Literature DB >> 29087037

Using weighted power mean for equivalent square estimation.

Sumin Zhou1, Qiuwen Wu2, Xiaobo Li3, Rongtao Ma1, Dandan Zheng1, Shuo Wang1, Mutian Zhang1, Sicong Li1, Yu Lei1, Qiyong Fan1, Megan Hyun1, Tyler Diener1, Charles Enke1.   

Abstract

PURPOSE: Equivalent Square (ES) enables the calculation of many radiation quantities for rectangular treatment fields, based only on measurements from square fields. While it is widely applied in radiotherapy, its accuracy, especially for extremely elongated fields, still leaves room for improvement. In this study, we introduce a novel explicit ES formula based on Weighted Power Mean (WPM) function and compare its performance with the Sterling formula and Vadash/Bjärngard's formula.
METHODS: The proposed WPM formula is ESWPMa,b=waα+1-wbα1/α for a rectangular photon field with sides a and b. The formula performance was evaluated by three methods: standard deviation of model fitting residual error, maximum relative model prediction error, and model's Akaike Information Criterion (AIC). Testing datasets included the ES table from British Journal of Radiology (BJR), photon output factors (Scp ) from the Varian TrueBeam Representative Beam Data (Med Phys. 2012;39:6981-7018), and published Scp data for Varian TrueBeam Edge (J Appl Clin Med Phys. 2015;16:125-148).
RESULTS: For the BJR dataset, the best-fit parameter value α = -1.25 achieved a 20% reduction in standard deviation in ES estimation residual error compared with the two established formulae. For the two Varian datasets, employing WPM reduced the maximum relative error from 3.5% (Sterling) or 2% (Vadash/Bjärngard) to 0.7% for open field sizes ranging from 3 cm to 40 cm, and the reduction was even more prominent for 1 cm field sizes on Edge (J Appl Clin Med Phys. 2015;16:125-148). The AIC value of the WPM formula was consistently lower than its counterparts from the traditional formulae on photon output factors, most prominent on very elongated small fields.
CONCLUSION: The WPM formula outperformed the traditional formulae on three testing datasets. With increasing utilization of very elongated, small rectangular fields in modern radiotherapy, improved photon output factor estimation is expected by adopting the WPM formula in treatment planning and secondary MU check.
© 2017 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine.

Entities:  

Keywords:  Akaike Information Criterion; equivalent square; linac output factors; weighted power mean

Mesh:

Year:  2017        PMID: 29087037      PMCID: PMC5689911          DOI: 10.1002/acm2.12201

Source DB:  PubMed          Journal:  J Appl Clin Med Phys        ISSN: 1526-9914            Impact factor:   2.102


INTRODUCTION

Equivalent Square (ES) is a widely used and important concept in photon external beam radiation dose calculation. ES postulates that, for an arbitrary rectangular field, there exists an equivalent square field sharing certain dosimetric characteristics. That concept provides us with a pathway for estimating a rectangular field's properties (e.g., central axis percentage depth dose, scatter factor) from measurements performed on square fields. The crucial step in the success of this approach is to identify the optimal formula that will predict the correct equivalent square. For a rectangular field with width a and length b, Sterling's formula1 was historically the first widely used, explicit ES formula for such a purpose. It remains the primary choice in current medical physics practice. Originally proposed in 1964 for studying the rectangular radiation field's central axis percentage depth dose that is generated by X‐ray units and 60Co machines, Sterling's formula has enjoyed success from that point to this, and now is almost a synonym for ES because of its simple mathematical structure and good prediction power for conventionally shaped and sized fields in many applications. The next major milestone in explicit ES formula history occurred about thirty years after the introduction of the Sterling's formula. To include the collimator exchange effect observed in their study of linear accelerator (linac) head‐scatter factors, Vadash and Bjärngard (VB) presented their modified version of Sterling's ES, 2, 3 specificallywith an adjustable parameter A > 0. Compared with Sterling's formula, this new form explicitly accounts for the collimator exchange effect.4 It can also reduce the maximum discrepancy in linac rectangular field head‐scatter factor prediction, down to about 1% for the resulting clinically relevant field sizes.2 The radiation oncology field has recently witnessed remarkable developments in the technology of radiotherapy delivery. Flattening‐filter‐free photon mode has become commonplace; very small and elongated photon fields are frequently used in both the IMRT and VMAT delivery processes. When dealing with these extreme situations, use of the previous ES formulae may lead to worrisome discrepancies. Thus, current radiation delivery modalities warrant an update to the explicit ES formula. Therefore, the aim of this study, using two well‐known datasets, was to propose a revised formula and demonstrate that it offers an improvement over the two most popular, conventional formulae.

