A R Farajollahi1, F Bouzarjomehri2, M Kiani3.
Abstract
INTRODUCTION: In radiation therapy centers across Iran, protection of normal tissues is usually accomplished by either Cerrobend or lead block shielding. In this study, the influence of these two shielding methods on central axis dose distribution of photon beam a Cobalt unit was investigated in clinical conditions.Entities:
Keywords: Cerrobend; Dosimetry; Lead block; Radiation therapy; Shield; TLD dosimetry
Year: 2015 PMID: 26157733 PMCID: PMC4479389
Source DB: PubMed Journal: J Biomed Phys Eng ISSN: 2251-7200
Mean ±SD Percent Depth Doses from 0.5 to 20cm Water Depths, for 23 Radiation Fields Shielded by Lead and Cerrobend Blocks
|
|
|
|
|
|---|---|---|---|
| 0.5 | 100±0.00 | 100±0.00 | - |
| 1 | 98.97±.68 | 99.12±.64 | 0.45 |
| 5 | 80.92±2.48 | 80.67±2.39 | 0.72 |
| 10 | 58.08±2.89 | 58.4±3.17 | 0.72 |
| 15 | 41.52±2.88 | 41.5±3.36 | 0.98 |
| 20 | 29.55±2.58 | 29.44±3.02 | 0.8 |
Figure 1PDD Curves of the Fields Shielded by Lead Block (red), Cerrobend Block (blue) and without Shield (green). These curves were plotted by Omni Pro software of the water phantom.
Mean ± SD of Absolute Doses at 1.5, 3, 5, 7, 10 and 12 cm Depths in Mini-Water Phantom for 23 Patient Field Sizes, SSD = 80 cm.
|
|
|
|
|
|---|---|---|---|
| 1.5 | 113.98±23.58 | 114.30±23.29 | 0.96 |
| 3 | 105.42±20.66 | 105.73±20.98 | 0.96 |
| 5 | 93.33±18.5 | 93.42±16.92 | 0.98 |
| 7 | 81.35±14.4 | 81.85±14.23 | 0.91 |
| 10 | 66.64±10.84 | 67.17±10.73 | 0.86 |
| 12 | 58.02±10.13 | 58±9.61 | 0.99 |
Mean ±SD of Doses at Surface, 0.5 and 1.5cm Solid Water Depths for 4 Patient Field Sizes Measured by TLD Dosimetery and Shielded by Lead and Cerrobend Blocks, SSD = 80 cm
|
|
|
|
|
|
|---|---|---|---|---|
| Surface | 111.96±7.69 | 109.13±6.31 | 0.26 | 2.53±1.63 |
| 0.5 | 115±7.07 | 114.08±6.26 | 0.16 | 2.05±0.86 |
| 1.5 | 108.91±6.67 | 107.76±6.88 | 0.63 | 1.09±.25 |