 0
dz 2.38 2.45 13.1 34
dr 2.79 2.36 13.19 40

Summary statistics for MRID3D geometrical distortion measurements.

STD: Standard deviation. The detailed measurements along all three coordinates and their absolute values with respect to MRI isocenter are shown in Figure 3. The B0 inhomogeneity along the z-direction was measured separately using an inverse gradient technique, and those data are also shown in Figure 3.
Figure 3

Static magnetic field (B0) inhomogeneity and distortion measurement results using an MRID3D phantom.

We used a QUASAR™ GRID3D Image Distortion Phantom and analysis system (Modus Medical, Canada) to evaluate image distortion in MR images due to the presence of an SRS frame and localizers. The data in the axial plane of the MRI images showed a maximum of 0.5 mm in the x-direction, 1.5 mm in the y-direction; in the z-direction, the maximum of 2.6 mm was observed at the phantom boundary (11 cm from MRI isocenter). The results showed a mean absolute deviation of 2.3 mm from isocenter. We defined our ACR phantom weekly B0 inhomogeneity and geometrical distortion baselines: the reference B0 inhomogeneity using MRID3D phantom was evaluated on all three axes (axial: 1.89 ppm, coronal: 0.135 ppm and sagittal: 0.068 ppm) as well as on average (0.699 ppm), all well-defined within our limits of ±2 ppm. The Base inhomogeneity based on Equation 1 was defined as 0.03 ppm. The Basegeometrical distortion was defined for our MRI ACR phantom at slice 5 as 0.6 mm in the x-direction, 0.8 mm in the y-direction, and 0.5 mm in the sagittal plane. All adopted ACR measures were within our defined tolerance, as summarized in Table 6. Specifically, the 20-channel and Rx/Tx RF head coils have been tested thoroughly for SNR, percent integral uniformity (PIU) and percentage signal ghosting (PSG), and the results were found to fall within our accepted limits.
Table 6

Magnetic resonance imaging stereotactic radiosurgery planning quality control results.

BW: Bandwidth; MRI: Magnetic resonance imaging; PMU: Phasor measurement unit; RF: Radiofrequency; Rx/Tx: Receiver/transmitter; PIU: Percent image uniformity; SNR: Signal-to-noise ratio.

MRI equipment evaluation summary
Setup and table position accuracy Pass
Center frequency Pass
Transmitter gain or attenuation Pass
Geometric accuracy measurements* Pass
High-contrast spatial resolution* Pass
Low-contrast detectability* Pass
Artifact evaluation Pass
Visual checklist Pass
Magnetic field homogeneity Pass
    Method of Testing BW diff
Slice-position accuracy* Pass
Slice-thickness accuracy* Pass
Radiofrequency coil checks (20-channel RF head coil)  
    a. SNR Pass
    b. Volume coil percent image uniformity Pass
    c. Percent signal ghosting Pass
Radiofrequency coil checks (Rx/Tx RF head coil)  
    a. SNR Pass
    b. Volume coil PIU Pass
    c. Percent signal ghosting Pass
Rx/Tx RF head coil check Pass
20-channel RF head coil check Pass
Dynamic field map Pass
Eddy current compensation Pass
Gradient delay Pass
Gradient sensitivity Pass
Body coil image brightness Pass
Magnet shim Pass
23 RX gain calibration Pass
24 Body coil tuning Pass
25 Spike Pass
26 PMU transmit Pass
27 Rx stability Pass
28 Tx stability Pass

Magnetic resonance imaging stereotactic radiosurgery planning quality control results.

BW: Bandwidth; MRI: Magnetic resonance imaging; PMU: Phasor measurement unit; RF: Radiofrequency; Rx/Tx: Receiver/transmitter; PIU: Percent image uniformity; SNR: Signal-to-noise ratio.

