| Literature DB >> 29152380 |
Christopher J Rowlands1, Demian Park2, Oliver T Bruns3, Kiryl D Piatkevich2, Dai Fukumura4, Rakesh K Jain4, Moungi G Bawendi3, Edward S Boyden1,2,5, Peter Tc So1,6.
Abstract
Three-photon wide-field depth-resolved excitation is used to overcome some of the limitations in conventional point-scanning two- and three-photon microscopy. Excitation of chromophores as diverse as channelrhodopsins and quantum dots is shown, and a penetration depth of more than 700 μm into fixed scattering brain tissue is achieved, approximately twice as deep as that achieved using two-photon wide-field excitation. Compatibility with live animal experiments is confirmed by imaging the cerebral vasculature of an anesthetized mouse; a complete focal stack was obtained without any evidence of photodamage. As an additional validation of the utility of wide-field three-photon excitation, functional excitation is demonstrated by performing three-photon optogenetic stimulation of cultured mouse hippocampal neurons expressing a channelrhodopsin; action potentials could reliably be excited without causing photodamage.Entities:
Keywords: biophotonics; multiphoton microscopy; optogenetics; temporal focusing; three-photon
Year: 2017 PMID: 29152380 PMCID: PMC5687557 DOI: 10.1038/lsa.2016.255
Source DB: PubMed Journal: Light Sci Appl ISSN: 2047-7538 Impact factor: 17.782
Figure 1Demonstration of wide-field three-photon excitation at 1300 nm. (a–c) Effect of increasing spot size. Sample consists of a drop of quantum dots on a glass microscope slide. Color scale is arbitrary, but all images share the same scale. Circles of diameter 50, 100, 150 and 200 μm are provided as a guide to the eye, and a yellow contour indicates where the intensity drops below 25% of peak. (a) × 2.5 beam expander, 0.7 s exposure. (b) × 3.5 beam expander, 1.5 s exposure. (c) × 5 beam expander, 10 s exposure. (d) Fixed HeLa cells stained with QDot 605 quantum dots. Power at the sample is 80 mW, integration time is 10 s, the scale bar is 100 μm and flat-field correction is applied to compensate for non-uniform excitation. (e) Excitation and emission spectra of the 525 nm quantum dots. Arrows at 650 and 433 nm indicate the location on the one-photon absorption spectrum corresponding to two- and three-photon excitation, respectively. (f) Fluorescence intensity as a function of average power, demonstrating three-photon excitation.
Figure 2Establishing the achievable penetration depth into fixed brain tissue. (a) Three-photon excitation at 1300 nm through fixed tissue slices from 200 to 800 μm, with a × 2.5 beam expander. Scale bars=25 μm. A projection along the Y axis is shown below each image. (b) Two-photon excitation at 800 nm through fixed tissue slices from 200 to 500 μm. Scale bars=25 μm. A projection along the Y axis is shown below each image. (c) Summarizes the data in a and b; it plots the intensity in the modulated spatial frequency as a fraction of the intensity over the whole image. (d) The MIT logo projected through 700 μm of fixed brain tissue onto a layer of quantum dots using a × 3.5 beam expander. The scale bar is 25 μm, and an illustration of the mask itself is inset. A discussion of the minor image distortion observable in this figure can be found in the Supplementary Information. (e) The optical layout for this experiment is shown.
Figure 3Three-photon optogenetic excitation at 1300 nm. (a) Representative examples of cultured neurons excited by three-photon excitation. The duration of exposure is indicated by the shaded area. (b) An example of the response to different excitation frequencies (2, 5, 10 and 20 Hz in ascending order). Exposure is indicated by the shaded area. (c) An illustration of the patched neuron with the three-photon excitation pattern superimposed. (d) Irradiance map of the spot, assuming 52 mW total integrated power at the sample. Units of irradiance are W mm−2. (e) Comparison of the 470 nm one-photon (n=19) excitation probability and the 1300 nm three-photon (n=9) excitation probability for cells expressing CoChR as a function of excitation frequency. Error bars indicate one standard deviation from the mean.