Literature DB >> 29265700

Automatic calculation of patient size metrics in computed tomography: What level of computational accuracy do we need?

Sandra Sarmento1,2, Bruno Mendes1,2, Margarida Gouvêa3.   

Abstract

OBJECTIVES: To compare the effectiveness of two different patient size metrics based on water equivalent diameter (Dw ), the mid-scan water equivalent diameter Dw_c , and the mean (average) water equivalent diameter in the imaged region, Dw_ave , for automatic detection of accidental changes in computed tomography (CT) acquisition protocols.
METHODS: Patient biometric data (height and weight) were available from a previous survey for 80 adult chest examinations, and 119 adult single-acquisition chest-abdomen-pelvis (CAP) examinations for two 16 slice scanners (GE LightSpeed and Toshiba Aquilion RXL) equipped with automatic tube current modulation (ATCM). Dw_c and Dw_ave were calculated from the archived CT images. Size-specific dose estimates (SSDE) were obtained from volume CT dose index (CTDIvol ), using the conversion factors for a patient diameter of Dw_c .
RESULTS: CTDIvol and SSDE correlate better with Dw_ave than with Dw_c . R-squared values of linear fits to CTDIvol of CAP examinations were 0.81-0.89 for Dw_c and 0.93-0.94 for Dw_ave (SSDE: 0.69-080 for Dw_c , 0.87-0.92 for Dw_ave ). Percentage differences between Dw_c and Dw_ave were -4 ± 4% for chest and +5 ± 4% for CAP examinations (in % of Dw_ave ). However, small Dw variations translated as larger variations in CTDIvol for these ATCM systems (e.g., a 24% increase in Dw doubled CTDIvol ). The dependence of CTDIvol on Dw_ave was similar for chest and CAP examinations performed with similar ATCM parameters, while use of Dw_c resulted in a clear separation of the same data according to examination type. Maximum Dw variation in the imaged region was 5.6 ± 1.6 cm for chest and 6.5 ± 1.4 cm for CAP examinations.
CONCLUSIONS: Dw_ave is a better metric than Dw_c for binning similar-sized patients in dose comparison studies, despite the additional computational effort required for its calculation Therefore, when implementing automatic determination of Dw for SSDE calculations, automatic calculation of Dw_ave should be considered.
© 2017 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine.

Entities:  

Keywords:  computed tomography; examination doses; patient size metrics; water equivalent diameter

Mesh:

Year:  2017        PMID: 29265700      PMCID: PMC5768030          DOI: 10.1002/acm2.12240

Source DB:  PubMed          Journal:  J Appl Clin Med Phys        ISSN: 1526-9914            Impact factor:   2.102


