| Literature DB >> 29208560 |
H-S Huang1, C-L Tsai2, J Chang3, T-C Hsu4, S Lin5, C-C Lee6.
Abstract
OBJECTIVES: To provide a summary of evidence for the diagnostic accuracies of three multiplex PCR systems (mPCRs)-BioFire FilmArray RP (FilmArray), Nanosphere Verigene RV+ test (Verigene RV+) and Hologic Gen-Probe Prodesse assays-on the detection of viral respiratory infections.Entities:
Keywords: Multiplex PCR; Point-of-care test; Respiratory virus infection
Mesh:
Year: 2017 PMID: 29208560 PMCID: PMC7128951 DOI: 10.1016/j.cmi.2017.11.018
Source DB: PubMed Journal: Clin Microbiol Infect ISSN: 1198-743X Impact factor: 8.067
Characteristics of BioFire FilmArray RP, Nanosphere Verigene RV+ Test and Hologic Gen-Probe Prodesse assays
| Name | BioFire FilmArray | Verigene | GenProbe Prodesse |
|---|---|---|---|
| Technology | Melting curve analysis | Gold nanoparticles with silver signal amplification | Melting curve analysis |
| Assays | Respiratory panel | Respiratory virus plus test | ProFlu+, ProFAST+, ProAdeno+, ProParaflu+, Pro hMPV+ |
| Targets | Adenovirus Coronavirus HKU1 Coronavirus NL63 Coronavirus 229E Coronavirus OC43 hMPV Human Rhinovirus/enterovirus FluA FluA/H1 FluA/H3 FluA/H1–2009 Influenza B Parainfluenza virus 1 Parainfluenza virus 2 Parainfluenza virus 3 Parainfluenza virus 4 RSV | FluA-H1 FluA-2009 H1N1 FluA-H3 FluA Influenza B RSV A RSV B | ProFlu+: FluA, influenza B, RSV ProFAST+: Seasonal FluA/H1, seasonal FluA/H3, 2009 H1N1 influenza ProAdeno+: Adenovirus ProParaflu+: Parainfluenza 1, parainfluenza 2, parainfluenza 3 Pro hMPV+: hMPV |
| Throughput | 1 sample per instrument | 1 sample per processor | 14 samples per run |
| Run time (hours) | 1 | <2.5 | 4–5 |
| Hands-on time | 2 minutes | 5 minutes | 1.5 hours |
| Sample preparation included? | Yes | Yes | No |
| Reagent storage conditions | Room temperature | 2–8°C and −20°C | −70°C |
FluA, influenza A virus; hMPV, human metapneumovirus; RSV, respiratory syncytial virus.
Fig. 1PRISMA flow diagram for selection of articles for meta-analysis.
Characteristics of the 20 included studies (25 data sets)
| Characteristic | |
|---|---|
| Age | |
| Children | 8 (32%) |
| Mixed | 10 (40%) |
| Unclear | 7 (28%) |
| Index test | |
| FilmArray RP | 10 (40%) |
| Verigene RV+ | 6 (24%) |
| Prodesse ProFlu+ | 9 (36%) |
| Reference standard | |
| Virus culture/direct fluorescent antibody | 4 (16%) |
| RT-PCR | 4 (16%) |
| Composite | 7 (28%) |
| Discrepant analysis | 10 (40%) |
| Specimen type | |
| Upper respiratory specimen | 22 (88%) |
| Mixed upper and lower respiratory specimen | 2 (8%) |
| Unclear | 1 (4%) |
| Study design | |
| Prospective | 11 (44%) |
| Retrospective | 12 (48%) |
| Mixed | 2 (8%) |
| Country | |
| United States | 18 (72%) |
| Belgium | 3 (12%) |
| Korea | 1 (4%) |
| Japan | 2 (8%) |
| Taiwan | 1 (4%) |
RT-PCR, reverse transcription–PCR.
Fig. 2QUADAS-2 for included studies. (a) QUADAS-2 bias assessment. (b) QUADAS-2 applicability assessment. QUADAS-2, Quality Assessment of Diagnostic Accuracy Studies 2.
Accuracy estimates of included studies
| Test | Sensitivity (95% confidence interval) | Specificity (95% confidence interval) | LR+ (95% confidence interval) | LR− (95% confidence interval) | AUC (95% confidence interval) |
|---|---|---|---|---|---|
| Influenza A virus | |||||
| FilmArray | 0.911 (0.848, 0.949) | 0.995 (0.988, 0.998) | 186 (74.9, 368) | 0.0928 (0.052, 0.153) | 0.99 (0.98, 1) |
| Verigene | 0.949 (0.882, 0.979) | 0.982 (0.944, 0.995) | 65.2 (15.9, 185) | 0.058 (0.0206, 0.122) | 0.99 (0.98, 1) |
| Prodesse | 0.954 (0.871, 0.985) | 0.983 (0.973, 0.989) | 57.65 (40.48, 76.94) | 0.053 (0.022, 0.0896) | 0.99 (0.99, 1) |
| Summary | 0.940 (0.902, 0.964) | 0.987 (0.979, 0.992) | 76.9 (42.4, 126) | 0.06 (0.03, 0.101) | 0.99 (0.98, 1) |
| Influenza B virus | |||||
| FilmArray | 0.822 (0.689, 0.905) | 0.994 (0.980, 0.998) | 167.50 (40.9, 503.00) | 0.188 (0.093, 0.313) | 0.98 (0.94, 1) |
| Prodesse | 0.963 (0.907, 0.986) | 0.992 (0.969, 0.998) | 136.73 (30.5, 385) | 0.04 (0.014, 0.097) | 0.99 (0.99, 1) |
| Summary | 0.932 (0.877, 0.963) | 0.993 (0.986, 0.997) | 154.4 (66.5, 304) | 0.072 (0.034, 0.124) | 0.99 (0.99, 1) |
| RSV | |||||
| FilmArray | 0.911 (0.821, 0.958) | 0.987 (0.971, 0.994) | 73.1 (29.4, 150) | 0.09 (0.0412, 0.172) | 0.98 (0.98, 0.99) |
| Verigene | 0.977 (0.929, 0.993) | 0.993 (0.962, 0.999) | 219.30 (23.5, 868) | 0.027 (0.0076, 0.072) | 0.99 (0.98, 1) |
| Summary | 0.942 (0.84, 0.972) | 0.991 (0.980, 0.996) | 109.45 (47.5, 226) | 0.06 (0.03, 0.118) | 0.99 (0.99, 1) |
| Adenovirus | |||||
| FilmArray | 0.670 (0.516, 0.794) | 0.991 (0.961, 0.998) | 86.9 (20.3, 273) | 0.337 (0.212, 0.494) | 0.89 (0.85, 0.91) |
| hMPV | |||||
| FilmArray | 0.914 (0.835, 0.956) | 0.999 (0.854, 1) | 74.5 (29.4, 188.98) | 0.167 (0.093, 0.301) | 0.98 (0.97, 0.99) |
AUC, area under the curve; hMPV, human metapneumovirus; LR−, negative likelihood ratio; LR+, positive likelihood ratio; RSV, respiratory syncytial virus.
Fig. 3Forest plot of sensitivity and specificity of multiplex PCRs for diagnosis of each virus.
Fig. 4Receiver operating curve analysis of multiplex PCRs for detection of different virus infections. Solid lines indicate the 95% confidence region, and the dashed lines indicate the 95% prediction region. AUC, area under the curve.