Literature DB >> 29204382

Effect of Material and Wall Thickness Buildup Caps on the Head Scatter Factor Measurements in Irregular Fields Shielded by Cerrobend.

Ali Jomehzadeh1, Daryoush Shahbazi-Gahrouei2, Vahid Jahanbakhsh2.   

Abstract

The head scatter factor (Sc ) is important to measurements radiation beam and beam modeling of treatment planning systems used for advanced radiation therapy techniques. This study aimed to investigate the design of a miniphantom to measurement variations in collimator Sc in the presence of shielding blocks for shaping the beam using different field sizes. Copper, Brass, and Perspex buildup caps were designed and fabricated locally as material with three different thicknesses for buildup caps (miniphantoms). Measurements were performed on an Elekta Compact medical linear accelerator (6 MV) in Shafa Kerman Hospital, Iran. The Farmer-type ion chamber FG65-P (Scanditronix, Wellhofer) was used for all measurements. To measure the Sc , miniphantom was positioned in a stand vertical to the beam central axis. The data indicate that the Sc measurements using different buildup cap materials and thicknesses in 5 × 10, 7.5 × 7.5, and two 10 × 10 cm Cerrobend shield blocks ranged 0.98 to 1.00, 1.04 to 1.05, and 1.04 to 1.06, respectively. Also, it was observed that by increasing the block shield area from 50 cm2 to both 56.25 and 100 cm2, the Sc increased in all situations. Results showed that using Brass compared to Perspex and Copper has less uncertainty due to its simple preparation and cutting which is useful to measurement of variations in collimator Sc and shaping the photon beam.

Entities:  

Keywords:  cerrobend block; linear accelerator; miniphantom; radiation therapy; scatter collimator factor

Year:  2017        PMID: 29204382      PMCID: PMC5691564     

Source DB:  PubMed          Journal:  J Med Signals Sens        ISSN: 2228-7477


Introduction

Radiation therapy (RT) uses high photon energy radiation beam which generated by medical linear accelerators (Linacs) is applied to destruction cancerous tissue.[123] The goal of RT is to deliver the prescribed dose to the planning target volume and simultaneously minimizing the unnecessary dose to the critical organs at risks (OARs).[345] To delivery tumor dose accurately and minimizing the dose to the OARs, new RT techniques, such as 3D-conformal RT, intensity modulated RT, and volumetric modulated arc therapy have been developed.[6] Each step in the integrated process of RT needs quality control and quality assurance to prevent errors and to give high confidence, those patients will receive the prescribed treatment correctly.[6] Modern RT treatment planning systems use advanced model-based algorithms to achieve a high degree of accuracy with motorized multileaf collimators (MLCs) which installed on the head of Linacs, and standard lead or Cerrobend shields for OARs. However, MLC is an optional on the head of Linacs and unavailable to all RT centers, particularly in Iran. Therefore, instead of using MLCs, Cerrobend shield blocks can be used for customized beam shaping which consists in a low melting point alloy.[7] Cerrobend is an alloy of 50% bismuth, 26.7% lead, 13.3% tin, and 10% cadmium by weight. Due to different OARs sizes, Cerrobend blocks can provide any shielding field sizes in radiotherapy.[8] The absorbed dose at a point within a patient/phantom can be divided into two components: primary radiation which is contributed by the primary or original photons emitted from the source and scattered photons which is scattered in the treatment head and phantom. Head scattered radiation is photon with a history of scattering from all structures in the treatment unit head called collimator scatter factor (S).[9] S describes the change of in-air output as a function of the collimator settings. According to the task group report No.74 from American Association of Physicists in Medicine, S is defined as the ratio of collision water kinetic energy released per unit mass in the free space of an arbitrary field to that of a reference field size (10 × 10 cm).[9] Scatter collimator factor depends on collimator design, scattering from the flattening filter, scattering from the dose monitoring chambers, beam-modifying devices, field sizes, thicknesses, and materials of buildup cap.[10] To measure the S, it is important to design the buildup cap or miniphantom, to use together with a suitable detector. The buildup cap thickness should provide full charged particle equilibrium (CPE) as in water medium, without any photon scattering, and be small enough to be fully covered by the radiation beam. The thickness of buildup cap should be equal to the depth of maximum dose (D), to provide CPE.[4] Several authors have been investigated the measurement of S, suitable materials for miniphantom design, and the buildup cap thicknesses.[39111213] Li and Zhu[3] showed that the in-air output ratio correction factors increase with miniphantom thickness for all the materials and for different energies. Iftikhar et al.[12] showed that the block tray factor increased with field size for both 6 and 15 MV photon beams. Farajollahi et al.[11] showed that there was no significant differences between lead and Cerrobend shielding methods for different field sizes in measuring S for Cobalt-60 therapy. Iftikhar[13] also found that the use of buildup caps with Zeff close to that of water such as acrylic is a good choice of output factors measurements. To the best of our knowledge, there is no study on assessment of buildup cap thicknesses and material for Cerrobend block fields using x-ray photon therapy. This study aimed to investigate the variation of collimator S in the presence of shielding blocks in the path of radiation beam using different wall thickness, material of buildup caps dependence.

