| Literature DB >> 29321059 |
Bas Habets1,2,3, J Bart Staal4,5, Marsha Tijssen6,5, Robert van Cingel6,5.
Abstract
OBJECTIVE: To determine the intrarater reliability of the Humac NORM isokinetic dynamometer for concentric and eccentric strength tests of knee and shoulder muscles.Entities:
Keywords: Isokinetic dynamometry; Knee; Muscle strength; Reliability; Shoulder; Technical report
Mesh:
Year: 2018 PMID: 29321059 PMCID: PMC5764011 DOI: 10.1186/s13104-018-3128-9
Source DB: PubMed Journal: BMC Res Notes ISSN: 1756-0500
Content and sequence of the test procedure for isokinetic tests of the knee flexors/extensors and shoulder rotators
| Knee flexors/extensors | Shoulder internal/external rotators |
|---|---|
| Concentric—concentric; 60°/s; 3 repetitionsa | Concentric—concentric; 60°/s; 3 repetitionsa |
| Concentric—concentric; 180°; 5 repetitionsa | Concentric—concentric; 120°/s; 5 repetitionsa |
| Concentric—eccentric extensors; 60°/s; 5 repetitionsa | Concentric—eccentric external rotators; 60°/s; 5 repetitionsa |
| Concentric—eccentric flexors; 60°/s; 5 repetitionsb | Concentric—eccentric external rotators; 120°/s; 5 repetitions |
aAfter each test, a 2 min rest interval was offered
bBetween the knee and shoulder tests, a 5 min rest interval was offered, followed by warming-up
Intramachine reproducibility values for concentric and eccentric knee tests
| ICC | SEM | SEM% | SDC | |
|---|---|---|---|---|
| KE concentric 60°/s | ||||
| Right | 0.840 (0.728–0.907) | 38.6 | 23.1 | 107.0 |
| Left | 0.762 (0.560–0.868) | 30.7 | 19.8 | 85.1 |
| KF concentric 60°/s | ||||
| Right | 0.744 (0.596–0.843) | 19.3 | 18.7 | 53.5 |
| Left | 0.804 (0.685–0.882) | 17.3 | 17.0 | 48.0 |
| KE concentric 180°/s | ||||
| Right | 0.829 (0.723–0.897) | 19.7 | 18.3 | 54.6 |
| Left | 0.861 (0.772–0.917) | 17.7 | 17.1 | 49.1 |
| KF concentric 180°/s | ||||
| Right | 0.762 (0.617–0.856) | 15.7 | 22.1 | 43.5 |
| Left | 0.826 (0.713–0.896) | 13.5 | 19.4 | 37.4 |
| KE eccentric 60°/s | ||||
| Right | 0.849 (0.755–0.910) | 27.7 | 14.8 | 76.8 |
| Left | 0.787 (0.659–0.871) | 33.9 | 19.3 | 94.0 |
| KF eccentric 60°/s | ||||
| Right | 0.891 (0.817–0.936) | 17.5 | 13.0 | 48.5 |
| Left | 0.840 (0.594–0.925) | 20.9 | 16.0 | 57.9 |
ICC intraclass correlation coefficient, CI confidence interval, SEM standard error of measurement, SDC smallest detectable change, KE knee extensors, KF knee flexors
Intramachine reproducibility values for concentric and eccentric shoulder tests
| ICC2,1 | SEM | SEM% | SDC | |
|---|---|---|---|---|
| SIR 60°/s | ||||
| Right | 0.935 (0.888–0.963) | 2.7 | 9.5 | 7.5 |
| Left | 0.924 (0.869–0.957) | 2.7 | 9.8 | 7.5 |
| SER concentric 60°/s | ||||
| Right | 0.898 (0.811–0.944) | 2.4 | 8.6 | 6.7 |
| Left | 0.934 (0.886–0.962) | 1.9 | 6.9 | 5.3 |
| SIR concentric 120°/s | ||||
| Right | 0.878 (0.786–0.931) | 3.7 | 14.4 | 10.3 |
| Left | 0.944 (0.897–0.969) | 2.4 | 9.8 | 6.7 |
| SER concentric 120°/s | ||||
| Right | 0.813 (0.662–0.896) | 3.0 | 12.1 | 8.3 |
| Left | 0.935 (0.882–0.964) | 1.8 | 7.3 | 5.0 |
| SER eccentric 60°/s | ||||
| Right | 0.783 (0.385–0.906) | 4.3 | 11.8 | 11.9 |
| Left | 0.905 (0.784–0.953) | 3.1 | 8.6 | 8.6 |
| SER eccentric 120°/s | ||||
| Right | 0.722 (0.555–0.834) | 4.6 | 13.0 | 12.8 |
| Left | 0.797 (0.665–0.881) | 3.9 | 11.2 | 10.8 |
ICC intraclass correlation coefficient, CI confidence interval, SEM standard error of measurement, SDC smallest detectable change, SIR shoulder internal rotators, SER shoulder external rotators