Leaflet fracture in Edwards-Duromedics bileaflet valves.

W Klepetko1, A Moritz, J Mlczoch, H Schurawitzki, E Domanig, E Wolner.   

Abstract

Two cases of leaflet fracture in the Edwards-Duromedics valve at 36 and 38 months after implantation are reported. Both patients were immediately reoperated on and recovered well. In one valve an older housing fracture with partial tissue ingrowth was noted beside a recent transverse leaflet fracture. In the other valve the leaflet was fractured near the pivot mechanism. All larger embolized parts were detected in the iliac artery region by computed tomographic scan and were subsequently removed. Problems in diagnosis and the importance of immediate reoperation, even without exact diagnosis, are discussed. Technical evaluation of the valve revealed crack growth and arrest, giving evidence of fatigue fracture. Scanning electron microscopic examination revealed several areas of pitting and erosion. Although the exact cause of mechanical disruption remains speculative, pyrolytic carbon seems to have the characteristic of fatigue fracture as well as erosion damage. A connection between the two might exist.

Entities:  

Mesh:

Year:  1989        PMID: 2911200

Source DB:  PubMed          Journal:  J Thorac Cardiovasc Surg        ISSN: 0022-5223            Impact factor:   5.209


  18 in total

1.  Role of vortices in cavitation formation in the flow at the closure of a bileaflet mitral mechanical heart valve.

Authors:  Chi-Pei Li; Sheng-Fu Chen; Chi-Wen Lo; Po-Chien Lu
Journal:  J Artif Organs       Date:  2011-10-21       Impact factor: 1.731

Review 2.  Analysis of prosthetic cardiac devices: a guide for the practising pathologist.

Authors:  J Butany; M J Collins
Journal:  J Clin Pathol       Date:  2005-02       Impact factor: 3.411

3.  A case of missing mitral valve leaflet: radiological detection.

Authors:  Altan Yildiz; Caner Ozer; Engin Kara; F Demir Apaydin; Meltem Nass Duce; Hulusi Eğilmez
Journal:  Eur Radiol       Date:  2003-02-18       Impact factor: 5.315

4.  Fracture embolization of a Duromedics mitral prosthesis.

Authors:  F J Baumgartner; A I Munro; W R Jamieson
Journal:  Tex Heart Inst J       Date:  1997

5.  Characteristics of cavitation intensity in a mechanical heart valve using a pulsatile device: synchronized analysis between visual images and pressure signals.

Authors:  Hwansung Lee; Eiki Akagawa; Eisuke Tatsumi; Yoshiyuki Taenaka
Journal:  J Artif Organs       Date:  2008-07-06       Impact factor: 1.731

6.  Observation of cavitation pits on a mechanical heart valve surface in an artificial heart used in in vivo testing.

Authors:  Hwansung Lee; Akihiko Homma; Eisuke Tatsumi; Yoshiyuki Taenaka
Journal:  J Artif Organs       Date:  2009-06-18       Impact factor: 1.731

7.  Wavelet analysis of valve closing sound detects malfunction of bileaflet mechanical valve.

Authors:  Hiroshi Sugiki; Toshifumi Murashita; Norihiko Shiiya; Yoshiro Matsui; Kenji Sugiki
Journal:  J Artif Organs       Date:  2008-04-15       Impact factor: 1.731

8.  Dynamics of a mechanical monoleaflet heart valve prosthesis in the closing phase: effect of squeeze film.

Authors:  C Gill-Jeong; K B Chandran
Journal:  Ann Biomed Eng       Date:  1995 Mar-Apr       Impact factor: 3.934

9.  Instantaneous back flow through peripheral clearance of Medtronic Hall tilting disc valve at the moment of closure.

Authors:  C S Lee; K B Chandran
Journal:  Ann Biomed Eng       Date:  1994 Jul-Aug       Impact factor: 3.934

10.  Prosthetic mitral valve leaflet escape.

Authors:  Darae Kim; Sin Sang Hun; In-Jeong Cho; Chi-Young Shim; Jong-Won Ha; Namsik Chung; Hyun Chul Ju; Jang Won Sohn; Geu-Ru Hong
Journal:  J Cardiovasc Ultrasound       Date:  2013-06-26
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.