Material and Methods

Photon beam, shield, and buildup cap design

The measurements were performed on an Elekta Compact linear accelerator (6 MV) at RT center in Shafa Kerman Hospital. Ionization chamber is the best choice for output factor measurements in extended water phantom for large field sizes,[14] which is reason why the Farmer type ionization was used here for all dose measurements. The Farmer-type ion chamber FC65-P (Scanditronix, Wellhofer) is a cylindrical ionization chamber with a fairly small sensitive volume of 0.65 cm3, outer diameter of 7 mm, inner diameter of 6.2 mm, and total active length of 23.1 mm with an inner electrode of aluminum, which used for all measurements.[14] Block shields were produced through melting of Cerrobend. It is normally fabricated in the molding room using the styrofoam block cutter system (PAR Scientific Model ACD-4MK4, Odense, Denmark).[11] This shielding block has been used to compare the effect of block shield sizes on the head Sc measurements [Figure 1].
Figure 1

Design of the block shield size for 6 MV photon beam. The dashed area illustrates the block shield region. (a) Shield (5 × 10 cm) was used in (10 × 10 cm) open field size, (b) shield (7.5 × 7.5 cm) was used in (15 × 15 cm) open field size, (c) two shield (5 × 10 cm) was used in (20 × 20 cm) open field size.

Design of the block shield size for 6 MV photon beam. The dashed area illustrates the block shield region. (a) Shield (5 × 10 cm) was used in (10 × 10 cm) open field size, (b) shield (7.5 × 7.5 cm) was used in (15 × 15 cm) open field size, (c) two shield (5 × 10 cm) was used in (20 × 20 cm) open field size. Copper, Brass and, Perspex buildup caps (miniphantom) were designed locally to have different material with three different wall thicknesses, to evaluate the effects of material and the wall thickness of buildup cap on S measurements. The Brass was designed with a wall thickness of 6 mm (equal to Dmax) and two wall thicknesses larger than Dmax (7 and 8 mm). The Copper buildup cap was designed with a wall thickness of 12 (equal of Dmax) and two wall thicknesses larger than Dmax (14 and 16 mm) and Perspex designed with a wall thickness of 14 (equal to Dmax) and two wall thicknesses larger than Dmax (17 and 19 mm) [Table 1].
Table 1

Material and wall thickness buildup caps for 6MV photon beam and Cerrobend block sizes

Material and wall thickness buildup caps for 6MV photon beam and Cerrobend block sizes

Head scatter factor measurement

To measure the head S, the designed miniphantoms were positioned in a stand vertical to the beam central axis and the measurements were performed using ion chamber for (10 × 10, 15 × 15, and 20 × 20 cm) field sizes to compare the effect of different field sizes on S measurements [Figure 2]. In this situation, the field size is fully covering the miniphantom. Table 1 shows the buildup caps material, the wall thicknesses, and block shield sizes.
Figure 2

The miniphantom and chamber position during scatter factor measurements.

The miniphantom and chamber position during scatter factor measurements. Measurements were carried out with perpendicular orientation of the chamber to the beam axis with a source-to-axis distance (SAD) 100 cm.[9] For each measurement, the SAD sets to source to central electrode of chamber which was different for each miniphantom. The measurements were also carried out for different block Cerrobend shield sizes to evaluate the effect of shield size on the S measurements. The field sizes were big enough to cover the buildup cap. Indeed, the chamber was not positioned behind the block shield to prevent the block attenuation effect. Each measurement was performed three times and then, the average and standard deviation values were calculated from the measurements. To obtain the head S, the average values were normalized to the reference open field (10 × 10 cm) readings.

