C K Ross1, N V Klassen, K R Shortt, G D Smith.
Abstract
Considerable effort has been devoted to measuring the absorbed dose to water using water calorimetry. Most of these efforts have been hampered by a lack of adequate knowledge of the heat defect of water. We argue that there is now sufficient information to establish with considerable confidence the heat defect of high-purity water containing various dissolved gases. For the present work we used water saturated with a 50/50 mixture of H2 and O2 gases, for which the heat defect is calculated to be -2.1%. As a test of this assignment, we have compared the absorbed dose to water as measured using water calorimetry with that obtained from Fricke dosimetry. The water calorimeter consisted of a small sealed vessel containing 100 ml of stirred water saturated with a 50/50 mixture of H2 and O2 gases. It was irradiated with 20 MV x-rays at a dose rate of about 0.4 Gy s-1. The same vessel was then filled with Fricke dosemeter solution, and irradiated under identical conditions. Our Fricke dosimetry is based on the Svensson and Brahme value of epsilon G (3.515 x 10(-3) 1 cm-1 J-1) and agrees to within 0.2% with the dose to water for 60Co gamma-rays obtained via graphite calorimetry. We find that for 20 MV x-rays, the dose to water determined by water calorimetry is 1.006 +/- 0.004 times the dose determined by Fricke dosimetry. Within 0.6(+/- 0.4)%, this result supports the calculated heat defect of -2.1% for water saturated with a 50/50 mixture of H2 and O2 gases.Entities:
Mesh:
Year: 1989 PMID: 2928376 DOI: 10.1088/0031-9155/34/1/003