INTRODUCTION

Computed tomography (CT) is a powerful diagnostic tool, but CT imaging protocols should be optimized to minimize radiation exposure. Diagnostic reference levels (DRLs) have been a powerful tool in dose optimization, by establishing typical values of volume CT Dose Index (CTDIvol) or dose‐length product (DLP), for certain types of examinations performed on standard‐sized (70 ± 3 kg) patients.1, 2, 3, 4 Modern CT scanners are equipped with automatic tube current modulation (ATCM), which adjusts tube current according to patient size and anatomical region, based on parameters set by the user.5 The functioning of an ATCM system and the conditions to be set depend on the scanner manufacturer. For GE and Toshiba systems, three parameters must be specified: an “image quality index” related to image noise, and the minimum and maximum values of tube current, I min and I max. An adequate value of I max avoids unnecessary dose escalation in large or obese patients. I min is equally important to prevent excessive noise in smaller patients, particularly in low attenuation regions such as the lungs.6, 7 GE defines a parameter called noise index (NI) to specify the noise level, while Toshiba uses the standard deviation (SD).8, 9 ATCM systems have some limitations, and pediatric patients need separate imaging protocols, with different parameters (such as lower kV and lower I min). The range of sizes in pediatric patients is immense, from babies to adolescents, and different protocols should be used according to child size/age.8, 10, 11 Only adult patients will be considered in the present work. Optimization of ATCM settings is a time‐consuming process involving a radiologist who assesses image quality after each acquisition. This is usually done for only a few patients. If the results are considered acceptable, the protocol is implemented in a provisional fashion. Postacquisition assessment of examination doses and image quality is continued for some time, to confirm that settings are optimized for all patient sizes.7 In this context, it is useful to have reference dose levels for different‐sized adult patients. The American Association of Physicists in Medicine (AAPM) lists approximate reference values for different weight ranges in some typical examination protocols.12, 13 Reliance on ATCM systems increases the potential detriment of nonoptimized settings and accidental changes to previously optimized protocols, as follows. An accidental increase in target image noise will result in a degradation of image quality for all patients, which should be quickly detected by radiologists. However, a decrease in target image noise will increase examination doses and image quality for smaller adults, while scanner output for large patients is limited by I max. Likewise, an unnecessarily high value of I min increases examination doses for small adults, with no degradation of image quality. Both situations lead to saturation of the tube current (at I max or I min) for an increased number of patients,7 but this may go unnoticed under a heavy workload, or be dismissed through overconfidence in the automated system. With the introduction of PACS (Picture Archiving and Communication Systems) in radiology, several vendors have developed radiation dose index monitoring (RDIM) software, which collects dosimetric information from imaging studies and stores it in a relational database.14 RDIM systems are a powerful tool to identify accidental changes and outliers, and determine where optimization is needed. However, patient biometric data (height and weight) are not usually available in PACS. Therefore, an accidental change which affects only small adults is hard to recognize quickly, because individual examination doses are still in the expected range (e.g., a 50‐kg adult imaged with a CTDIvol adequate for a 90‐kg patient). Naturally, dosimetric data from thousands of examinations will include patients of all sizes and can be compared between different institutions and scanners. But this is population‐dependent and also impractical for quick detection of changes and nonoptimized protocols. The AAPM Task Group 204 proposed the use of size‐specific dose estimates (SSDE) for patient dose comparisons. SSDE is an estimate of patient dose at the center of the imaged region, obtained from CTDIvol using conversion factors f(D eff) related to the effective diameter of the patient, calculated from the measured anteroposterior (AP) and lateral (LAT) patient dimensions, D eff = √(AP∙LAT).15 To improve the calculation of SSDE by taking into account patient attenuation, the AAPM Task Group 220 proposed describing patient size in terms of water equivalent diameter (D w). TG220 also suggested that D w could be calculated automatically by the CT scanner for all patients, with no user intervention, and the results stored in the DICOM header of CT images.16 An automatically calculated D w would allow binning of similar‐sized patients in RDIM databases for comparison of examination doses. SSDE values for adults may vary with patient size, depending on the ATCM system.17 Leng et al. have shown that SSDE can be calculated with less than 10% error using the examination CTDIvol and the value of D w obtained from the mid‐scan CT slice (D w_c).16 However, values of D w along the imaged region, D w(z), may be useful for estimating organ doses.16, 18 Obtaining D w(z) values requires longer computational times, but it also allows calculation of the mean value (average) of D w(z), D w_ave. As the response of ATCM systems is based on patient attenuation, D w_ave is the quantity more closely related to the examination's mean CTDIvol. 19Anam et al. recently reported on the implementation of automatic contouring for calculation of D w, and showed that D w_ave could be obtained with reasonable accuracy from only nine images, for head and thorax examinations.20 This is still nine times the computational effort required for calculating D w_c. Differences between D w_c and D w_ave were found to be less than 10%,20 which agrees well with data reported by other authors.21 The aim of this study was to compare D w_c and D w_ave as patient size metrics, for the purpose of ATCM optimization and detection of accidental changes; and to determine whether the difference between the two is sufficient to justify the additional computational effort required to automatically determine D w_ave in addition to D w_c. This study also assessed the interdependence of metrics and the feasibility of using D w metrics in nonautomated scenarios, for retrospective comparison with older data.

MATERIALS AND METHODS

Data collection

This study took advantage of existing biometric data, which had been collected during a routine internal survey, after confirmation by the radiologists that image quality was satisfactory. Patient biometric data (height and weight) were available for 80 chest and 119 single‐acquisition chest–abdomen–pelvis (CAP) examinations, performed in either of the two scanners available at our institution: a GE LightSpeed in use since 2011 (CT11) and the Toshiba Aquilion RXL acquired in March 2014 (CT14). The available data, summarized in Tables 1 and 2, pertain only to adult patients (21–89 years, mean 62 years), because pediatric examinations use separate protocols. CAP and chest examinations were chosen because they are frequently performed and because the corresponding protocols are used as a basis for the examination protocols of more complex examinations, which are harder to optimize independently.
Table 1