Discussion

The methodology discussed herein describes practical strategies we have implemented through lessons learned performing clinical MRI QA and SRS treatment planning [11-14]. We focus on discussion of major issues encountered during our QC procedures. The MRI machine specifications which have the highest potential to affect SRS treatment planning are B0 and B1 inhomogeneity and gradient non-linearity, which affect the geometrical accuracy and intensity uniformity of MRI images. Use of single-channel Rx/Tx RF head coils, a Leksell G frame (SRS frame) and accessories for SRS treatment planning only exacerbates these issues. Using an MRID3D phantom over a 37 cm x 32 cm (W x L) area gives enough information about MRI image distortion due to B0 inhomogeneity and gradient non-linearity to allow acquired images to be used for SRS treatment planning. As we expected, geometrical distortion is within 1 mm accuracy in the axial plane (x- and y-directions), and 2 mm along the z-direction, 10 cm from isocenter (almost head size), but it worsens to the order of 5 mm at the boundaries (16 cm away from isocenter). Immobilization devices constructed from materials optimized for radiation therapy may not necessarily be optimal for MRI (e.g., carbon fiber) [15-17]. In our experience, it is no longer sufficient for immobilization device materials (Leksell G frame, screws, adaptor and MRI localizer) to be simply MRI-compatible; these materials and devices should be MRI-optimal. Poor material choices can contribute to magnetic susceptibility-induced geometric distortions. Our phantom results specifically on 3D axial T2 SPACE (Sampling Perfection with Application optimized Contrasts using different flip angle Evolution) and axial T2 CISS (Three-dimensional constructive interference in steady state) sequences show artifacts even after pulse sequence optimization and use of different orientations. It is essential that some MRI sequences reviewed by the SRS team be repeated using a different sequence, such as 2D axial T2 or T1-weighted Turbo Spin Echo (TSE). However, our results from Quasar™ GRID3D shows that images acceptable for treatment planning can be obtained with the use of Laksell frame and localization devices by using the right MRI pulse sequence and a Tx/Rx RF head coil. Our proposed SRS MRI QA program has been reviewed and approved by our QA committee, and peer reviewed at every step by SRS committee members. Our aim is to minimize the scanning time and maximize efficiency. One major change proposed was the use of MRI ACR phantom for weekly geometrical accuracy checks rather than the MRID3D. This streamlines the process and the technologist can incorporate these results into the regular weekly checks. Our results indicated that gradient nonlinearity-induced geometric distortions can be severe and must be corrected using 3D distortion correction prior to using MR images for SRS treatment planning. However, even with 3D distortion correction, residual distortions can persist for large FOV prescriptions. One compounding factor is that some MRI scanners permit acquisition of image volumes positioned off-center from isocenter in the superior/inferior direction. This approach increases the likelihood of scanning in regions of nonlinear gradients and, therefore, increases the likelihood of residual distortions. At a minimum, the magnitude of these residual distortions should be characterized as a function of radial distance from isocenter for each scanner. Ideally, the residual distortions would be corrected. High MRI image intensity uniformity is critical in SRS treatment planning. Phased-array RF coils require correction for differences in the sensitivity profiles of each coil element during reconstruction to optimize image uniformity. These revisions, often based on a quick prescan image, become increasingly important when flexible phased-array RF coils, wrapped around the patient in various positions, are utilized. Our results indicate that by using prescan normalization and post-processing corrections the MRI images collected are within preset limits and SNR, PIU and PSG tests serve as good indications for variation. The participation of the dedicated SRS team, including the medical physicist, radiation oncologist, and neurosurgeon in the quantification, protocol modification and development of quality assurance procedures, as well as verification of MRI data used for SRS planning, is critical. Moreover, the scanner selection considerations, specifications, chosen MRI pulse sequences, and post-processing packages are critical in having a successful program of MRI-guided SRS treatment planning.

