Magnetic resonance imaging metal artifact reduction for eye plaque patient with dental braces.

H Michael Gach1,2,3, Stacie L Mackey4, Sana Rehman1, Mo Kadbi5, Jacqueline E Zoberi1, Jose Garcia-Ramirez1, Perry W Grigsby1.   

Abstract

PURPOSE: To determine if metal artifact reduction can minimize magnetic susceptibility artifacts in the orbits for an eye plaque brachytherapy patient with metallic dental braces.
MATERIAL AND METHODS: A 62-year-old female patient with a choroidal melanoma in the right eye received a 1.5 T magnetic resonance imaging (MRI) simulation for 3D eye plaque brachytherapy planning. The protocol included conventional 3D T1-weighted and 2D T2-weighted MRIs. A vendor-supplied T2-weighted metal artifact reduction sequence was added to the protocol to reduce magnetic susceptibility artifacts from the metallic dental braces. The metal artifact reduction sequence combined turbo spin echo acquisitions, high RF excitation and readout bandwidths, and view angle tilting and slice encoding for metal artifact correction with z-shimming to correct in-plane and through-plane image distortions, respectively.
RESULTS: Dental braces caused significant signal loss and image distortion in the orbits on the conventional T1-weighted and T2-weighted MRIs, and the MRIs were unusable for treatment planning. The metal artifact reduction sequence with 13 z-phase encodes minimized distortion and signal loss in the orbits, allowing the tumor to be clearly delineated.
CONCLUSIONS: T2-weighted MRI with metal artifact reduction was successfully applied to minimize artifacts in the orbits resulting from the dental braces, thus allowing the MRIs to be used in 3D brachytherapy treatment planning.

Entities:  

Keywords:  SEMAC; dental braces; eye plaque; metal artifact reduction; orbits

Year:  2017        PMID: 29204171      PMCID: PMC5705839          DOI: 10.5114/jcb.2017.71184

Source DB:  PubMed          Journal:  J Contemp Brachytherapy        ISSN: 2081-2841


Purpose

Episcleral eye plaque brachytherapy is an effective treatment for ocular melanomas that avoids removal of the affected eye [1, 2]. The treatment involves suturing a radioactive plaque to the sclera adjacent to the tumor. The plaque remains in place until the desired dose is administered. Radionuclides may include gamma emitters 125I, 103Pd, 192Ir, and beta emitters 106Ru and 90Sr [3, 4, 5, 6]. Fundoscopy and ultrasonography are the current imaging standards for assessing the dimensions of intraocular melanomas for eye plaque brachytherapy [7]. 3D treatment planning based on computed tomography (CT) was demonstrated for episcleral plaque brachytherapy, but tumor delineation was unsatisfactory due to poor soft tissue contrast [8]. At our institution, eye plaque patients typically receive fundoscopy, ultrasound, and magnetic resonance imaging (MRI) for brachytherapy treatment planning [9]. MRI has superior soft tissue contrast to CT and is non-ionizing. T1-weighted (T1W) and T2-weighted (T2W) (MRI) were shown to provide excellent spatial delineation of the tumor [10, 11, 12]. The T1W MRIs have excellent contrast between the tumor (bright) and the vitreous humor (dark), but cannot distinguish between retinal detachment (if present) and the tumor. On T2W MRI, the tumor appears hypointense compared to the vitreous humor and retinal detachment (if present). Orthodontic appliances typically contain stainless steel and titanium that may produce severe magnetic susceptibility artifacts in the head and neck, and degrade the diagnostic quality of the MRI exam [13, 14]. Temporary removal of orthodontic devices is frequently justified to minimize susceptibility artifacts when diagnostic MRIs are required [15]. Obviously, it is far preferable to leave the orthodontic devices intact if the MRI can be successfully performed with minimal artifacts. This case report describes the use of a metal artifact reduction MRI pulse sequence to reduce magnetic susceptibility artifacts related to dental braces in an eye plaque brachytherapy patient.

