| Literature DB >> 29187798 |
Gus Koerbin1,2, Jiakai Liu3, Alex Eigenstetter1, Chin Hon Tan3, Tony Badrick4, Tze Ping Loh5,6.
Abstract
INTRODUCTION: A product recall was issued for the Roche/Hitachi Cobas Gentamicin II assays on 25th May 2016 in Australia, after a 15 - 20% positive analytical shift was discovered. Laboratories were advised to employ the Thermo Fisher Gentamicin assay as an alternative. Following the reintroduction of the revised assay on 12th September 2016, a second reagent recall was made on 20th March 2017 after the discovery of a 20% negative analytical shift due to erroneous instrument adjustment factor.Entities:
Keywords: bias; error; moving average; quality assurance; quality control
Mesh:
Substances:
Year: 2017 PMID: 29187798 PMCID: PMC5701776 DOI: 10.11613/BM.2018.010705
Source DB: PubMed Journal: Biochem Med (Zagreb) ISSN: 1330-0962 Impact factor: 2.313
Sequence of events leading to the discovery of the analytical shift in the gentamicin assay at the laboratory reviewed
| 17th May 2016 | Roche advised of stock supply issues with GENT2 assay |
| 25th May 2016 | Roche advised of increased recovery of patient results and reagent recall. Advised use of alternate assay. |
| 31st May 2016 | Cessation of Roche assay |
| 1st Jun 2016 | Switched to Thermo Fisher Gentamicin reagent |
| 12th Sep 2016 | Roche advised GENT2 assay now available again but with 0.8 instrument adjustment factor due to ‘calibrator matrix issue’ |
| 20th Oct 2016 | Revised Roche assay with instrument adjustment factor of 0.8 reintroduced into clinical service |
| 20th Mar 2017 | Therapeutic Goods Administration (Australia) recalled Roche GENT2 assay due to decreased patient recovery (Reference RC-2017-RN-00374-1). Instrument adjustment factor of 0.8 recommended by Roche was inappropriately low and laboratory was advised to reset instrument adjustment factor to 1. |
| 24th Mar 2017 | Laboratory decided to reintroduce the Thermo Fisher Gentamicin reagent |
Figure 1Internal quality control (IQC) data of the index laboratory. The upper panel shows the IQC chart showing acceptable results up to the 1st Roche GENT2 reagent recall secondary to a 15 - 20% positive bias. The lower panel shows the IQC data during the use of the revised Roche assay with a 20% negative bias. The areas shaded light grey and grey represent 2 - 3 standard deviations and > 3 standard deviations, respectively.
Figure 2Analytical shift missed by the external quality assurance programme before the second reagent recall secondary to the presence of a negative bias, as the peer laboratories against which the index laboratory was compared were similarly using the affected reagents.
Figure 3Moving average control chart of gentamicin from 1st July 2015 to 5th April 2017 for the index laboratory (upper panel) and laboratory W (lower panel). The horizontal lines represent the upper and lower control limits, respectively.
Figure 4Moving sum of outlier control chart of gentamicin from 1st July 2015 to 5th April 2017 for the index laboratory (Panel A) and laboratory W (Panel B). A result of > 5.0 mg/L was considerd as an outlier. The horizontal lines represented the upper and lower control limits, respectively.