Lilie Wang1, Anthony J Cmelak1, George X Ding1.
Abstract
Entities:
Keywords: Eclipse AAA; Monte Carlo calculations; model-based dose calculation; skin dose
Mesh:
Year: 2018 PMID: 29411506 PMCID: PMC5849836 DOI: 10.1002/acm2.12275
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1Comparisons of surface doses (entrance and exit) measured by OSL dosimeters and the doses calculated by Monte Carlo (MC) for (a) 6 MV, and (b) 15 MV photon beams of field size 4 × 4 cm2, delivered 200 MU on a water slab of 20 cm thickness. The arrows indicate the in‐air data points for which MC calculations include simulation of OSL dosimeter of size 10 × 10 × 1.5 mm3. The OSL measurement uncertainty is 3–5%, and the MC calculation statistical uncertainty is 1%. Also shown are schematic diagram and the photo of experimental arrangement. One OSL detector is on the thin tape in air and the other is on the solid water as shown in the photo insert.
Figure 2An axial view of patient comparison of dose distributions for an H&N cancer patient treated with 6 MV VMAT beams. (a) Eclipse with default body contour; (b) Eclipse with external body contour including head mask and table; (c) Eclipse with an extended external body contour including table; (d) Monte Carlo calculated with CT images including table.
Figure 3(a) Axial CT slice showing the locations of the contoured skin structure on right neck (in yellow) for the H&N cancer patient; (b) Comparison of dose profiles calculated by AAA and MC for the H&N case treated with 6 MV VMAT beams. The dose profile is indicated by the white horizontal line in (a). For AAA, the default body contour, 2 cm‐enlarged body contour, and body contour including head mask are used in the calculations. The MC calculation uncertainty is 1%.
Figure 4(a) Axial CT slice showing the locations of the contoured skin structure on left supraclavicular area (in yellow) for the H&N cancer patient; (b) Comparison of dose profiles calculated by AAA and MC for the H&N case. The dose profile is indicated by the white vertical line in (a). For AAA, the default body contour, 2 cm‐enlarged body contour, and body contour including head mask are used in the calculations. The MC calculation uncertainty is 1%.
Figure 5Dose difference between AAA and MC, with and without the air gap, from skin surface to 2 cm depth. (a) for the H&N dose profiles in Fig. 3(b); (b) for the H&N dose profiles in Fig. 4(b).
Figure 6Comparison of dose volume histograms (DVH) for skin tissues and target for the H&N cancer patient, calculated from MC and AAA. For AAA, the default body contour, 2 cm‐enlarged body contour, and body contour including head mask are used in the calculations.
Mean percentage (%) skin dose and target dose for the H&N cancer case. Calculated by MC and Eclipse AAA of different scenarios (see text for details). The MC calculation uncertainty is 1%
| AAA, default | AAA, 2 cm air, w/tbl | AAA, mask w/tbl | MC | |
|---|---|---|---|---|
| Right neck | 75.1 | 83.9 | 83.9 | 87.6 |
| Left scv | 53.7 | 64.7 | 64.9 | 66.8 |
| PTV | 102.5 | 99.9 | 100 | 100 |
Mean percentage (%) skin dose for all the three cases. Calculated by MC and Eclipse AAA of different scenarios (see text for details). The MC calculation uncertainty is 1%
| Skin location | AAA, default | AAA, 2 cm air | MC | |
|---|---|---|---|---|
| 15 MV rib | Anterior | 64.5 | 78.9 | 75.1 |
| 6 MV lung | Anterior | 33.8 | 40.2 | 39.0 |
| Right | 20.2 | 24.8 | 24.8 | |
| 6 MV H&N | Right neck | 75.1 | 85.2 | 87.6 |
| Left scv | 53.7 | 65.5 | 66.8 |