Material and methods

A 62-year-old female patient presented with a slowly growing eye tumor consistent with a right choroidal melanoma. She was scheduled to receive definitive therapy with eye plaque brachytherapy and received an MRI simulation for treatment planning. The MRI was performed using a Philips Ingenia 1.5 Tesla MRI scanner (Philips Healthcare, The Netherlands). The patient was placed supine in a 15 channel phased array head coil. At isocenter, the patient was instructed to stare at a decal that is located on the top of the bore and to minimize blinking. The nominal MRI protocol for our eye plaque patients includes the following pulse sequences: 1) three-plane T1W scouts; 2) 2D T2W turbo spin echo (TSE) MRI; and 3) 3D T1W MRI. The pulse sequence parameters are summarized in Table 1. The structural sequences are acquired in the axial orientation with the phase encode in the right/left direction to prevent flow artifacts from appearing in the orbits. Intravenous MRI contrast is not used in these studies.
Table 1

Pulse sequence parameters

ParameterT1W Scout3D T1W2D T2W2D T2W with SEMAC*
TE/TR (ms)5/155/25120/7924100/4099
No. of slices5423341
Slice thickness/gap (mm)10/101.2/01.5/0.22/0
Flip angle20°30°90°90°
Acquired and reconstructedvoxel sizes (mm)2.38 x 1.000.94 x 0.941.22 x 1.180.94 x 0.941.17 x 0.980.44 x 0.441.22 x 0.990.97 x 0.97
Receiver BW (Hz/Pixel)187109139932
No. of averages1131
Acquisition time (s)18152293406

SEMAC used 13 z-phase encodes

SEMAC – slice encoding for metal artifact correction

Pulse sequence parameters SEMAC used 13 z-phase encodes SEMAC – slice encoding for metal artifact correction MRI safety screening identified that the patient had dental braces with metal and was claustrophobic. Dental braces are generally considered safe for nominal MRI field strengths. However, we limit the specific absorption rate to < 2.5 W/kg to minimize the risk of radiofrequency heating. The patient was coached to immediately use the squeeze bulb to alert the technologist in case of experiencing unusual heating or other adverse events. The braces produced magnetic susceptibility artifacts observed as dramatic signal loss in the anterior portion of the head on the T1W scout images and image geometric distortion in the orbit area on T2W. The patient was withdrawn from the bore with the intention to abort the exam. The MRI technologist contacted the MRI physicist and physician to determine the best course of action. In an effort to salvage the exam and obtain satisfactory MRI planning information for brachytherapy, the team decided to rescan the patient using Philips’ multislice 2D orthopedic metal artifact reduction (O-MAR) research sequence. The O-MAR sequence combines TSE acquisitions, high RF excitation, and readout bandwidths as well as view angle tilting (VAT) and slice encoding for metal artifact correction (SEMAC) with z-shimming to correct in-plane and through-plane image distortions, respectively [16, 17]. The SEMAC strength can be adjusted as weak, medium, and strong corresponding to specific frequency ranges per slice (z-phase encodes) of ± 5 kHz (7), ± 10 kHz (13), and ± 15 kHz (21), respectively. The acquisition time increases with the number of z-phase encodes, so it is advisable to use the weakest setting that corrects the artifact. We routinely use the weak SEMAC setting for cervical cancer patients with titanium tandems that would otherwise produce susceptibility artifacts with millimeter dimensions using conventional sequences [18]. The O-MAR sequence can currently be run with 4 contrast options: 1) T1W; 2) T2W; 3) proton density weighting (PDW); and 4) short TI inversion recovery (STIR). The Philips O-MAR product sequence (version 5.2 and later) is FDA approved. For this study, our Philips MRI used an earlier software version (version 5.1.7), and the O-MAR pulse sequence was run as a software patch. Our Institutional Review Board allowed us to use the O-MAR research sequence for clinical use without requiring informed consent, due to the clinical benefit/risk ratio and the similarity between the research and product sequences. The O-MAR software patch was installed by running a program that updates the pulse sequence. The patient was returned to isocenter, and an additional set of scouts and conventional 3D T1W images were acquired. We also acquired T2W O-MAR MRIs (Table 1). We selected the medium SEMAC setting (13 z-phase encodes) due to the severity of the artifact and in the interest of time. The patient was claustrophobic and wanted to end the exam, therefore no further sequences (e.g., T1W with O-MAR) or pulse sequence parameter options (e.g., increasing the receiver bandwidth) were completed. The MRIs were imported into an Eclipse Brachyvision treatment planning system (Varian Medical Systems, Palo Alto, CA, USA), and used to construct a 3D treatment plan as briefly described herein. The details of our MRI-based eye plaque brachytherapy methodology were recently published [9]. The tumor and critical structures (e.g., lens, macula, and optic nerve) were contoured. The treatment plan was designed to provide one fraction of 85 Gy delivered over 96 hours from Collaborative Ocular Melanoma Study (COMS) eye plaques loaded with 125I seeds. The plaque dosimetry used a TG-43 based homogenous dose calculation [19]. The seed groups were based on TG-129 seed center coordinates [20].