Results

The measurements were performed in air, using Farmer chamber type placed in Perspex, Copper, and Brass buildup caps on a 6-MV photon energy Linac. The S was measured at isocenter for various open square fields as well as different block shielded as described in previous section. Table 2 shows the average reading of standard open and block shielded fields, S measurements with different buildup caps, and the S value obtained during commissioning of Linac.
Table 2

Head scatter factor for the block shield (5 × 10 cm) and deviation relative to open field (10 × 10 cm)

Head scatter factor for the block shield (5 × 10 cm) and deviation relative to open field (10 × 10 cm) Table 3 represents the measured S values for 15 × 15 open field size which shielded by 7.5 × 7.5 cm Cerrobend block shield as illustrated in Figure 1. The comparison of the head S with and without shield (commissioning data) presents a maximum and minimum deviation of 4.90 and 1.96% for Perspex 1 and Copper 2, Copper 3, Brass 2, respectively.
Table 3

Head scatter factor for the block shielded (7.5 × 7.5 cm) and deviation relative to open field (15 × 15 cm)

Head scatter factor for the block shielded (7.5 × 7.5 cm) and deviation relative to open field (15 × 15 cm) Table 4 shows the S measurements for field size (20 × 20 cm) and two block shielded field (5 × 10 cm). It is clear that among all buildup caps, Perspex 3 has the maximum deviation (4.9%).
Table 4

Head scatter factor for the two-shielded blocks (5 × 10 cm) and deviation relative to open field (20 × 20 cm)

Head scatter factor for the two-shielded blocks (5 × 10 cm) and deviation relative to open field (20 × 20 cm) Figure 3 shows a comprehensive comparison of S with buildup cap materials, wall thicknesses using different block shield in different open field sizes. It is obvious that by increasing the block shield area from 50 cm2 (in an open 10 × 10 cm field size) to both shield areas of 56.25 cm2 (in an open 15 × 15 cm field size) and 100 cm2 (in an open 20 × 20 cm field size), the measured S in all situations increased.
Figure 3

Variation of scatter factor measurements using different buildup cap materials, wall thicknesses, block shield areas, and open field sizes.

Variation of scatter factor measurements using different buildup cap materials, wall thicknesses, block shield areas, and open field sizes.

Discussion

In the most RT centers in Iran (except modern centers which have MLCs adjusted to the accelerator head), protection of normal tissues is usually accomplished by either Cerrobend or lead block shielding. Cerrobend block is the most common system for customized beam shaping due to its low melting point, high attenuation coefficient for photon beam and nontoxicity, and also reduction of bremsstrahlung rays by electron beam. Using materials with atomic number close to the tissue is recommended as a suitable material for designing Cerrobend. Interaction of radiation with matter causes its intensity reduced or attenuated exponentially which means that they do not have a fixed range in materials. The increase in thickness of the material resulted in photon intensity decrease. In this work, Perspex, Copper, and Brass with different thickness were used as buildup materials for attenuation photon intensity or shaping the photon beam. Findings of the present study showed that the value of Dmax for photon in water is about 1.5 cm; hence, the CPE conditions existed. Variations in the S factor (in-air output, collimator scatter, or head scatter) of buildup cap were observed with increases of the thickness of the used materials. As can be seen from Tables 2 to 4, increases in buildup cap thickness resulted in decreasing the amount of reading doses and are in good agreement with the exponentially law of attenuation (I = I0 e−μ). The atomic number of the materials used in buildup caps had been shown previously to have an effect on the measurement of head S.[1114] Table 2 demonstrated that maximum deviation of measured S in block shield situations relative to the commissioning data observed in Perspex 2 which was 2.04%. However, there is a same deviation (1.01%) using Copper 1, Copper 3, Brass 2, and Perspex 3, but no deviation was obtained for Copper 2, Brass 1, Brass 3, and Perspex 1. The S deviation using the same buildup cap material is more prominent between Perspex 1 and Perspex 2 by 2.04%. As Table 3 showed, the deviation of measured S relative to the S measured in commissioning process using Perspex buildup caps is prominent than Copper and Brass materials. While the deviation of S among Brass and Copper buildup caps is less than 1.00%, the deviating between Perspex 1 and Perspex 2/Perspex 3 is 1.90%. Table 4 illustrated that using the same buildup cap, the maximum deviation observed between Perspex 1/Perspex 2 (66.67%). Moreover, no deviation observed between Copper 2/Copper 3, Brass 1/Brass 3, and Perspex 2/Perspex 3. Figure 3 indicated that by increasing the thickness of buildup caps, in the case of two-block 5 × 10 and 7.5 × 7.5 cm, the S was decreased. In block shield (5 × 10 cm), Brass 2 had the minimum value and the shield (two blocks 5 × 10 cm) increasing with wall thicknesses and for the shield (7.5 × 7.5 cm), no variation was observed. In shield (5 × 10 cm), Perspex 2 had the minimum value. Moreover, other shield blocks decrease with thickness of buildup caps. The amount of variations in the S of fields with and without shield results had consistency with literature that as field size increases, the outer regions of the filter became exposed to the detector and buildup cap.[791114] Another key point finding here is that the ratio of area to surface of shields is an important factor. When a Cerrobend block is placed in the path way of an x-ray beam, scatter radiation will not be produced and hence, scatter radiation will be decreased. Increases in incident photon flow of less than 2% were observed due to the less scatter of a Cerrobend block. This amount of scatter depends on the length of the inner edge of the block and block size which is irradiated. Variations of the collimator setting could change the photon energy fluence, hence affecting the shape of photon beam. Results of the present work and other researchers have shown that only minor deviations in the resulting S values presented when the construction details of collimator were changed.[1114] Overall, use of melt shielding methods and Brass material is a very easy and fast shield-making technique to reduce variations of the S.