Summary of patient data for chest examinations. Weight, height, and BMI are indicated as mean values, with the range in brackets

# of PatientsWeight (kg)Height (cm)BMI (kg/m2)
CT1131 (24 M; 7 F)69 (45–105)167 (150–183)25 (16–38)
CT1449 (27 M; 22 F)69 (43–117)164 (147–183)25 (17–53)
Total8029 F64 (43–117)159 (147–175)26 (17–53)
51 M72 (45–105)169 (157–183)25 (16–38)
Table 2

Summary of patient data for CAP examinations. Weight, height, and BMI are indicated as mean values, with the range in brackets

No. of PatientsWeight (kg)Height (cm)BMI (kg/m2)
CT1141 (22 M; 19 F)65 (44–90)163 (145–180)25 (17–37)
CT1478 (38 M; 40 F)70 (43–104)163 (144–185)26 (17–34)
Total11959 F65 (43–90)159 (144–175)26 (17–37)
60 M71 (45–104)168 (145–185)5 (17–34)
Summary of patient data for chest examinations. Weight, height, and BMI are indicated as mean values, with the range in brackets Summary of patient data for CAP examinations. Weight, height, and BMI are indicated as mean values, with the range in brackets Both CT11 and CT14 are 16‐slice scanners, equipped with ATCM in both the longitudinal direction (z‐axis modulation) and the perpendicular plane (xy or angular modulation). The combination of these two is known as 3D modulation. Patients are randomly assigned to one scanner or the other, depending on equipment availability and internal logistics. Two orthogonal scout images (tube positions 0° and 90°) were acquired before each examination, in the order recommended by each manufacturer. The acquisition parameters are summarized in Table 3.
Table 3

Acquisition parameters used in both scanners

kVCollimation (mm)PitchTime (s)/rotImage quality index (NI/SD) I min (mA) I max (mA)
CT1112020 (16 × 1.25)1.3750.8NI = 18100440
CT1412032 (16 × 2)0.9380.5SD = 12.580/100a 500

While data were being collected, the I min value was changed for the CT14 scanner. This had little influence on patient doses. Therefore, only one dataset was considered. Image quality remained satisfactory to the radiologists in the department.

Acquisition parameters used in both scanners While data were being collected, the I min value was changed for the CT14 scanner. This had little influence on patient doses. Therefore, only one dataset was considered. Image quality remained satisfactory to the radiologists in the department. Proper functioning of the ATCM system and scanner indications of dosimetric parameters were checked at acceptance, and then annually, following the protocols and recommendations of the Spanish Medical Physics Society.22 To reduce patient dose in CT examinations, it is important to limit anatomical coverage (scan range) to the area of clinical concern.7 Appropriate restriction of anatomical coverage minimizes the scan length, whereas the optimization of ATCM parameters is reflected in the examination's mean CTDIvol. Both influence the dose‐length product (DLP). In this work, the examination CTDIvol (mean CTDIvol for the 32 cm diameter CTDI phantom) was chosen as the dosimetric parameter of interest and obtained from the dose summary archived in PACS.

