Literature DB >> 2913869

Clinical evaluation of a Raman scattering multiple gas analyzer for the operating room.

D R Westenskow1, K W Smith, D L Coleman, D E Gregonis, R A Van Wagenen.   

Abstract

A Raman spectrometer multiple gas analyzer was used to monitor inspired and expired concentrations of oxygen (O2), nitrogen (N2), carbon dioxide (CO2), nitrous oxide (N2O), halothane, and isoflurane in 10 patients. The Raman spectrometer and a dedicated mass spectrometer were connected to each patient to provide a comparison of the two instruments. Results show that readings from the Raman spectrometer are within 0.62 vol% of known gas standards for O2, N2, N2O; within 0.03 vol% for CO2; and within 0.04 vol% for halothane, enflurane and isoflurane. Clinical results show that Raman spectrometer readings are within 1.36 vol% of the mass spectrometer readings for O2, N2, N2O; within 0.01 vol% for CO2; and within 0.22 vol% for halothane and isoflurane. The clinical and laboratory results indicate the Raman spectrometer monitors airway gases and vapors as accurately as a dedicated mass spectrometer.

Entities:  

Mesh:

Year:  1989        PMID: 2913869     DOI: 10.1097/00000542-198902000-00028

Source DB:  PubMed          Journal:  Anesthesiology        ISSN: 0003-3022            Impact factor:   7.892


  8 in total

1.  Laboratory evaluation of the vital signs (ICOR) piezoelectric anesthetic agent analyzer.

Authors:  D R Westenskow; F H Silva
Journal:  J Clin Monit       Date:  1991-04

2.  Static and dynamic accuracy required of gas concentration measuring instruments.

Authors:  P C Beatty
Journal:  Med Biol Eng Comput       Date:  1991-03       Impact factor: 2.602

Review 3.  Basic principles of optical radiation and some common applications in anesthesia.

Authors:  D Gravenstein; S Lampotang; W Huda; A Sultan
Journal:  J Clin Monit       Date:  1996-11

4.  Can the Raman scattering analyzer compete with mass spectrometers: an affirmative reply.

Authors:  D R Westenskow; D L Coleman
Journal:  J Clin Monit       Date:  1989-01

5.  Accuracy and cross-sensitivity of 10 different anesthetic gas monitors.

Authors:  B Walder; R Lauber; A M Zbinden
Journal:  J Clin Monit       Date:  1993-11

6.  Stability and long-term durability of Raman spectroscopy.

Authors:  D Lawson; S Samanta; P T Magee; D E Gregonis
Journal:  J Clin Monit       Date:  1993-09

7.  Continuous end-tidal CO2 sampling within the proximal endotracheal tube estimates arterial CO2 tension in infants.

Authors:  G F Rich; J M Sconzo
Journal:  Can J Anaesth       Date:  1991-03       Impact factor: 5.063

8.  Accuracy of end-tidal carbon dioxide tension analyzers.

Authors:  D B Raemer; I Calalang
Journal:  J Clin Monit       Date:  1991-04
  8 in total