Literature DB >> 18375467

The additional dose to radiosensitive organs caused by using under-collimated X-ray beams in neonatal intensive care radiography.

Hanan Datz1, Avi Ben-Shlomo, David Bader, Siegal Sadetzki, Ada Juster-Reicher, Kyla Marks, Tatiana Smolkin, Samuel Zangen, Menachem Margaliot.   

Abstract

Radiographic technique and exposure parameters were recorded in five Israeli Neonatal Intensive Care Units for chest, abdomen and both chest and abdomen X-ray examinations. Equivalent dose and effective dose values were calculated according to actual examination field size borders and proper technique field size recommendations using PCXMC, a PC-based Monte Carlo program. Exposure of larger than required body areas resulted in an increase of the organ doses by factors of up to 162 (testes), 162 (thyroid) and 8 (thyroid) for chest, abdomen and both chest and abdomen examinations, respectively. These exposures increased the average effective dose by factors of 2.0, 1.9 and 1.3 for the chest, abdomen and both chest and abdomen examinations, respectively. Differences in exposure parameters were found between the different neonatal intensive care units-tube voltage, current-time product and focal to skin distance differences up to 13, 44 and 22%, respectively. Reduction of at least 50% of neonate exposure is feasible and can be implemented using existing methodology without any additional costs.

Mesh:

Year:  2008        PMID: 18375467     DOI: 10.1093/rpd/ncn090

Source DB:  PubMed          Journal:  Radiat Prot Dosimetry        ISSN: 0144-8420            Impact factor:   0.972


  4 in total

1.  Organ doses in preterm and full-term neonates and infants - a retrospective study on 1,064 chest radiographs.

Authors:  Birgit Kammer; Karl O Schneider; Evi Dell'Agnolo; Michael C Seidenbusch
Journal:  Pediatr Radiol       Date:  2022-03-18

2.  Balloon pulmonary valvuloplasty in neonates with critical pulmonary stenosis: Jugular or femoral.

Authors:  Erick Hoetama; Radityo Prakoso; Poppy Surwianti Roebiono; Indriwanto Sakidjan; Yovi Kurniawati; Sisca Natalia Siagian; Olfi Lelya; Anna Ulfah Rahajoe; Ganesja Moelia Harimurti; Oktavia Lilyasari
Journal:  Ann Pediatr Cardiol       Date:  2019-11-07

3.  Radiation dose to newborns in neonatal intensive care units.

Authors:  Mohammad Taghi Bahreyni Toossi; Malakeh Malekzadeh
Journal:  Iran J Radiol       Date:  2012-09-17       Impact factor: 0.212

4.  Reducing radiation hazard opportunities in neonatal unit: quality improvement in radiation safety practices.

Authors:  Priyantha Edison; Pi Sun Chang; Guan Hong Toh; Li Na Lee; Sarat Kumar Sanamandra; Varsha Atul Shah
Journal:  BMJ Open Qual       Date:  2017-10-21
  4 in total