Retrospective data analysis using attenuation metrics

Reconstructed CT images can be used to calculate D w, provided the reconstruction kernel is linear and quantitative (not edge‐enhancing or otherwise nonlinear).16 The image series obtained with the SOFT (GE) and FC08 (Toshiba) reconstruction kernels were used for this study. The field of view (FOV) used in chest and CAP examinations usually includes the outer contour of the patient. Visual observation of the CT images confirmed that, for the majority of the examinations, the whole contour of the skin was visible in the entire imaged region, except near the shoulders. Examinations where a large part of the patient's contour was outside the FOV were excluded from the dataset. These situations were too rare to justify a correction based on air border proportion, as suggested by Ikuta et al.23 According to the AAPM Task Group 220, the water equivalent diameter (D w) of an object is related to its water equivalent area (A w): .16 If ROI is the mean CT number in a ROI (region of interest) of area, AROI, containing the object, then A w can be determined from a CT image using eq. (1), 16: The air surrounding the object should have negligible impact on the result.16 To account for the attenuation of the CT table, A w (table) was determined by manually contouring the table (ROI_table) in one CT image for each scanner, and then substituting AROI_table and ROI_table in eq. (1). For each examination included in the study, a complete sequence of CT images (image series) was downloaded from the PACS and analyzed using ImageJ software (National Institute of Health, Bethesda, MA, USA), with a macro written by the authors. For each CT image, this macro extracted the values of table position (z) and tube current (I ) from the DICOM header, then drew a region of interest (ROI) encompassing the entire FOV, and determined its area (A ROI) and mean CT number (ROI). The results were transferred to a spreadsheet, and D w(z) was calculated using eq. (2), 16: The automated method to obtain D w was tested using two cylindrical acrylic phantoms (32 cm and 24 cm in diameter), filled with water, and imaged with the clinical protocol for chest. The D w results obtained were in good agreement (better than 0.2 cm) with expected values. The values of D w obtained with the automated method were compared with D w obtained from manual patient contouring in two images (one in the thorax and one in the abdomen) for a total of ten examinations (five in CT11 and five in CT14). The values of D w(z) obtained from patient images were used to calculate two different quantities: the mid‐scan or central D w_c, which is the value of D w(z) in the middle of the scanned region; and the D w_ave, calculated as the mean of all D w(z) values in the imaged region. D w_c and D w_ave were determined for all examinations, and SSDE was calculated as CTDIvol × f (D w_c), where f (D w_c) is the conversion factor related to a patient diameter of D w_c, obtained from AAPM tables.15, 16

Mathematically simulated scenarios

To simulate acquisitions with different values of I min, the original values of tube current were thresholded in the spreadsheet, so that all I(z) values below a certain I min were made equal to that I min. The mean of I(z) was calculated for each simulated scenario, and the corresponding examination CTDIvol determined. Simulated values of I min were 210 mA and 280 mA for CT11 and 140 mA and 180 mA for CT14. This approach has some limitations, because only values of I min higher than the original can be simulated. Moreover, it assumes a mere cutoff at the limiting value. Preliminary phantom measurements (B. Mendes, F. Dias, D. Oliveira, S. Sarmento, unpublished data) suggest that real ATCM systems are more sophisticated and adjust spatial variations according to the intended range. But the resulting difference between real and simulated CTDIvol was found to be less than 5% for these two scanners.

RESULTS

Dosimetric plots as a function of different metrics

Values of CTDIvol are plotted as a function of patient weight (a), D w_c (b), and D w_ave (c) in Figs. 1 and 2, for chest and CAP examinations performed in CT11 and CT14, respectively.
Figure 1

CTDI vol as a function of patient weight (a), Dw_c (b), and Dw_ave (c), for chest and CAP examinations performed in the CT11 scanner. AAPM reference values12, 13 are shown as horizontal/vertical lines (gray — chest and black — CAP).

Figure 2

CTDI vol as a function of patient weight (a), Dw_c (b), and Dw_ave (c), for chest and CAP examinations performed in the CT14 scanner. AAPM reference values12, 13 are shown as horizontal/vertical lines (gray — chest and black — CAP).

CTDI vol as a function of patient weight (a), Dw_c (b), and Dw_ave (c), for chest and CAP examinations performed in the CT11 scanner. AAPM reference values12, 13 are shown as horizontal/vertical lines (gray — chest and black — CAP). CTDI vol as a function of patient weight (a), Dw_c (b), and Dw_ave (c), for chest and CAP examinations performed in the CT14 scanner. AAPM reference values12, 13 are shown as horizontal/vertical lines (gray — chest and black — CAP). The approximate CTDIvol values for different weight ranges listed in AAPM protocols are shown for comparison.12, 13 The dispersion of CTDIvol data in Figs. 1(a) and 2(a) reflects the different heights of patients with similar weight, as well as different mass distributions in the body. When CTDIvol is plotted as a function of D w_c, in Figs. 1(b) and 2(b), there is a separation of data for chest and CAP examinations, related to the different anatomical location of the mid‐scan slice. D w_c is obtained in the middle of the lungs (low attenuation region) in chest examinations and closer to the liver in CAP examinations. Plotting CTDIvol as a function of D w_ave reduces data dispersion to a minimum, as shown in Figs. 1(c) and 2(c). Details become clearer, like the flattening of the curves for very small and very large patient sizes, which is probably related to I min and I max. As expected, the dependence of CTDIvol on D w_ave is the same for CAP and chest examinations — these examinations are performed with the same ATCM settings, in both scanners. SSDE values are plotted in Fig. 3(a) as a function of D w_c (CAP examinations) and as a function of D w_ave in Fig. 3(b) (CAP) and 3(c) (chest). The use of D w_ave reduces data dispersion in plots of SSDE, as it did for CTDIvol. To allow a more quantitative comparison, R‐squared values are presented in Table 4 for linear fits to CTDIvol and SSDE data for CAP examinations.
Figure 3

SSDE plotted as a function of Dw_c (a) for CAP examinations and as a function of Dw_ave for CAP (b) and chest (c) examinations, for both CT scanners.

