Literature DB >> 28976314

A Method for Measuring the Directional Response of Ultrasound Receivers in the Range 0.3-80 MHz Using a Laser-Generated Ultrasound Source.

James A Guggenheim, Edward Z Zhang, Paul C Beard.   

Abstract

A simple method for measuring the directivity of an ultrasound receiver is described. The method makes use of a custom-designed laser ultrasound source which generates a large diameter (>1 cm) broadband monopolar plane wave with a continuous frequency content extending from to . The plane wave is highly uniform in amplitude (±5% over >8 mm) and phase (equivalent to at 80 MHz over ). To measure directivity, the source is rotated around the receiver under test in a compact centimeter-scale setup. To demonstrate the method, it was used to measure the directivity of two broadband small aperture Fabry-Perot ultrasound sensors over an angular range of ±50° at frequencies up to 80 MHz. Measurements were found to be highly repeatable with an estimated typical repeatability <4% in the range of 0.5-25 MHz. Due to the broad bandwidth, large size, and adjustable nature of the source, the method is widely applicable and could aid the characterization of receivers used in medical ultrasound, ultrasonic nondestructive testing. and ultrasound metrology.

Year:  2017        PMID: 28976314     DOI: 10.1109/TUFFC.2017.2758173

Source DB:  PubMed          Journal:  IEEE Trans Ultrason Ferroelectr Freq Control        ISSN: 0885-3010            Impact factor:   2.725


  6 in total

1.  Considerations for Choosing Sensitive Element Size for Needle and Fiber-Optic Hydrophones-Part I: Spatiotemporal Transfer Function and Graphical Guide.

Authors:  Keith A Wear
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2018-12-10       Impact factor: 2.725

2.  Directivity and Frequency-Dependent Effective Sensitive Element Size of Membrane Hydrophones: Theory Versus Experiment.

Authors:  Keith A Wear; Christian Baker; Piero Miloro
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2019-07-24       Impact factor: 2.725

3.  Directivity and Frequency-Dependent Effective Sensitive Element Size of a Reflectance-Based Fiber-Optic Hydrophone: Predictions From Theoretical Models Compared With Measurements.

Authors:  Keith A Wear; Samuel M Howard
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2018-10-01       Impact factor: 2.725

4.  Directivity and Frequency-Dependent Effective Sensitive Element Size of Needle Hydrophones: Predictions From Four Theoretical Forms Compared With Measurements.

Authors:  Keith A Wear; Christian Baker; Piero Miloro
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2018-07-13       Impact factor: 2.725

5.  Large area laser scanning optical resolution photoacoustic microscopy using a fibre optic sensor.

Authors:  Thomas J Allen; Olumide Ogunlade; Edward Zhang; Paul C Beard
Journal:  Biomed Opt Express       Date:  2018-01-18       Impact factor: 3.732

6.  All-optical forward-viewing photoacoustic probe for high-resolution 3D endoscopy.

Authors:  Rehman Ansari; Edward Z Zhang; Adrien E Desjardins; Paul C Beard
Journal:  Light Sci Appl       Date:  2018-10-10       Impact factor: 17.782

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.