Abstract
A simple formula is derived for the calculation of an equivalent square field that gives the same head scatter factor as a given rectangular field. This formula is based strictly on the configuration of a medical linear accelerator treatment head. The geometric parameters used are the distances between the target and the top of each field-defining aperture. The formula accounts for both the effect of field elongation and the collimator exchange effect. This method predicts the output to within 1% accuracy for both open and wedged fields and does not require any new measured data other than the field size dependence of head scatter for a range of square field sizes. Interestingly, the formula we derived has the same format as the formula that was empirically obtained by Vadash and Bjärngard [Med. Phys. 20, 733-734 (1993)].Mesh:
Year: 1997 PMID: 9394284 DOI: 10.1118/1.597963