Table 4

R‐squared values obtained for linear fits to CAP CTDIvol and SSDE values vs. different patient size metrics

R‐squared values for linear fits
CTDIvolSSDE
CT14CT11CT14CT11
Weight0.720.800.680.78
BMI0.730.780.730.76
D w_c 0.810.890.690.80
D w_ave 0.930.940.870.92
SSDE plotted as a function of Dw_c (a) for CAP examinations and as a function of Dw_ave for CAP (b) and chest (c) examinations, for both CT scanners. R‐squared values obtained for linear fits to CAP CTDIvol and SSDE values vs. different patient size metrics Before the widespread use of ATCM systems, Menke tested different surrogates for mean patient attenuation and concluded that the correlation between patient attenuation and body mass index (BMI) was no better than with patient weight.19 A similar result was obtained in this study, as shown in Table 4. In Fig. 4, CTDIvol and SSDE are plotted as a function of D w_ave, for the original dosimetric data in CT14 and for the mathematically simulated CAP examinations with I min values of 140 mA and 180 mA. When I min is set at 180 mA, CTDIvol remains approximately constant for small patients (D w_ave<25 cm) and then increases gradually until, for large patients (D w_ave ≈ 30 cm), it reaches the values obtained at lower I min settings. As a result, a high value of I min reduces the range of variation of CTDIvol with patient size. A similar effect is observed for SSDE (Fig. 4).
Figure 4

CTDI vol and SSDE plotted as a function of Dw_ave for the mathematically simulated CAP examinations in CT14 with Imin values of 140 mA and 180 mA. The original dosimetric data from the CAP examinations are plotted in the same graphs for comparison.

CTDI vol and SSDE plotted as a function of Dw_ave for the mathematically simulated CAP examinations in CT14 with Imin values of 140 mA and 180 mA. The original dosimetric data from the CAP examinations are plotted in the same graphs for comparison. The mean values and standard deviation (SD) of CTDIvol and SSDE are presented in Table 5, for the real examinations and for the simulated scenarios.
Table 5

Statistics of CTDIvol and SSDE, presented as mean values ± standard deviations, for the real examinations and the simulated scenarios

CTDIvol (mGy)SSDE (mGy)
CT1411.8 ± 4.014.6 ± 3.6
CT14_Imin14012.2 ± 3.515.2 ± 3.0
CT14_Imin18013.1 ± 2.916.5 ± 2.3
Statistics of CTDIvol and SSDE, presented as mean values ± standard deviations, for the real examinations and the simulated scenarios

Interdependence of different metrics

The patient sample considered in this work is representative of a particular population of oncological patients (Tables 1 and 2). Mean male and female heights agree well with known statistics for the Portuguese population in this age‐group.24 The data obtained in this work were compared with the earlier study by Menke of a different patient population, aged 18–87 years, with weight 46–108 kg, height 153–200 cm (mean 170 cm), and BMI 16–38 kg/m2. 19 In Fig. 5, D w_ave is plotted as a function of patient weight for chest and CAP examinations. The correlations found by Menke for chest and abdominal examinations are shown for comparison. For chest CT, the data from this study fall mostly within the 95% prediction limits previously obtained for this unrelated population 19 (Fig. 5), despite the difference in the obtained regression equation. It is unclear whether this difference results from intrinsic population metrics, or there is some additional bias in this study due to oncological risk factors and/or effect of oncological treatments.
Figure 5

Dw_ave plotted as a function of patient weight, for chest and CAP examinations. The broken lines represent the regression equations obtained for our own data, without separation of male and female patients (equations shown above), while the light gray lines represent the regression equation (thick middle line) and the 95% prediction limits (outer thin lines) obtained by Menke for chest and abdominal examinations, respectively.19