METHODS

Herein, a Weighted Power Mean (WPM)5, 6 based formula was introduced, with the aim of achieving better accuracy than that obtained by the Sterling's and VB's formulae:with two adjustable parameters: power index and weighting factor . It is worth mentioning that our formula [see Eq. (3)] can be reduced to Sterling's formula (α ≡ −1 and ) [see Eq. (1)] or VB's formula (α ≡ −1 and ) [See Eq. (2)]. We included these two cases in our parameter selection, therefore the proposed formula is guaranteed to fit any dataset no worse than the two well‐known formulae would. By definition, all ES formulae should give a when a rectangular field degenerates to a square field (i.e. ). Our WPM‐based ES certainly satisfies this requirement. With any positive field size a, no matter what the values of α and w are: We tested this new WPM‐based ES formula on two publicly available datasets: (a) tables of equivalent square fields for central axis dose calculations in the British Journal of Radiology (BJR), supplement 257; and (b) in‐water output factor ratios (S ) from the Varian TrueBeam Representative Beam Data for Eclipse (Varian Medical Systems, Palo Alto, CA, USA).8 The latter provided the S measurements for linac jaw‐defined rectangular and square fields with field size between 3 cm and 40 cm in an SAD setup at SSD = 95 cm, with the ion chamber placed in the center of the photon field at 5 cm in‐water depth. We considered two methods for parameter fitting of the proposed formula: Performing a nonlinear least‐squares fitting to determine parameter values (i.e., w and α) based on known values for rectangular fields with different size combinations : Given a radiation property R and its measured values R(a,b), from a radiation generating device with rectangular/square field size combinations , w and α are determined through the following numerical procedure: since when a = b [See Eq. (4)], in the plot, all square field measurement points will be fixed points when we adjust the parameters w and α in the ES formula [See Eq. (3)]. As a result, we can set the constructed cubic‐spline (CS)9 based on the square field measurement points only (See solid lines in Fig. 1) as our target curve, and select the two parameters α and w values as those that minimize the sum of the squares of the difference between and the predicted value for all rectangular field measurements.
Figure 1

S as a function of estimated ES for various photon energies [(a) 6X FFF, (b) 15X, and (c) 10X FFF Edge]. Open‐field S values from the Varian representative dataset and the published Edge dataset were plotted as a function of ES predicted by three models. The solid lines in the plots are cubic splines with construction based only on square‐field measurements.

S as a function of estimated ES for various photon energies [(a) 6X FFF, (b) 15X, and (c) 10X FFF Edge]. Open‐field S values from the Varian representative dataset and the published Edge dataset were plotted as a function of ES predicted by three models. The solid lines in the plots are cubic splines with construction based only on square‐field measurements. With the BJR supplement 25 dataset, where the target value was given (as in Scenario 1 above), the standard deviation of the residual error after nonlinear least‐square fitting was used to judge the quality of the ES formula. For Scp in the Varian dataset (method 2 above) and the published measurement data from a Varian TrueBeam Edge unit,10 the quality of an ES formula was visualized in the plot. The reason for this was that if an ES formula correctly identified the underlying relationship between a rectangular field and its corresponding ES measured in Scp, then all data points, no matter whether they were from square field measurements or rectangular field measurements, should fall onto the same curve in a plot of . Clinically, the maximum absolute value in relative error between the predicted and the measured Scp value for an ES formula is most relevant to radiotherapy delivery (specifically MU calculation) and, therefore, can be used as a criterion for clinical model comparison. The Akaike Information Criterion (AIC)11 is a statistical tool for comparing different model performances on the same dataset wherein models are allowed to use different numbers of free parameters. For the ES formulae presented here, there are zero, one, and two free parameters in the Sterling, VB, and our WPM based formula respectively. The relative score of a model's performance can be expressed as:where k is the number of free parameters in the formula, and L is the maximum value of the likelihood function for the formula. Under the assumption that the residuals are distributed according to independent identical normal distributions (with zero mean), we have Here, n is the number of data points, is the residual sum of squares for formula with the optimal parameter set , and the constant term is model‐independent for a given dataset. We applied AIC as an objective comparison between our formula and the two established formulae. Please note that for the same dataset, only the relative value of AIC is meaningful; therefore, we will set the constant term to zero for the rest of this paper. For model comparison, the lower the AIC value, the better the model's performance.