Conclusions

In conclusion, we describe an MRI machine QC procedure to maintain clinically acceptable MR image acquisition for SRS treatment planning purposes. MRI examinations for SRS planning can benefit from the improved localization and planning possible with the superior image quality and soft tissue contrast achieved with appropriate MRI QA. We recommend convening a team of experts who meet periodically to review cases, discuss new MRI pulse sequences and technology, including newly available post-processing software packages, and who can develop a custom QA program for the facility. We firmly believe this type of dialog opens opportunities for advance use of MRI images in SRS treatment planning, especially in a new era of MRI-guided radiotherapy available in commercial machines.
  15 in total

Review 1.  Sources of distortion in functional MRI data.

Authors:  P Jezzard; S Clare
Journal:  Hum Brain Mapp       Date:  1999       Impact factor: 5.038

Review 2.  Aspects of MR image distortions in radiotherapy treatment planning.

Authors:  A Fransson; P Andreo; R Pötter
Journal:  Strahlenther Onkol       Date:  2001-02       Impact factor: 3.621

3.  Intensity non-uniformity correction in MRI: existing methods and their validation.

Authors:  Boubakeur Belaroussi; Julien Milles; Sabin Carme; Yue Min Zhu; Hugues Benoit-Cattin
Journal:  Med Image Anal       Date:  2005-11-22       Impact factor: 8.545

Review 4.  New developments in MRI for target volume delineation in radiotherapy.

Authors:  V S Khoo; D L Joon
Journal:  Br J Radiol       Date:  2006-09       Impact factor: 3.039

Review 5.  Anatomical imaging for radiotherapy.

Authors:  Philip M Evans
Journal:  Phys Med Biol       Date:  2008-05-21       Impact factor: 3.609

6.  A two-step scheme for distortion rectification of magnetic resonance images.

Authors:  Lesley N Baldwin; Keith Wachowicz; B Gino Fallone
Journal:  Med Phys       Date:  2009-09       Impact factor: 4.071

7.  The influence of MRI scan position on image registration accuracy, target delineation and calculated dose in prostatic radiotherapy.

Authors:  S Hanvey; A H Sadozye; M McJury; M Glegg; J Foster
Journal:  Br J Radiol       Date:  2012-12       Impact factor: 3.039

Review 8.  MRI simulation for radiotherapy treatment planning.

Authors:  Slobodan Devic
Journal:  Med Phys       Date:  2012-11       Impact factor: 4.071

9.  Quality assurance methods and phantoms for magnetic resonance imaging: report of AAPM nuclear magnetic resonance Task Group No. 1.

Authors:  R R Price; L Axel; T Morgan; R Newman; W Perman; N Schneiders; M Selikson; M Wood; S R Thomas
Journal:  Med Phys       Date:  1990 Mar-Apr       Impact factor: 4.071

10.  Comprehensive QA for radiation oncology: report of AAPM Radiation Therapy Committee Task Group 40.

Authors:  G J Kutcher; L Coia; M Gillin; W F Hanson; S Leibel; R J Morton; J R Palta; J A Purdy; L E Reinstein; G K Svensson
Journal:  Med Phys       Date:  1994-04       Impact factor: 4.071

View more
  2 in total

1.  Geometric distortion in magnetic resonance imaging systems assessed using an open-source plugin for scientific image analysis.

Authors:  Takahiro Aoyama; Hidetoshi Shimizu; Ikuo Shimizu; Atsushi Teramoto; Naoki Kaneda; Kazuhiko Nakamura; Masaru Nakamura; Takeshi Kodaira
Journal:  Radiol Phys Technol       Date:  2018-09-25

2.  Evaluating the accuracy of geometrical distortion correction of magnetic resonance images for use in intracranial brain tumor radiotherapy.

Authors:  Seyed Mehdi Bagherimofidi; Claus Chunli Yang; Roberto Rey-Dios; Madhava R Kanakamedala; Ali Fatemi
Journal:  Rep Pract Oncol Radiother       Date:  2019-10-19
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.