Suitability of the porcine aortic model for transcatheter aortic root repair.

Changtian Wang1,2, Mario Lachat1, Evelyn Regar1, Ludwig Karl von Segesser3, Francesco Maisano1, Enrico Ferrari1,4.   

Abstract

OBJECTIVES: To treat aortic valve disease and concomitant root disease with transcatheter techniques, 'composite graft' implants are required. Our goal was to assess the suitability of the porcine aortic root for transcatheter root repair tests.
METHODS: Eight pig hearts explanted from domestic pigs used in experimental surgery were compared to data from the literature on human hearts. The measured diameters included those of the annulus, sinuses of Valsalva, coronary ostia, sinotubular junction, ascending aorta, innominate artery and aortic arch. The measured distances were from the coronary ostia to the nadir of the corresponding annulus; from the innominate artery to the nadir of the corresponding annulus; from the small curvature of the arch to the nadir of the corresponding annulus.
RESULTS: The mean weight of the pigs was 89 ± 5.4 kg. The mean aortic annulus diameter was 20 ± 1.2 mm (human: 23.0 ± 2.5 mm), the sinus of Valsalva diameter was 20.5 ± 0.5 mm (human: 31.4 ± 3.4 mm) and the sinotubular junction diameter was 20 ± 0.9 mm (human: 27.2 ± 3.0 mm). The diameter of the mean ascending aorta was 19 ± 0.7 mm (human: 29.3 ± 4 mm); the diameter of the innominate artery was 8.5 ± 0.7 mm, that of the aortic arch was 15 ± 0.7 mm and that of the coronary ostia was 5 ± 0.5 mm (left) and 4.7 ± 0.5 mm (right) (human: 4.8 ± 0.5 mm and 3.7 ± 0.9 mm). The distances from the left and right coronary orifices to the corresponding annuli were 8 ± 1.5 mm and 14 ± 2.4 mm, respectively (human: 14.7 ± 1.3 mm; 15.4 ± 1.7 mm). The distances from the innominate artery to the nadirs of the left and right coronary annuli were 44 ± 4.3 mm and 41 ± 4 mm (human: 80 ± 17 mm). The distance from the curvature of the small arch to the annulus was 35 ± 4.9 mm.
CONCLUSIONS: The porcine heart can be used as an experimental model to design and test new devices for catheter-based composite repair of the aortic root. Nevertheless, caution is required in using devices with tailored dimensions that must be adapted to the smaller pig's root.

Entities:  

Mesh:

Year:  2018        PMID: 29415164     DOI: 10.1093/icvts/ivx381

Source DB:  PubMed          Journal:  Interact Cardiovasc Thorac Surg        ISSN: 1569-9285


  3 in total

1.  Ex vivo biomechanical analysis of the Ross procedure using the modified inclusion technique in a 3-dimensionally printed left heart simulator.

Authors:  Yuanjia Zhu; Mateo Marin-Cuartas; Matthew H Park; Annabel M Imbrie-Moore; Robert J Wilkerson; Sarah Madira; Danielle M Mullis; Y Joseph Woo
Journal:  J Thorac Cardiovasc Surg       Date:  2021-09-16       Impact factor: 5.209

2.  Novel bicuspid aortic valve model with aortic regurgitation for hemodynamic status analysis using an ex vivo simulator.

Authors:  Yuanjia Zhu; Annabel M Imbrie-Moore; Michael J Paulsen; Bryant Priromprintr; Hanjay Wang; Haley J Lucian; Justin M Farry; Y Joseph Woo
Journal:  J Thorac Cardiovasc Surg       Date:  2020-06-29       Impact factor: 5.209

3.  Simultaneous transcatheter treatment of ascending aortic aneurysm with aortic and mitral regurgitation: an in vitro study.

Authors:  Yuxi Zhao; Tao Li; Mingwei Wu; Zhaoxiang Zeng; Minxin Gao; Xianhao Bao; Ziyi Xu; Jian Zhou; Jiaxuan Feng; Rui Feng; Zaiping Jing
Journal:  Interact Cardiovasc Thorac Surg       Date:  2021-08-18
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.