H Michael Gach1,2,3, Stacie L Mackey4, Sana Rehman1, Mo Kadbi5, Jacqueline E Zoberi1, Jose Garcia-Ramirez1, Perry W Grigsby1.
Abstract
PURPOSE: To determine if metal artifact reduction can minimize magnetic susceptibility artifacts in the orbits for an eye plaque brachytherapy patient with metallic dental braces.Entities:
Keywords: SEMAC; dental braces; eye plaque; metal artifact reduction; orbits
Year: 2017 PMID: 29204171 PMCID: PMC5705839 DOI: 10.5114/jcb.2017.71184
Source DB: PubMed Journal: J Contemp Brachytherapy ISSN: 2081-2841
Pulse sequence parameters
| Parameter | T1W Scout | 3D T1W | 2D T2W | 2D T2W with SEMAC |
|---|---|---|---|---|
| TE/TR (ms) | 5/15 | 5/25 | 120/7924 | 100/4099 |
| No. of slices | 5 | 42 | 33 | 41 |
| Slice thickness/gap (mm) | 10/10 | 1.2/0 | 1.5/0.2 | 2/0 |
| Flip angle | 20° | 30° | 90° | 90° |
| Acquired and reconstructed | 2.38 x 1.00 | 1.22 x 1.18 | 1.17 x 0.98 | 1.22 x 0.99 |
| Receiver BW (Hz/Pixel) | 187 | 109 | 139 | 932 |
| No. of averages | 1 | 1 | 3 | 1 |
| Acquisition time (s) | 18 | 152 | 293 | 406 |
SEMAC used 13 z-phase encodes
SEMAC – slice encoding for metal artifact correction
Fig. 1Conventional T1W 2D scout (A, B) and 3D structural (C, D) magnetic resonance images (MRIs) showing signal loss in the anterior portion of the head from dental braces. A) Sagittal scout. B) Axial scout. C) Axial 3D T1W MRI of orbits in slice containing the tumor. D) Slice through maxillary sinuses 16 mm inferior of the tumor. Images were windowed to display the artifacts
Fig. 2A) Conventional axial T2W magnetic resonance image (MRI) of the orbits showing the geometric distortion associated with magnetic susceptibility and motion artifacts. B) Axial T2W with O-MAR. The tumor is indicated by the arrow. The conventional T2W images appear grainier than the O-MAR images due to the former’s smaller reconstruction pixel size and slice thickness
Fig. 3Magnetic resonance imaging-based 3D treatment plan for patient’s eye plaque brachytherapy using T2W O-MAR images
Total dose based on 3D treatment plan and 96 hour exposure
| Reference point | x [cm] | y [cm] | z [cm] | Total dose (cGy) |
|---|---|---|---|---|
| RX height, AP | –3.43 | –0.03 | 8.21 | 8615.6 |
| Tumor apex, AP | –3.38 | –0.11 | 8.19 | 10827.7 |
| Max. basal, AP | –3.19 | –0.37 | 8.12 | 27580.3 |
| Lens, AP | –3.90 | 0.89 | 9.03 | 1098.9 |
| Fovea, AP | –3.69 | 0.32 | 7.37 | 2848.6 |
| Optic disc, AP | –3.40 | 0.18 | 7.47 | 4979.9 |
IEC 61217 coordinate system