Dw_ave plotted as a function of patient weight, for chest and CAP examinations. The broken lines represent the regression equations obtained for our own data, without separation of male and female patients (equations shown above), while the light gray lines represent the regression equation (thick middle line) and the 95% prediction limits (outer thin lines) obtained by Menke for chest and abdominal examinations, respectively.19 For CAP examinations, the variation of D w_ave with patient weight is similar to that obtained by Menke for abdominal examinations (Fig. 5). There is good correlation between D w_c and D w_ave, for both examination types studied, as shown in Fig. 6. The two metrics are very similar, with maximum percentage differences below 15% (in % of D w_ave) as shown in Fig. 7 and summarized quantitatively in Table 6. There appears to be some separation of male and female patients, probably related to differences in body habitus. The mean D w_c − D w_ave difference for chest CT was −4 ± 4% (in % of D w_ave), which is comparable with the −1 ± 4% reported by Anam et al.20
Figure 6

Dw_c plotted as a function of Dw_ave for chest and CAP examinations. The dashed lines represent the linear fits to all the data (male and female patients). Fits to the two subsets (male and female) are shown only as equations.

Figure 7

Percentage difference between Dw_c and Dw_ave (above) and maximum absolute Dw variation (below), plotted as a function of Dw_ave, for all examinations and patients in the study.

Table 6

Differences between D w_c − D w_ave and D w(max) − D w(min), represented as mean ± standard deviation, for different examinations and patient groups (see plots in Fig. 7)

D w_c − D w_ave D w(max) − D w(min)
(% of D w_ave)(cm)(% of D w_ave)(cm)
ChestF−2 ± 3−0.6 ± 0.822 ± 95.4 ± 1.7
M−5 ± 4−1.3 ± 1.023 ± 76.1 ± 1.5
All−4 ± 4−1.0 ± 1.022 ± 75.8 ± 1.6
CAPF+3 ± 3+0.9 ± 0.823 ± 76.4 ± 1.5
M+8 ± 3+2.2 ± 0.924 ± 56.7 ± 1.4
All+5 ± 4+1.5 ± 1.124 ± 66.5 ± 1.4
Dw_c plotted as a function of Dw_ave for chest and CAP examinations. The dashed lines represent the linear fits to all the data (male and female patients). Fits to the two subsets (male and female) are shown only as equations. Percentage difference between Dw_c and Dw_ave (above) and maximum absolute Dw variation (below), plotted as a function of Dw_ave, for all examinations and patients in the study. Differences between D w_c − D w_ave and D w(max) − D w(min), represented as mean ± standard deviation, for different examinations and patient groups (see plots in Fig. 7) The variation of D w found in each examination, D w_max − D w_min, is similar for male and female patients, as shown in Fig. 7 and Table 6. The values agree well with those reported by Leng et al. for a different population (weight 37–183 kg, mean 85 kg, BMI 15–57 kg/m2, mean 29 kg/m2), where D w variation was 5.2 ± 1.4 cm for chest and 6.5 ± 1.3 cm for CAP examinations, and maximum D w_max − D w_min was 10.5 cm (32% of D w_ave).21 In this study, D w variation was 5.8 ± 1.6 cm for chest and 6.5 ± 1.4 cm for CAP examinations (Table 6).

Feasibility of small‐scale studies in nonautomated scenarios

At our institution, D w_ave proved easier to obtain than patient weight. Recording patient biometric data requires technologist time and interferes with workflow, while downloading images from PACS may be carried out by medical physicists elsewhere. Using D w_ave also allowed retrospective studies and comparisons, for all CT examinations where full FOV images were available.