Results

The T1W MRIs without metal artifact reduction are shown in Figure 1. There is severe signal degradation in the orbits. The T2W MRIs with and without SEMAC are shown in Figure 2. O-MAR minimized the susceptibility artifact in the region of the tumor for the T2W MRIs (Figure 2). The dimensions of the eyes were 20-30% longer in the anterior/posterior (readout) direction for the T2W images without O-MAR versus with O-MAR. The specific absorption rate for the T2W sequences with and without O-MAR were 0.3 and 0.2 W/kg, respectively. The patient experienced no unusual heating during the MRI exam.
Fig. 1

Conventional T1W 2D scout (A, B) and 3D structural (C, D) magnetic resonance images (MRIs) showing signal loss in the anterior portion of the head from dental braces. A) Sagittal scout. B) Axial scout. C) Axial 3D T1W MRI of orbits in slice containing the tumor. D) Slice through maxillary sinuses 16 mm inferior of the tumor. Images were windowed to display the artifacts

Fig. 2

A) Conventional axial T2W magnetic resonance image (MRI) of the orbits showing the geometric distortion associated with magnetic susceptibility and motion artifacts. B) Axial T2W with O-MAR. The tumor is indicated by the arrow. The conventional T2W images appear grainier than the O-MAR images due to the former’s smaller reconstruction pixel size and slice thickness

Conventional T1W 2D scout (A, B) and 3D structural (C, D) magnetic resonance images (MRIs) showing signal loss in the anterior portion of the head from dental braces. A) Sagittal scout. B) Axial scout. C) Axial 3D T1W MRI of orbits in slice containing the tumor. D) Slice through maxillary sinuses 16 mm inferior of the tumor. Images were windowed to display the artifacts A) Conventional axial T2W magnetic resonance image (MRI) of the orbits showing the geometric distortion associated with magnetic susceptibility and motion artifacts. B) Axial T2W with O-MAR. The tumor is indicated by the arrow. The conventional T2W images appear grainier than the O-MAR images due to the former’s smaller reconstruction pixel size and slice thickness The MRIs with O-MAR were used to create a 3D model (Figure 3 and Table 2), which was necessary due to the irregular shape of the target volume and its proximity to critical structures (e.g., lens, optic disk, and fovea). On fundoscopy, the tumor was measured to be 7.9 x 8.7 mm with an apical height of 3.34 mm. On ultrasound, the tumor was measured to be 7.96 x 8.72 mm with an apical height of 3.29 mm. The tumor dimension was measured in MRI to be 7.9 x 9.1 x 4.1 mm with a maximal dimension of 9.4 mm. The posterior tumor edge to the fovea and disk were 6 and 4 mm, respectively. The prescription (RX) height was 4.3 mm.
Fig. 3

