Literature DB >> 29453685

Transcatheter Closure of Arterial Duct in Infants < 6 kg: Amplatzer Duct Occluder Type I vs Amplatzer Duct Occluder II Additional Sizes.

Giuseppe Santoro1, Mario Giordano2, Gianpiero Gaio2, Maria Teresa Palladino2, Giovanbattista Capozzi2, Carola Iacono2, Maria Giovanna Russo2.   

Abstract

Transcatheter closure of arterial duct (AD) remains challenging in low-weight patients and using Amplatzer Duct Occluder devices is still considered off-label in infants < 6 kg. This study aimed to report a large, single-center experience of percutaneous AD closure in infants < 6 kg as well as to compare the most frequently used devices, Amplatzer Duct Occluder type I (ADO I) and Amplatzer Duct Occluder type II Additional Sizes devices (ADO II-AS) (St. Jude Medical Corp, St. Paul, MN, USA). From March 2000 to March 2017, among the 762 patients submitted to percutaneous closure of AD at our Institution, 33 were infants < 6 kg (age 4.8 ± 2.1 months; weight 5.0 ± 0.9 kg). Fourteen patients (45%) underwent ADO I (Group I) and 19 patients ADO II-AS (Group II) device implantation. AD diameter was 2.6 ± 0.8 (range 1.5-4.0) mm resulting in QP/QS of 2.6 ± 0.0.9 (range 1.1-4.5). Successful device deployment was achieved in all patients without procedural morbidity or mortality. Procedural and fluoroscopy times were not significantly different between the groups. However, total X-ray absorbed dose was significantly higher in Group I (121 ± 69 vs 29 ± 16 mGy/cm2, p < 0.01). Immediate, 24 h, and mid-term (46 ± 37 months, median 39) complete occlusion were recorded in 72.7, 90.9, and 97% of patients, respectively without significant difference between the groups. In conclusion, transcatheter closure of AD with Amplatzer Duct Occluder devices is feasible, safe, and effective also in infants less than 6 kg, without significant difference between the most commonly used devices, namely, ADO I and ADO II-AS.

Entities:  

Keywords:  Arterial duct; Device; Low-weight infant

Mesh:

Year:  2018        PMID: 29453685     DOI: 10.1007/s00246-018-1828-5

Source DB:  PubMed          Journal:  Pediatr Cardiol        ISSN: 0172-0643            Impact factor:   1.655


  22 in total

1.  Comparison of percutaneous closure of large patent ductus arteriosus by multiple coils versus the Amplatzer duct occluder device.

Authors:  Giuseppe Santoro; Maurizio Cappelli Bigazzi; Maria Teresa Palladino; Maria Giovanna Russo; Marianna Carrozza; Raffaele Calabrò
Journal:  Am J Cardiol       Date:  2004-07-15       Impact factor: 2.778

2.  Comparison of the efficacy of different-sized Amplatzer duct occluders (I, II, and II AS) in children weighing less than 10 kg.

Authors:  Osman Baspinar; Ahmet Irdem; Ercan Sivasli; Derya Aydin Sahin; Metin Kilinc
Journal:  Pediatr Cardiol       Date:  2012-05-31       Impact factor: 1.655

3.  Long-term clinical experience with Amplatzer Ductal Occluder II for closure of the persistent arterial duct in children.

Authors:  Sok-Leng Kang; Gareth Morgan; Jonathan Forsey; Andrew Tometzki; Robin Martin
Journal:  Catheter Cardiovasc Interv       Date:  2014-01-31       Impact factor: 2.692

4.  Single-center experience in percutaneous closure of arterial duct with Amplatzer duct Occluder II additional sizes.

Authors:  Heba Talat Mahmoud; Giuseppe Santoro; Gianpiero Gaio; Fabio Angelo D'Aiello; Cristina Capogrosso; Maria Teresa Palladino; Maria Giovanna Russo
Journal:  Catheter Cardiovasc Interv       Date:  2016-11-10       Impact factor: 2.692

5.  Initial clinical experience with the Medtronic Micro Vascular Plug™ in transcatheter occlusion of PDAs in extremely premature infants.

Authors:  Shyam Sathanandam; Henri Justino; B Rush Waller; Wolfgang Radtke; Athar M Qureshi
Journal:  Catheter Cardiovasc Interv       Date:  2016-11-26       Impact factor: 2.692

6.  Novel use of the medtronic micro vascular plug for PDA closure in preterm infants.

Authors:  Elizabeth W Wang-Giuffre; John P Breinholt
Journal:  Catheter Cardiovasc Interv       Date:  2016-11-10       Impact factor: 2.692

7.  Transcatheter closure of moderate to large patent ductus arteriosus with the Amplatzer duct occluder.

Authors:  Jou-Kou Wang; Mei-Hwan Wu; Juey-Jen Hwang; Fu-Tein Chiang; Ming-Tai Lin; Hung-Chi Lue
Journal:  Catheter Cardiovasc Interv       Date:  2007-03-01       Impact factor: 2.692

8.  Closure of patent ductus arteriosus in children, small infants, and premature babies with Amplatzer duct occluder II additional sizes: multicenter study.

Authors:  Metin Sungur; Cemsit Karakurt; Nazan Ozbarlas; Osman Baspinar
Journal:  Catheter Cardiovasc Interv       Date:  2013-04-08       Impact factor: 2.692

9.  Transcatheter closure of persistent ductus arteriosus with the Amplatzer duct occluder in very young symptomatic children.

Authors:  G Butera; G De Rosa; M Chessa; L Piazza; A Delogu; A Frigiola; M Carminati
Journal:  Heart       Date:  2004-12       Impact factor: 5.994

10.  Percutaneous Patent Ductus Arteriosus (PDA) Closure in Very Preterm Infants: Feasibility and Complications.

Authors:  Carl H Backes; Sharon L Cheatham; Grace M Deyo; Scott Leopold; Molly K Ball; Charles V Smith; Vidu Garg; Ralf J Holzer; John P Cheatham; Darren P Berman
Journal:  J Am Heart Assoc       Date:  2016-02-12       Impact factor: 5.501

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.