As part of the method development, the injection volume as a critical quality attribute in fast fused-core chromatography was evaluated. Spilanthol, a pharmaceutically interesting N-alkylamide currently under investigation in our laboratory, was chosen as the model compound. Spilanthol was dissolved in both PBS and MeOH/H2O (70/30, v/v) and subsequently analyzed using a fused-core system hereby selecting five chromatographic characteristics (retention time, area, height, theoretical plates and symmetry factor) as responses. We demonstrated that the injection volume significantly influenced both the qualitative and quantitative performance of fused-core chromatography, a phenomenon which is confounded with the nature of the used sample solvent. From 2 μL up to 100 μL injection volume with PBS as solvent, the symmetry factor decreased favorably by 20%. Moreover, the theoretical plates and the quantitative parameters (area and height) increased up to 30%. On the contrary, in this injection volume range, the theoretical plates for the methanol-based samples decreased by more than 60%, while the symmetry factor increased and the height decreased, both by 30%. The injection volume is thus a critical and often overlooked parameter in fused-core method description and validation.Entities:
Keywords: Critical quality method-attribute (CQA); Fused-core (pore-shell, Halo®) HPLC; Injection volume; Spilanthol; Superficially porous particles (SPP); Topical patch
Year: 2013 PMID: 29403835 PMCID: PMC5760949 DOI: 10.1016/j.jpha.2013.02.002
Source DB: PubMed Journal: J Pharm Anal ISSN: 2214-0883
Fig. 1Chromatographic characteristics of PBS (A) and methanol-based (B) samples expressed as percentage relatively to the injection volume of 2 μL (=100%). Conditions: 0.1% formic acid in MeOH/H2O (70/30, v/v) mobile phase composition on aged column.
Fig. 2Typical chromatograms demonstrating the effect of injection volume for PBS and MeOH/H2O (70/30, v/v) as sample solvent. Conditions: 0.1% formic acid in MeOH/H2O (70/30, v/v) mobile phase composition on aged column.
Mean transdermal parameters for spilanthol in four different patches [mean±RSD (%)].
| Transdermal parameters | Patch composition (%TEC/±OA) | |||
| 5% TEC/−OA | 5% TEC/+OA | 20% TEC/−OA | 20% TEC/+OA | |
| Observed secondary parameters | ||||
| 0.50±60.08 | 0.52±44.88 | 0.64±12.72 | 0.43±22.50 | |
| 8.99±72.15 | 7.61±53.29 | 9.14±8.97 | 5.92±27.89 | |
| Lag time (h) | 6.22±30.15 | 5.64±40.28 | 5.22±27.59 | 6.70±19.32 |
| Apparent primary parameters | ||||
| 4.91±62.26 | 5.11±48.35 | 5.79±7.92 | 4.34±22.82 | |
| 4.51±28.70 | 5.87±64.53 | 5.34±30.08 | 5.00±19.01 | |
| 4.56±67.60 | 4.55±57.50 | 4.64±31.54 | 4.50±36.59 | |
n=10.
n=9.
n=7.