Priyantha Edison1, Pi Sun Chang2, Guan Hong Toh3, Li Na Lee3, Sarat Kumar Sanamandra3, Varsha Atul Shah1.
Abstract
AIM: Guided by the ALARA - "As Low As Reasonably Achievable" principle in radiation safety, a quality improvement project to optimise the bedside diagnostic imaging process to the best standards of care was conducted over a six month period. The goal was too reduce the radiation hazard opportunities in the neonatal intensive care unit by at least 75% from the existing level at Q2/2015, within 6 months.Entities:
Keywords: ALARA; cumulative radiation exposure; preterm infants; radiation dosimetry; radiation safet
Year: 2017 PMID: 29450287 PMCID: PMC5699129 DOI: 10.1136/bmjoq-2017-000128
Source DB: PubMed Journal: BMJ Open Qual ISSN: 2399-6641
Number of radiographs performed in different gestational age groups
| Radiograph | 23–28 | 29–33 | 34–36 | 37–42 |
| Chest X-ray | 126 | 93 | 37 | 61 |
| Abdominal X-ray | 15 | 18 | 1 | 0 |
| Combined X-ray | 192 | 188 | 40 | 23 |
| Other X-rays | 14 | 38 | 7 | 6 |
| Total | 347 | 337 | 85 | 90 |
Figure 1Radiation hazard opportunities: pre and post project implementation. PDSA, Plan Do Study Act.
Figure 2Pre project and post-project survey.
Figure 3Radiograph discrepancy rate. PDSA, Plan Do Study Act.
Measurement of neonate’s radiation dose during mobile chest X-ray
| Unfors patient skin dosimeter | Neonate’s skin dose during chest X-ray (mGy) | Source to image distance (cm) | Exposure factors |
| 1 | 0.069 | 95 | 52/2.2 |
| 2 | 0.064 | 95 | 52/2.2 |
| 3 | 0.054 | 95 | 50/2.2 |
| 4 | 0.057 | 95 | 50/2.2 |
| 5 | 0.052 | 95 | 50/2.2 |
| 6 | 0.057 | 95 | 51/2.5 |
| 7 | 0.047 | 95 | 49/2.5 |
| 8 | 0.072 | 95 | 53/2.5 |
| 9 | 0.040 | 95 | 45/2.0 |
| 10 | 0.044 | 95 | 45/2.5 |
| Average skin dose (mGy) | 0.056 | ||
Unfors patient skin dosimeter was placed at the anterior of the left shoulder to measure the radiation dose during chest X-ray examination.
Measurement of scattered radiation using a 4 cm thick PMMA to simulate a neonate
| Mobile X-ray unit, Fuji GO | kV | mAs | SID | PMMA phantom (18×24 cm) | FOV |
| Exposure factors | 50 | 2.5 | 95 cm | 4 cm thick | 15×15 cm |
FOV, Field of view; PMMA, polymethylmethacrylate; SID, Source to image distance.
Scattered radiation measured at various distances using Unfors Xi Survey Meter
| Distance from centre of PMMA (cm) | Scattered radiation measured in μGy | Equivalent hours of background radiation (using natural background radiation of 3 mSv/year) |
| 30 | 0.227 | 0.663 |
| 50 | 0.106 | 0.309 |
| 100 | 0.029 | 0.085 |
| 200 | 0.007 | 0.020 |
PMMA, polymethylmethacrylate.