ANALYSIS AND DISCUSSION

Despite the very similar values of D w_c and D w_ave, total examination doses (CTDIvol or SSDE) clearly have a stronger correlation with D w_ave than with D w_c, as reflected by the lower dispersion of dosimetric data (Figs. 1, 2, and 3, Table 4). The values of D w_c probably reflect both variations in scan length (which alter the anatomical location of the midscan slice) and localized anatomy characteristics like abdominal obesity, or large breasts in some female patients. This makes D w_ave the most advantageous metric for the purpose of protocol optimization and automatic detection of outliers or accidental changes, despite the additional computational effort involved in its calculation. Another advantage of D w_ave is the similar dependence of CTDIvol and SSDE for both examination types (Figs. 1 and 2). Identifying examination type can be challenging for automatic systems, because examination and protocol nomenclature are rarely standardized. Moreover, some examinations have more than one sequence (e.g., before and after contrast injection), making it difficult to compare total DLP.25 Our data suggest that comparing CTDIvol vs. D w_ave for a group of different examinations (performed with similar protocols) may be a feasible alternative in some situations. The mathematically simulated scenario with I min = 180 mA illustrates how patient doses may increase by nearly 50% for adults under 70 kg, while mean examination doses increase by only 12–13%, and maximum values are not exceeded (Fig. 4 and Table 5). This highlights the importance of establishing diagnostic reference levels as a function of patient size, to allow quick detection of nonoptimized protocols by dose monitoring software. Comparison with established references could also be made before irradiation, if D w were determined from scout images as suggested by TG220.16 It is not the purpose of this work to determine ideal dose levels for small and large adults, or discuss whether CTDIvol should increase linearly with patient size. For adult patients, optimum variation of examination dose with size remains a matter of debate. Noise constant systems such as GE and Toshiba result in a linear increase in CTDIvol with patient weight, but some ATCM systems intentionally decrease dose less for thinner adults.26 A comparison of CT scanners from three manufacturers showed that the Philips system had the least variation of DLP with patient weight, when compared with GE and Siemens.27 Some authors using GE scanners divide adult patients into weight categories.7 More data are necessary, especially as automatic selection of tube voltage may soon be a widespread option as well.28 The example presented here merely highlights the importance of choosing a patient size metric which reduces data dispersion to a minimum, to improve detection of normal trends and outliers. The comparisons shown in Fig. 5 are an encouraging result, suggesting that the study of a sufficiently large number of different populations and anatomical regions might provide a conversion between patient weight and D w_ave, for each examination type. This would allow comparison of newer large‐scale data based on D w metrics with the existing studies and standards based on patient weight. As reported by other authors, the impact of D w_ave − D w_c differences on SSDE values is small,21 because the two metrics have quite similar values. This is reflected in the small data dispersion seen in Fig. 3(b) and 3(c). However, the lower dispersion of dosimetric data when D w_ave is used as a metric for patient size suggests that the effect of small D w_ave − D w_c differences is probably amplified by the large variation in CTDIvol (from ~8 mGy to ~16 mGy, a nearly 100% increase) which occurs for a relatively small increase in D w_ave (from ~25 cm to ~31 cm, ~24% increase), as shown in Figs. 1(c) and 2(c). The effect may be less pronounced when using ATCM systems from different vendors. These results highlight the importance of automatic calculation of D w_ave, despite the additional computational effort involved. Moreover, these data suggest that D w_ave needs to be determined with great accuracy and in a standardized manner, if it is to be used for comparisons between different CT scanners and different institutions. In this work, the mean difference between automated and manual D w was found to be 0.2 ± 1.2% (0.05 ± 0.32 cm), but the mean absolute difference was 1.0 ± 0.7% (0.26 ± 0.19 cm), with a maximum difference of 2.2% (0.6 cm). Drawing a ROI including the entire FOV is computationally fast and simple. But the greater accuracy (maximum 0.5% difference) reported for automatic contouring20 should prove useful for dose comparisons, particularly during initial studies and acquisition of baseline data.

CONCLUSIONS

This study highlights the importance of automatic calculation of D w(z), not just for organ dose estimation as already recommended,16, 18 but also to make D w_ave available to end users as a patient size metric for binning similar‐sized patients in RDIM systems. Despite the small percentage difference between D w_c and D w_ave (−4 ± 4% for chest and +5 ± 4% for CAP examinations in this study), both CTDIvol and SSDE present a stronger correlation with D w_ave than they do with D w_c. Our data suggest D w_c values reflect localized anatomy characteristics. The lower dispersion of dosimetric data obtained with D w_ave makes it easier to identify trends and outliers. This is useful for ATCM optimization and detection of accidental changes. Use of D w_ave also reduces dependence on examination type, which may be difficult to identify accurately in large‐scale databases. Therefore, when implementing automatic determination of D w_c for SSDE calculations, automatic calculation of D w_ave should definitely be considered as well, despite the additional computational effort involved. Use of D w metrics is not yet widely implemented in CT scanners and RDIM systems, but it is important to acquire baseline data for D w_ave metrics and to establish comparisons with existing standards based on patient weight. Our experience shows that small‐scale studies using D w_ave metrics are feasible in nonautomated scenarios and may be used initially to acquire baseline data from new and retrospective studies.

