Literature DB >> 29235898

Flash Glucose Measurements in Children with Type 1 Diabetes in Real-Life Settings: To Trust or Not to Trust?

Agnieszka Szadkowska1, Andrzej Gawrecki2, Arkadiusz Michalak1, Dorota Zozulińska-Ziółkiewicz2, Wojciech Fendler1,3, Wojciech Młynarski1.   

Abstract

BACKGROUND AND AIMS: To evaluate the clinical accuracy of a flash glucose monitoring device FreeStyle Libre (FSL) among children with type 1 diabetes in real-world settings during a summer camp.
MATERIALS AND METHODS: During a summer camp, children with type 1 diabetes (n = 79, aged 8-18 years) were provided with FSLs for 12 days. On days 3, 7, and 11 of the study, they underwent supervised glucose testing at 8 timepoints. Glycemia was estimated by using FSL and measured with a personal glucometer within a period of 2 min. The glucose trend arrows were recorded.
RESULTS: The study was completed by 78 children (median: age 12.8 years, diabetes duration 5.8 years, HbA1c 58.5 mmol/mol). Mean absolute relative difference (MARD) between the FSL and the glucometer was 13.5% ± 12.9%. FSL was the most accurate in stable glycemic conditions: MARD 11.4% ± 10.4%, less accurate when glycemia was falling >2 mg/(dL·min) [0.111 mmol/(L·min)-MARD 22.6% ± 18.6%; P < 0.001 vs. stable conditions] and when the device could not determine the glucose trends (16.5% ± 16.3%, P = 0.01 vs. stable conditions). The FSL demonstrated lower accuracy during the day than the night [MARD 14.9% ± 14% vs. 11.2% ± 10.6%, P < 0.0001]. Out of 1655 data pairs of glucometer and FSL, using the Surveillance Error Grid methodology we determined that 80.36% of FSL readings were associated with no clinical risk, 18.73% with slight risk and only one high-risk measurement was detected.
CONCLUSION: FSL is accurate in children, but its accuracy depends on the glucose trend. Results flagged by the rapid fall flag and "trend undetermined" should be verified by blood glucose measurements.

Entities:  

Keywords:  Accuracy; Blood glucose self-monitoring; Children; Flash glucose monitoring; FreeStyle Libre; Type 1 diabetes mellitus

Mesh:

Substances:

Year:  2017        PMID: 29235898     DOI: 10.1089/dia.2017.0287

Source DB:  PubMed          Journal:  Diabetes Technol Ther        ISSN: 1520-9156            Impact factor:   6.118


  8 in total

1.  Estimating plasma glucose with the FreeStyle Libre Pro continuous glucose monitor during oral glucose tolerance tests in youth without diabetes.

Authors:  Nejla Ghane; Miranda M Broadney; Elisabeth K Davis; Robert W Trenschel; Shavonne M Collins; Sheila M Brady; Jack A Yanovski
Journal:  Pediatr Diabetes       Date:  2019-08-29       Impact factor: 4.866

Review 2.  Type 1 diabetes mellitus management in young children: implementation of current technologies.

Authors:  Michal Nevo-Shenker; Moshe Phillip; Revital Nimri; Shlomit Shalitin
Journal:  Pediatr Res       Date:  2019-11-12       Impact factor: 3.756

Review 3.  Technologies for Diabetes Self-Monitoring: A Scoping Review and Assessment Using the REASSURED Criteria.

Authors:  Jessica Hanae Zafra-Tanaka; David Beran; Beatrice Vetter; Rangarajan Sampath; Antonio Bernabe-Ortiz
Journal:  J Diabetes Sci Technol       Date:  2021-03-09

Review 4.  Flash Glucose Monitoring: A Review of the Literature with a Special Focus on Type 1 Diabetes.

Authors:  Giulia Mancini; Maria Giulia Berioli; Elisa Santi; Francesco Rogari; Giada Toni; Giorgia Tascini; Roberta Crispoldi; Giulia Ceccarini; Susanna Esposito
Journal:  Nutrients       Date:  2018-07-29       Impact factor: 5.717

Review 5.  Use of Continuous Glucose Monitoring in the Assessment and Management of Patients With Diabetes and Chronic Kidney Disease.

Authors:  James Ling; Jack K C Ng; Juliana C N Chan; Elaine Chow
Journal:  Front Endocrinol (Lausanne)       Date:  2022-04-22       Impact factor: 6.055

6.  Safety and feasibility of a factory-calibrated continuous glucose monitoring system in term and near-term infants at risk of hypoglycemia.

Authors:  Eri Nishimura; Shuntaro Oka; Junichi Ozawa; Kosuke Tanaka; Taichi Momose; Kazuhiko Kabe; Fumihiko Namba
Journal:  Turk Arch Pediatr       Date:  2021-01-12

7.  An Evaluation of the Accuracy of a Flash Glucose Monitoring System in Children with Diabetes in comparison with Venous Blood Glucose.

Authors:  Bingyan Cao; Rui Wang; Chunxiu Gong; Di Wu; Chang Su; Jiajia Chen; Yajun Yi; Min Liu; Xuejun Liang; Wenjing Li
Journal:  J Diabetes Res       Date:  2019-09-09       Impact factor: 4.011

8.  Comparison between a flash glucose monitoring system and a portable blood glucose meter for monitoring dogs with diabetes mellitus.

Authors:  Francesca Del Baldo; Claudia Canton; Silvia Testa; Harry Swales; Ignazio Drudi; Stefania Golinelli; Federico Fracassi
Journal:  J Vet Intern Med       Date:  2020-10-30       Impact factor: 3.333