Mitra Azadeh1, Boris Gorovits2, John Kamerud3, Stephen MacMannis4, Afshin Safavi5, Jeffrey Sailstad5, Perceval Sondag6.
Abstract
The accuracy of reported sample results is contingent upon the quality of the assay calibration curve, and as such, calibration curves are critical components of ligand binding and other quantitative methods. Regulatory guidance and lead publications have defined many of the requirements for calibration curves which encompass design, acceptance criteria, and selection of a regression model. However, other important aspects such as preparation and editing guidelines have not been addressed by health authorities. The goal of this publication is to answer many of the commonly asked questions and to present a consensus and the shared views of members of the ligand binding assay (LBA) community on topics related to calibration curves with focus on providing recommendations for the preparation and editing of calibration curves.Keywords: accuracy profile; calibration curve; curve editing; ligand binding assay; non-linear regression