Magnetic resonance imaging-based 3D treatment plan for patient’s eye plaque brachytherapy using T2W O-MAR images

Table 2

Total dose based on 3D treatment plan and 96 hour exposure

Reference pointx [cm]*y [cm]*z [cm]*Total dose (cGy)
RX height, AP–3.43–0.038.218615.6
Tumor apex, AP–3.38–0.118.1910827.7
Max. basal, AP–3.19–0.378.1227580.3
Lens, AP–3.900.899.031098.9
Fovea, AP–3.690.327.372848.6
Optic disc, AP–3.400.187.474979.9

IEC 61217 coordinate system

Total dose based on 3D treatment plan and 96 hour exposure IEC 61217 coordinate system Magnetic resonance imaging-based 3D treatment plan for patient’s eye plaque brachytherapy using T2W O-MAR images The patient received a COMS 14 mm plaque (unnotched) containing 13 seeds with 3.84 U/seed (3.02 mCi/seed). The treatment time was 95 hours and 12 minutes.

Discussion

MRI metal artifact reduction was previously used to reduce the effects of metallic dental implants in MRI and PET/MRI [21, 22]. This is the first published report of metal artifact reduction to suppress MRI artifacts resulting from dental braces. Unfortunately, we were unable to evaluate the T1W O-MAR sequence or try other pulse sequence settings due to the inability of the patient to remain in the MRI because of her claustrophobia. Previously, there were few options for minimizing artifacts from dental braces. Increasing the receiver bandwidth (readout gradient amplitude) could significantly decrease the image distortion observed in the conventional T2W images near the orbits albeit at the cost of decreased signal-to-noise and contrast-to-noise ratios. However, increasing the receiver bandwidth on the conventional T1W MRIs would not have adequately corrected the artifacts and signal loss. In this patient, the medium SEMAC setting was satisfactory for imaging the orbits in the presence of dental braces. Assuming no signal averaging and an equivalent number of slices, the medium SEMAC T2W MRIs required a roughly three times longer acquisition time than the conventional T2W MRIs. Hence, the acquisition time was a major disadvantage of the O-MAR method. However, MRI acquisitions of seven minutes are routine for imaging the whole brain (e.g., 3D T1W MRIs). In addition, accelerated O-MAR acquisitions were recently demonstrated using compressed sensing to significantly reduce the acquisition time [23]. The strong SEMAC setting (with 21 z-phase encodes) may be useful for more severe artifacts or imaging volumes that are closer to the source of the artifact. However, the strong setting increases the acquisition time by ~ 60% relative to the medium SEMAC setting. Alternatively, O-MAR with the weak SEMAC setting may be useful for volumes of interest near metals that produce mild susceptibility artifacts e.g., the eye after implantation of the metal eye plaque. Without the metal artifact reduction, the MRIs would not have been satisfactory to generate a 3D treatment plan. Instead, the treatment plan would have relied on either a 2D model based on fundoscopy and ultrasound imaging [24] or a 3D plan based on CT images. Therefore, the advantages of O-MAR in facilitating an MRI-based 3D treatment plan for this patient outweighed the disadvantage of increased acquisition time.

Conclusions

In conclusion, O-MAR was demonstrated to provide excellent reduction of susceptibility artifacts and geometric distortion in T2W MRI of the orbits caused by dental braces. The MRIs with metal artifact reduction had sufficient quality to be used in the patient’s brachytherapy treatment planning whereas the conventional MRIs were unusable. Therefore, temporary removal of the dental braces was avoided by using O-MAR.

