| Literature DB >> 29204382 |
Ali Jomehzadeh1, Daryoush Shahbazi-Gahrouei2, Vahid Jahanbakhsh2.
Abstract
The head scatter factor (Sc ) is important to measurements radiation beam and beam modeling of treatment planning systems used for advanced radiation therapy techniques. This study aimed to investigate the design of a miniphantom to measurement variations in collimator Sc in the presence of shielding blocks for shaping the beam using different field sizes. Copper, Brass, and Perspex buildup caps were designed and fabricated locally as material with three different thicknesses for buildup caps (miniphantoms). Measurements were performed on an Elekta Compact medical linear accelerator (6 MV) in Shafa Kerman Hospital, Iran. The Farmer-type ion chamber FG65-P (Scanditronix, Wellhofer) was used for all measurements. To measure the Sc , miniphantom was positioned in a stand vertical to the beam central axis. The data indicate that the Sc measurements using different buildup cap materials and thicknesses in 5 × 10, 7.5 × 7.5, and two 10 × 10 cm Cerrobend shield blocks ranged 0.98 to 1.00, 1.04 to 1.05, and 1.04 to 1.06, respectively. Also, it was observed that by increasing the block shield area from 50 cm2 to both 56.25 and 100 cm2, the Sc increased in all situations. Results showed that using Brass compared to Perspex and Copper has less uncertainty due to its simple preparation and cutting which is useful to measurement of variations in collimator Sc and shaping the photon beam.Entities:
Keywords: cerrobend block; linear accelerator; miniphantom; radiation therapy; scatter collimator factor
Year: 2017 PMID: 29204382 PMCID: PMC5691564
Source DB: PubMed Journal: J Med Signals Sens ISSN: 2228-7477
Figure 1Design of the block shield size for 6 MV photon beam. The dashed area illustrates the block shield region. (a) Shield (5 × 10 cm) was used in (10 × 10 cm) open field size, (b) shield (7.5 × 7.5 cm) was used in (15 × 15 cm) open field size, (c) two shield (5 × 10 cm) was used in (20 × 20 cm) open field size.
Material and wall thickness buildup caps for 6MV photon beam and Cerrobend block sizes
Figure 2The miniphantom and chamber position during scatter factor measurements.
Head scatter factor for the block shield (5 × 10 cm) and deviation relative to open field (10 × 10 cm)
Head scatter factor for the block shielded (7.5 × 7.5 cm) and deviation relative to open field (15 × 15 cm)
Head scatter factor for the two-shielded blocks (5 × 10 cm) and deviation relative to open field (20 × 20 cm)
Figure 3Variation of scatter factor measurements using different buildup cap materials, wall thicknesses, block shield areas, and open field sizes.