Quantitative Effect of Pelvic Position on Radiographic Assessment of Acetabular Component Position.

Dimitri E Delagrammaticas1, Hasham M Alvi1, Aaron J Kaat2, Ryan R Sullivan1, Michael D Stover1, David W Manning1.   

Abstract

BACKGROUND: Intraoperative fluoroscopy aims to improve component position in total hip arthroplasty. Measurement bias related to image quality, however, has not been quantified. We aim to quantify measurement bias in the interpretation of acetabular component position as a function of pelvis and fluoroscopic beam position in a simulated supine total hip arthroplasty model.
METHODS: Posterior-anterior pelvis and hip images were obtained using a previously described pelvic model with known acetabular component position. Pelvic position was varied in 5° increments of pelvis rotation (iliac-obturator) and tilt (inlet-outlet), and in 1 cm increments from beam center in cranial-caudal and medial-lateral planes. Multiple regression analyses were conducted to evaluate the relationship between the resulting bias in interpretation of component position relative to pelvis position.
RESULTS: Anteversion and abduction measurement bias increased exponentially with increasing deviation in rotation and tilt. Greater bias occurred for anteversion than for abduction. Hip centered images were less affected by pelvis malposition than pelvis centered images. Deviations of beam center within 5 cm in the coronal plane did not introduce measurement bias greater than 5°. An arbitrarily defined acceptable bias of ±5° for both abduction and anteversion was used to identify a range of optimum pelvic positioning each for hip and pelvis centered imaging.
CONCLUSION: Accurate measurement of acetabular component abduction and anteversion, especially anteversion, is sensitive to proper pelvic position relative to the chosen radiographic plane. An acceptable measurement bias of ±5° is achieved when the pelvis is oriented within a newly identified range of optimum pelvic positioning.
Copyright © 2017 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  acetabular component position; anterior approach; fluoroscopy; pelvis position; radiographic interpretation

Mesh:

Year:  2017        PMID: 29066106     DOI: 10.1016/j.arth.2017.09.033

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  7 in total

1.  Accuracy of plain antero-posterior radiographic-based methods for measurement of acetabular cup version.

Authors:  Mohammad Ali Alzohiry; Mohammad Kamal Abdelnasser; Mohamed Moustafa; Mohamed Mahran; Hatem Bakr; Yaser Khalifa; Ahmed Abelaal; Haisam Atta; Galal Zaki Said
Journal:  Int Orthop       Date:  2018-06-04       Impact factor: 3.075

Review 2.  Radiographic assessment of the cup orientation after total hip arthroplasty: a literature review.

Authors:  Jing-Xin Zhao; Xiu-Yun Su; Zhe Zhao; Ruo-Xiu Xiao; Li-Cheng Zhang; Pei-Fu Tang
Journal:  Ann Transl Med       Date:  2020-02

3.  Three-dimensional orientation and location-dependent varying rules of radiographic angles of the acetabular cup.

Authors:  Jing-Xin Zhao; Xiu-Yun Su; Zhe Zhao; Ruo-Xiu Xiao; Li-Cheng Zhang; Pei-Fu Tang
Journal:  Int Orthop       Date:  2018-02-17       Impact factor: 3.075

4.  The synergetic effect of pelvic rotation and X-ray offset on radiographic angles of the acetabular cup.

Authors:  Jing-Xin Zhao; Xiu-Yun Su; Zhe Zhao; Ruo-Xiu Xiao; Guo-Fei Sun; Li-Cheng Zhang; Pei-Fu Tang
Journal:  Med Biol Eng Comput       Date:  2019-09-09       Impact factor: 2.602

5.  An Online Learning Tool to Obtain, Optimize, and Interpret Radiographs During Total Hip Arthroplasty.

Authors:  John P Livingstone; Makoa Mau; Jeffery K Harpstrite
Journal:  Hawaii J Health Soc Welf       Date:  2022-03

6.  Accuracy and Reliability of Software Navigation for Acetabular Component Placement in THA: An In Vitro Validation Study.

Authors:  Alex W Brady; Jakub Tatka; Lorenzo Fagotti; Bryson R Kemler; Bradley W Fossum
Journal:  Medicina (Kaunas)       Date:  2022-05-14       Impact factor: 2.948

7.  Letter to the Editor: The Effect of Postural Pelvic Dynamics on the Three-dimensional Orientation of the Acetabular Cup in THA Is Patient Specific.

Authors:  Jing-Xin Zhao; Xiu-Yun Su; Zhe Zhao; Li-Cheng Zhang; Pei-Fu Tang
Journal:  Clin Orthop Relat Res       Date:  2021-08-01       Impact factor: 4.755

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.