Active content determination of non-coated pharmaceutical pellets by near infrared spectroscopy: method development, validation and reliability evaluation.

J Mantanus1, E Ziémons, P Lebrun, E Rozet, R Klinkenberg, B Streel, B Evrard, Ph Hubert.   

Abstract

A robust near infrared (NIR) method able to quantify the active content of pilot non-coated pharmaceutical pellets was developed. A protocol of calibration was followed, involving 2 operators, independent pilot batches of non-coated pharmaceutical pellets and two different NIR acquisition temperatures. Prediction models based on Partial Least Squares (PLS) regression were then carried out. Afterwards, the NIR method was fully validated for an active content ranging from 80 to 120% of the usual active content using new independent pilot batches to evaluate the adequacy of the method to its final purpose. Conventional criteria such as the R(2), the Root Mean Square Error of Calibration (RMSEC), the Root Mean Square Error of Prediction (RMSEP) and the number of PLS factors enabled the selection of models with good predictive potential. However, such criteria sometimes fail to choose the most fitted for purpose model. Therefore, a novel approach based on accuracy profiles of the validation results was used, providing a visual representation of the actual and future performances of the models. Following this approach, the prediction model using signal pre-treatment Multiplicative Scatter Correction (MSC) was chosen as it showed the best ability to quantify accurately the active content over the 80-120% active content range. The reliability of the NIR method was tested with new pilot batches of non-coated pharmaceutical pellets containing 90 and 110% of the usual active content, with blends of validation batches and industrial batches. All those batches were also analyzed by the HPLC reference method and relative errors were calculated: the results showed low relative errors in full accordance with the results obtained during the validation of the method, indicating the reliability of the NIR method and its interchangeability with the HPLC reference method. Copyright (c) 2009 Elsevier B.V. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 20152407     DOI: 10.1016/j.talanta.2009.10.019

Source DB:  PubMed          Journal:  Talanta        ISSN: 0039-9140            Impact factor:   6.057


  4 in total

1.  Calibration Curves in Quantitative Ligand Binding Assays: Recommendations and Best Practices for Preparation, Design, and Editing of Calibration Curves.

Authors:  Mitra Azadeh; Boris Gorovits; John Kamerud; Stephen MacMannis; Afshin Safavi; Jeffrey Sailstad; Perceval Sondag
Journal:  AAPS J       Date:  2017-12-27       Impact factor: 4.009

2.  High-Throughput NIR-Chemometric Method for Meloxicam Assay from Powder Blends for Tableting.

Authors:  Ioan Tomuta; Rares Iovanov; Andreea Loredana Vonica; Sorin E Leucuta
Journal:  Sci Pharm       Date:  2011-10-13

3.  RhB@MOF-5 Composite Film as a Fluorescence Sensor for Detection of Chilled Pork Freshness.

Authors:  Jingyi Li; Ning Zhang; Xin Yang; Xinting Yang; Zengli Wang; Huan Liu
Journal:  Biosensors (Basel)       Date:  2022-07-20

4.  Development of NIRS method for quality control of drug combination artesunate-azithromycin for the treatment of severe malaria.

Authors:  Chantal Boyer; Karen Gaudin; Tina Kauss; Alexandra Gaubert; Abdelhakim Boudis; Justine Verschelden; Mickaël Franc; Julie Roussille; Jacques Boucher; Piero Olliaro; Nicholas J White; Pascal Millet; Jean-Pierre Dubost
Journal:  J Pharm Biomed Anal       Date:  2012-04-23       Impact factor: 3.935

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.