CONFLICT OF INTEREST

The authors declare that they have no conflict of interests.
  21 in total

1.  Diagnostic reference levels from the ACR CT Accreditation Program.

Authors:  Cynthia McCollough; Theresa Branham; Vince Herlihy; Mythreyi Bhargavan; Lavonne Robbins; Krista Bush; Michael McNitt-Gray; J Thomas Payne; Tom Ruckdeschel; Doug Pfeiffer; Dianna Cody; Robert Zeman
Journal:  J Am Coll Radiol       Date:  2011-11       Impact factor: 5.532

Review 2.  Reducing radiation dose in body CT: a practical approach to optimizing CT protocols.

Authors:  Alice R Goldman; Pierre D Maldjian
Journal:  AJR Am J Roentgenol       Date:  2013-04       Impact factor: 3.959

3.  Comparison of different body size parameters for individual dose adaptation in body CT of adults.

Authors:  Jan Menke
Journal:  Radiology       Date:  2005-08       Impact factor: 11.105

4.  Computed tomography automatic exposure control techniques in 18F-FDG oncology PET-CT scanning.

Authors:  Gareth R Iball; Deborah Tout
Journal:  Nucl Med Commun       Date:  2014-04       Impact factor: 1.690

5.  Size-specific Dose Estimates for Chest, Abdominal, and Pelvic CT: Effect of Intrapatient Variability in Water-equivalent Diameter.

Authors:  Shuai Leng; Maria Shiung; Xinhui Duan; Lifeng Yu; Yi Zhang; Cynthia H McCollough
Journal:  Radiology       Date:  2015-02-25       Impact factor: 11.105

Review 6.  Techniques and applications of automatic tube current modulation for CT.

Authors:  Mannudeep K Kalra; Michael M Maher; Thomas L Toth; Bernhard Schmidt; Bryan L Westerman; Hugh T Morgan; Sanjay Saini
Journal:  Radiology       Date:  2004-10-21       Impact factor: 11.105

7.  Size-specific dose estimates for adult patients at CT of the torso.

Authors:  Jodie A Christner; Natalie N Braun; Megan C Jacobsen; Rickey E Carter; James M Kofler; Cynthia H McCollough
Journal:  Radiology       Date:  2012-10-22       Impact factor: 11.105

8.  Estimating patient dose from x-ray tube output metrics: automated measurement of patient size from CT images enables large-scale size-specific dose estimates.

Authors:  Ichiro Ikuta; Graham I Warden; Katherine P Andriole; Ramin Khorasani; Aaron Sodickson
Journal:  Radiology       Date:  2013-10-28       Impact factor: 11.105

9.  AAPM medical physics practice guideline 6.a.: Performance characteristics of radiation dose index monitoring systems.

Authors:  Dustin A Gress; Renee L Dickinson; William D Erwin; David W Jordan; Robert J Kobistek; Donna M Stevens; Mark P Supanich; Jia Wang; Lynne A Fairobent
Journal:  J Appl Clin Med Phys       Date:  2017-05-12       Impact factor: 2.102

10.  Automated Calculation of Water-equivalent Diameter (DW) Based on AAPM Task Group 220.

Authors:  Choirul Anam; Freddy Haryanto; Rena Widita; Idam Arif; Geoff Dougherty
Journal:  J Appl Clin Med Phys       Date:  2016-07-08       Impact factor: 2.102

View more
  3 in total

1.  An Improved Method of Automated Noise Measurement System in CT Images.

Authors:  Choirul Anam; Idam Arif; Freddy Haryanto; Fauzia P Lestari; Rena Widita; Wahyu S Budi; Heri Sutanto; Kusworo Adi; Toshioh Fujibuchi; Geoff Dougherty
Journal:  J Biomed Phys Eng       Date:  2021-04-01

2.  Real-time fully automated dosimetric computation for CT images in the clinical workflow: A feasibility study.

Authors:  Massimiliano Porzio; Choirul Anam
Journal:  Front Oncol       Date:  2022-08-11       Impact factor: 5.738

3.  Size-Specific Dose Estimates of Radiation Based on Body Weight and Body Mass Index for Chest and Abdomen-Pelvic CTs.

Authors:  Jian Xu; Xiangquan Wang; Panfeng Yang; Kuangnan Luo; Xiaolong He
Journal:  Biomed Res Int       Date:  2020-07-10       Impact factor: 3.411

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.