Verification of out-of-control situations detected by "average of normal" approach.

Jiakai Liu1, Chin Hon Tan1, Tze Ping Loh2, Tony Badrick3.   

Abstract

OBJECTIVES: "Average of normal" (AoN) or "moving average" is increasingly used as an adjunct quality control tool in laboratory practice. Little guidance exists on how to verify if an out-of-control situation in the AoN chart is due to a shift in analytical performance, or underlying patient characteristics. DESIGN AND METHODS: Through simulation based on clinical data, we examined 1) the location of the last apparently stable period in the AoN control chart after an analytical shift, and 2) an approach to verify if the observed shift is related to an analytical shift by repeat testing of archived patient samples from the stable period for 21 common analytes.
RESULTS: The number of blocks of results to look back for the stable period increased with the duration of the analytical shift, and was larger when smaller AoN block sizes were used. To verify an analytical shift, 3 archived samples from the analytically stable period should be retested. In particular, the process is deemed to have shifted if a difference of >2 analytical standard deviations (i.e. 1:2s rejection rule) between the original and retested results are observed in any of the 3 samples produced. The probability of Type-1 error (i.e., false rejection) and power (i.e., detecting true analytical shift) of this rule are <0.1 and >0.9, respectively.
CONCLUSIONS: The use of appropriately archived patient samples to verify an apparent analytical shift is preferred to quality control materials. Nonetheless, the above findings may also apply to quality control materials, barring matrix effects.
Copyright © 2016 The Canadian Society of Clinical Chemists. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Analytical; Error; Laboratory management; Quality control; Statistics

Mesh:

Year:  2016        PMID: 27452179     DOI: 10.1016/j.clinbiochem.2016.07.012

Source DB:  PubMed          Journal:  Clin Biochem        ISSN: 0009-9120            Impact factor:   3.281


  7 in total

1.  An Objective Approach to Deriving the Clinical Performance of Autoverification Limits.

Authors:  Tze Ping Loh; Rui Zhen Tan; Chun Yee Lim; Corey Markus
Journal:  Ann Lab Med       Date:  2022-09-01       Impact factor: 4.941

2.  Missed detection of significant positive and negative shifts in gentamicin assay: implications for routine laboratory quality practices.

Authors:  Gus Koerbin; Jiakai Liu; Alex Eigenstetter; Chin Hon Tan; Tony Badrick; Tze Ping Loh
Journal:  Biochem Med (Zagreb)       Date:  2017-11-24       Impact factor: 2.313

3.  Moving Rate of Positive Patient Results as a Quality Control Tool for High-Sensitivity Cardiac Troponin T Assays.

Authors:  Tingting Li; Shunwang Cao; Yi Wang; Yujuan Xiong; Yuting He; Peifeng Ke; Xianzhang Huang
Journal:  Ann Lab Med       Date:  2020-08-25       Impact factor: 3.464

4.  Optimization and validation of patient-based real-time quality control procedure using moving average and average of normals with multi-rules for TT3, TT4, FT3, FT3, and TSH on three analyzers.

Authors:  Chao Song; Jun Zhou; Jun Xia; Deli Ye; Qian Chen; Weixing Li
Journal:  J Clin Lab Anal       Date:  2020-05-03       Impact factor: 2.352

5.  Impact of combining data from multiple instruments on performance of patient-based real-time quality control.

Authors:  Qianqian Zhou; Tze Ping Loh; Tony Badrick; Chun Yee Lim
Journal:  Biochem Med (Zagreb)       Date:  2021-04-15       Impact factor: 2.313

6.  Intelligent Quality Management 2 with IntraSpect™ technology for quality control of GEM® Premier™ 5000 blood gas analyzers- A novel application of the patient sample as its own control.

Authors:  James O Westgard; Jose Cervera
Journal:  Pract Lab Med       Date:  2022-04-08

7.  Moving average procedures as an additional tool for real-time analytical quality control: challenges and opportunities of implementation in small-volume medical laboratories.

Authors:  Vera Lukić; Svetlana Ignjatović
Journal:  Biochem Med (Zagreb)       Date:  2021-12-15       Impact factor: 2.313

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.