RESULTS

When we applied the WPM‐based formula to the BJR ES table, due to the intrinsic symmetry between the two field sides of every rectangular field in the dataset, β must be . Next, we performed least squares fitting to obtain the optimal α value:where is the known ES for a rectangular field with sizes a and b. For the BJR dataset, the fitting procedure above led to , and the standard deviation of residual error was reduced by 20% compared with using α = −1 (i.e., the Sterling‐type ES formulae). The 99% confidence interval of α was . The optimal value for α was unlikely to be −1 for this dataset. Therefore, using the BJR dataset to support the use of Sterling's formula was based more upon clinical practicality rather than statistical analysis. The best fitting values of α and w for open field S in the Varian dataset for Eclipse are listed in Table 1 for some representative photon energies. Again, none of them selected α = −1.
Table 1

The best fitting values of parameters α and w for open field S of selected photon energies from the Varian dataset

Fitting parameterBeam energy/mode
6X15X6X FFF10X FFF
α−1.17−1.40−1.32−1.49
w0.560.590.530.56
The best fitting values of parameters α and w for open field S of selected photon energies from the Varian dataset All open‐field values of S from the Varian dataset were plotted against the predicted ES values from the three explicit ES formulae for four different photon energies (See Fig. 1). We can see that our WPM formula did a better job minimizing the spread of data points around the measured square‐field curve, indicating better modeling performance. The largest magnitude relative errors for open fields at all photon energies from the Varian dataset were graphed in Fig. 2 for all three ES formulae. We observed the same order of performance for all photon energies: WPM performed the best, followed by VB's, and then Sterling's.
Figure 2

Maximum absolute value in relative error in open‐field S prediction from the three models for various photon energies are listed in the Varian dataset and the reference.10

Maximum absolute value in relative error in open‐field S prediction from the three models for various photon energies are listed in the Varian dataset and the reference.10 The more rigorous comparison was performed based on the Akaike theory. The AIC value of the WPM formula was consistently lower than its counterparts from the traditional methods when we performed model fitting on the photon open field S tables in the Varian dataset (See Fig. 3). The lowest values among the three formulae indicated that the newly proposed WPM‐based ES formula outperformed both the Sterling and VB formulae, even after we took into account the number of adjustable model parameters.
Figure 3

Relative AIC values of the three ES formulae are computed based on some open field photon S measurements from the Varian dataset and the reference.10 In all cases, the model performed the best, and Sterling the worst.

Relative AIC values of the three ES formulae are computed based on some open field photon S measurements from the Varian dataset and the reference.10 In all cases, the model performed the best, and Sterling the worst. The worst relative data fitting errors for all three formulae occurred at data entries where photon fields were very elongated, narrow fields. The relative model fitting error for the fields with a shorter side at 1cm is tabulated in Table 2. We interpolated Varian 10X FFF Edge S values of nonsquare shaped fields based only on the square field measurements and ES predicted by the three formulae. The improvements of the WPM formula over the other two traditional formulae were obvious for these hard‐to‐fit field cases of large aspect ratios and very small short side lengths. From Table 2, we can see that the VB's formula has lower maximum relative error than the Sterling's formula while the WPM consistently outperformed the other two traditional ES formulae when it was applied to elongated small fields.
Table 2

Model prediction relative error for the three ES formulae when they are applied to S collected from 10X FFF fields with a short side 1 cm in length on a Varian TrueBeam Edge unit.10

10X FFF edge dataRelative prediction error in S cp (%)
Field long side (cm)SterlingVBWPM
X Jaw size = 1 cmY Jaw size = 1 cmX Jaw size = 1 cmY Jaw size = 1 cmX Jaw size = 1 cmY Jaw size = 1 cm
10.00.00.00.00.00.0
21.70.30.90.50.40.5
33.51.22.32.30.10.1
44.52.23.23.60.10.5
55.73.04.34.40.40.7
66.13.34.64.80.30.7
76.43.74.95.20.30.7
86.84.05.25.50.40.7
107.14.15.55.70.20.5
127.64.65.96.20.40.7
157.64.46.06.00.20.3
208.04.56.36.10.30.2
258.14.86.46.50.20.3
308.15.16.46.70.10.4
358.35.16.66.70.20.3
408.34.66.56.30.10.1
Model prediction relative error for the three ES formulae when they are applied to S collected from 10X FFF fields with a short side 1 cm in length on a Varian TrueBeam Edge unit.10 When other factors are fixed, the relative error in dose calculation for a photon field equates to the relative error in the employed output factor. Therefore, we expect our new ES formula will improve the accuracy of dose calculation, particularly when the photon field has an elongated small rectangular shape.