Disclosure

Mo Kadbi is an employee of Philips Healthcare, which supplied the O-MAR pulse sequence used in this case report. There are no other conflicts of interest.
  23 in total

1.  Superficial ocular malignancies treated with strontium-90 brachytherapy: long term outcomes.

Authors:  Siddhartha Laskar; Lavanya Gurram; Sarbani Ghosh Laskar; Suresh Chaudhari; Nehal Khanna; Rituraj Upreti
Journal:  J Contemp Brachytherapy       Date:  2015-10-16

2.  MR technique for localization and verification procedures in episcleral brachytherapy.

Authors:  P V Houdek; J G Schwade; A J Medina; C A Poole; K R Olsen; D H Nicholson; S Byrne; R Quencer; R S Hinks; V Pisciotta
Journal:  Int J Radiat Oncol Biol Phys       Date:  1989-11       Impact factor: 7.038

3.  Magnetic resonance imaging of the inferior alveolar nerve with special regard to metal artifact reduction.

Authors:  Monika Probst; Vivien Richter; Jochen Weitz; Jan Stefan Kirschke; Carl Ganter; Matthias Troeltzsch; Mathias Nittka; Carl-Peter Cornelius; Claus Zimmer; Florian Andreas Probst
Journal:  J Craniomaxillofac Surg       Date:  2017-01-25       Impact factor: 2.078

4.  MRI-based treatment planning and dose delivery verification for intraocular melanoma brachytherapy.

Authors:  Jacqueline Esthappan Zoberi; Jose Garcia-Ramirez; Samantha Hedrick; Vivian Rodriguez; Carol G Bertelsman; Stacie Mackey; Yanle Hu; H Michael Gach; P Kumar Rao; Perry W Grigsby
Journal:  Brachytherapy       Date:  2017-08-14       Impact factor: 2.362

5.  Potential hazards and artifacts of ferromagnetic and nonferromagnetic surgical and dental materials and devices in nuclear magnetic resonance imaging.

Authors:  P F New; B R Rosen; T J Brady; F S Buonanno; J P Kistler; C T Burt; W S Hinshaw; J H Newhouse; G M Pohost; J M Taveras
Journal:  Radiology       Date:  1983-04       Impact factor: 11.105

6.  SEMAC: Slice Encoding for Metal Artifact Correction in MRI.

Authors:  Wenmiao Lu; Kim Butts Pauly; Garry E Gold; John M Pauly; Brian A Hargreaves
Journal:  Magn Reson Med       Date:  2009-07       Impact factor: 4.668

7.  Quantifying the dosimetric influences of radiation coverage and brachytherapy implant placement uncertainty on eye plaque size selection.

Authors:  Nolan L Gagne; Mark J Rivard
Journal:  Brachytherapy       Date:  2013-03-01       Impact factor: 2.362

8.  Compressed Sensing SEMAC: 8-fold Accelerated High Resolution Metal Artifact Reduction MRI of Cobalt-Chromium Knee Arthroplasty Implants.

Authors:  Jan Fritz; Shivani Ahlawat; Shadpour Demehri; Gaurav K Thawait; Esther Raithel; Wesley D Gilson; Mathias Nittka
Journal:  Invest Radiol       Date:  2016-10       Impact factor: 6.016

9.  103Pd versus 125I ophthalmic plaque brachytherapy: preoperative comparative radiation dosimetry for 319 uveal melanomas.

Authors:  Paul T Finger; Di Zhou; Nina Kalach; Ekaterina Semenova; Walter Choi
Journal:  J Radiat Oncol       Date:  2014-04-09

10.  Metal artifact reduction in patients with dental implants using multispectral three-dimensional data acquisition for hybrid PET/MRI.

Authors:  Jeanne M Gunzinger; Gaspar Delso; Andreas Boss; Miguel Porto; Helen Davison; Gustav K von Schulthess; Martin Huellner; Paul Stolzmann; Patrick Veit-Haibach; Irene A Burger
Journal:  EJNMMI Phys       Date:  2014-12-20
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.