Conclusion

Results showed that using Brass compared to Perspex and Copper has less uncertainty due to its simple preparation and cutting. Also, its application as a Cerrobend block is recommended for measuring the variations of collimator S.

Financial support and sponsorship

This work was financially supported (Grant No: 395266) by Isfahan University of Medical Sciences.

Conflicts of interest

There are no conflicts of interest.
  8 in total

1.  Monte Carlo simulation of the effect of miniphantom on in-air output ratio.

Authors:  Jun Li; Timothy C Zhu
Journal:  Med Phys       Date:  2010-10       Impact factor: 4.071

2.  Measurement of head scatter factors of linear accelerators with columnar miniphantoms.

Authors:  Paul A Jursinic
Journal:  Med Phys       Date:  2006-06       Impact factor: 4.071

3.  Commissioning a miniature multileaf collimator for small field radiotherapy.

Authors:  David Eaton; Nicola Twyman; Simon Thomas
Journal:  Med Dosim       Date:  2009-01-30       Impact factor: 1.482

4.  Report of AAPM Therapy Physics Committee Task Group 74: in-air output ratio, Sc, for megavoltage photon beams.

Authors:  Timothy C Zhu; Anders Ahnesjö; Kwok Leung Lam; X Allen Li; Chang-Ming Charlie Ma; Jatinder R Palta; Michael B Sharpe; Bruce Thomadsen; Ramesh C Tailor
Journal:  Med Phys       Date:  2009-11       Impact factor: 4.071

Review 5.  Quality assurance of radiotherapy in cancer treatment: toward improvement of patient safety and quality of care.

Authors:  Satoshi Ishikura
Journal:  Jpn J Clin Oncol       Date:  2008-11       Impact factor: 3.019

6.  Comparison between Clinically Used Irregular Fields Shielded by Cerrobend and Standard Lead Blocks.

Authors:  A R Farajollahi; F Bouzarjomehri; M Kiani
Journal:  J Biomed Phys Eng       Date:  2015-06-01

7.  An Approach in Radiation Therapy Treatment Planning: A Fast, GPU-Based Monte Carlo Method.

Authors:  Mojtaba Karbalaee; Daryoush Shahbazi-Gahrouei; Mohammad B Tavakoli
Journal:  J Med Signals Sens       Date:  2017 Apr-Jun

8.  Comparison of Head Scatter Factor for 6MV and 10MV flattened (FB) and Unflattened (FFF) Photon Beam using indigenously Designed Columnar Mini Phantom.

Authors:  Sigamani Ashokkumar; N Arunai Nambi Raj; Sujit Nath Sinha; Girigesh Yadav; Rajesh Thiyagarajan; Kothanda Raman; Manindra Bhushan Mishra
Journal:  J Med Phys       Date:  2014-07
  8 in total
  2 in total

1.  Evaluation of Targeted Image-Guided Radiation Therapy Treatment Planning System by Use of American Association of Physicists in Medicine Task Group-119 Test Cases.

Authors:  Mohsen Saeb; Daryoush Shahbazi-Gahrouei; Shahram Monadi
Journal:  J Med Signals Sens       Date:  2018 Apr-Jun

2.  Interpretation of In-air Output Ratio of Wedged Fields in Different Measurement Conditions.

Authors:  Parinaz Mehnati; Farideh Biglari; Ali Jomehzadeh
Journal:  J Med Signals Sens       Date:  2019 Apr-Jun
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.