DISCUSSION

Both the Sterling formula and the Vadash and Bjärngard formula have been used clinically for many decades. They are both mathematically simple and, more importantly, provide good approximations of ES for rectangular clinical treatment fields when the field shape is not too far away from a square. Rapid technical advances in radiotherapy, especially the use of elongated small treatment fields, have advanced radiotherapy in many ways. This provided us with the motivation to propose a new ES estimation formula for the new generation of radiotherapy, similar to the case that the collimator exchange effect motivated the introduction of the VB formula thirty years ago. The proposal we presented here is a natural generalization of the two well‐established formulae. Comparing with the generic classical formulae, the new formula contains additional variables which could be optimized for best prediction accuracy based on LINAC type and beam energy. However, tests using the AIC criteria presented in this work unequivocally established the superiority of the new formula, especially in clinical scenarios enabled by modern technological revolutions. Introducing a mathematically more complex WPM formula with an additional fitting parameter (i.e., α) comes with a relatively modest price and is justifiably offset by the gain in reducing the systematical error in S prediction for small elongated rectangular fields such as those presented in Table 2. The success of our proposed formula on the two widely used datasets revealed its compelling clinical potential to alleviate the number of required commissioning measurements, while maintaining the quality of beam data for TPS modeling. It could also improve the accuracy of secondary MU check, which currently often suffers from unacceptable accuracy in calculating IMRT and VMAT plans. Further validation may be necessary for linear accelerators from other vendors.

CONCLUSIONS

A novel WPM formula has been proposed for ES estimation, which outperformed Sterling's explicit ES formula and its variant proposed by Vadash and Bjärngard on two well‐known public datasets. Both the weighting factor and the power index in the WPM formula can be determined through simultaneous optimization to achieve better accuracy. The improvement of the WPM over the Sterling‐type explicit ES formulae is particularly obvious for very elongated small rectangular fields that have been used with increasing frequency in IMRT and VMAT delivery. Improved dose calculation accuracy is expected when the WPM formula is adopted into treatment planning and secondary MU check systems.

ACKNOWLEDGMENTS

The authors would like to thank Varian Medical Systems for research grant support.

CONFLICTS OF INTEREST

This project was funded by Varian Medical Systems.
  7 in total

1.  AUTOMATION OF RADIATION TREATMENT PLANNING. IV. DERIVATION OF A MATHEMATICAL EXPRESSION FOR THE PER CENT DEPTH DOSE SURFACE OF COBALT 60 BEAMS AND VISUALISATION OF MULTIPLE FIELD DOSE DISTRIBUTIONS.

Authors:  T D STERLING; H PERRY; L KATZ
Journal:  Br J Radiol       Date:  1964-07       Impact factor: 3.039

2.  An equivalent square field formula for determining head scatter factors of rectangular fields.

Authors:  S Kim; T C Zhu; J R Palta
Journal:  Med Phys       Date:  1997-11       Impact factor: 4.071

Review 3.  The equivalent field method for dose determinations in rectangular fields.

Authors:  M J Day; E G Aird
Journal:  BJR Suppl       Date:  1996

4.  Commissioning and dosimetric characteristics of TrueBeam system: composite data of three TrueBeam machines.

Authors:  Zheng Chang; Qiuwen Wu; Justus Adamson; Lei Ren; James Bowsher; Hui Yan; Andrew Thomas; Fang-Fang Yin
Journal:  Med Phys       Date:  2012-11       Impact factor: 4.071

5.  Head-scatter factors in rectangular photon fields.

Authors:  M Tatcher; B E Bjarngard
Journal:  Med Phys       Date:  1993 Jan-Feb       Impact factor: 4.071

6.  An equivalent-square formula for head-scatter factors.

Authors:  P Vadash; B Bjärngard
Journal:  Med Phys       Date:  1993 May-Jun       Impact factor: 4.071

7.  Characteristics of a novel treatment system for linear accelerator-based stereotactic radiosurgery.

Authors:  Ning Wen; Haisen Li; Kwang Song; Karen Chin-Snyder; Yujiao Qin; Jinkoo Kim; Maria Bellon; Misbah Gulam; Stephen Gardner; Anthony Doemer; Suneetha Devpura; James Gordon; Indrin Chetty; Farzan Siddiqui; Munther Ajlouni; Robert Pompa; Zane Hammoud; Michael Simoff; Steven Kalkanis; Benjamin Movsas; M Salim Siddiqui
Journal:  J Appl Clin Med Phys       Date:  2015-07-08       Impact factor: 2.102

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.