| Literature DB >> 28962603 |
Dachuan Liu1,2, Zhen Li2, Xuxiang Zhang2, Liping Zhao2, Jianping Jia3, Fei Sun4, Yaxing Wang5, Daqing Ma6, Wenbin Wei7.
Abstract
BACKGROUND: Ultrasonograpic retrobulbar optic nerve sheath diameter (ONSD) measurement is considered to be an alternative noninvasive method to estimate intracranial pressure,but the further validation is urgently needed. The aim of the current study was to investigate the association of the ultrasonographic ONSD and intracranial pressure (ICP) in patients.Entities:
Keywords: B-scan ultrasonography; Intracranial pressure; Optic nerve sheath diameter
Mesh:
Year: 2017 PMID: 28962603 PMCID: PMC5622417 DOI: 10.1186/s12883-017-0964-5
Source DB: PubMed Journal: BMC Neurol ISSN: 1471-2377 Impact factor: 2.474
Fig. 1Patient recruitment flow chart
Descriptive of participants’ information and main measurements
| minimum | maximum | Mean ± SD | |
|---|---|---|---|
| Age (year) | 12 | 68 | 38.3 ± 14.5 |
| BMI (kg/m2) | 17.7 | 35.5 | 24.9 ± 4.2 |
| ICP (mmH2O) | 50 | 450 | 215.3 ± 81.2 |
| ONSD-R(mm) | 3.87 | 8.38 | 5.70 ± 0.80 |
| ONSD-L (mm) | 3.46 | 7.55 | 5.80 ± 0.77 |
| ONSD-mean (mm) | 4.15 | 7.34 | 5.75 ± 0.71 |
ONSD-R: the optic nerve sheath diameter of the right eye measured with B-scan ultrasound;
ONSD-L: the optic nerve sheath diameter of the left eye measured with B-scan ultrasound;
ONSD-mean: the mean optic nerve sheath diameter of both eyes
Fig. 2The correlation of retrobulbar optic nerve sheath diameter measured in the right eye and in the left eye. Scatter plot of ONSD shows good correlation between two eyes (r = 0.656, P < 0.001)
Fig. 3The scatter plot of intracranial pressure (ICP) and the mean optic nerve sheath diameter (ONSD-mean) of both eyes and intracranial pressure and the body mass index. ICP was significantly associated with the mean ONSD (r = 0.61, P < 0.001) (a) and BMI (r = 0.55, P < 0.001) (b), not associated with age (r = −0.131, P = 0.174) (c) or gender (r = 0.03, P = 0.753) (d). In Fig. D,1 is male, 2 is female
Fig. 4Receiver operating characteristic (ROC) curve for the mean-ONSD. The area under the curve (AUC) of 0.861 means that 86.1% of the time, a randomly selected individual who with intracranial hypertension has a wider mean-ONSD than that of patients who with normal intracranial pressure
Classification results in the study group
| Group | Predicted Group Membership | Total | ||
|---|---|---|---|---|
| Normal ICP group | High ICP group | |||
| Original group | Normal ICP group | 45 (86.5%) | 7 (13.5%) | 52 |
| High ICP group | 9 (15.5%) | 49 (84.5%) | 58 | |
| Total | 54 | 56 | 110 | |
Classification results in validation group
| Group | Predicted Group | Total | ||
|---|---|---|---|---|
| Low ICP group | High ICP group | |||
| Original group | Normal ICP group | 12 (92.3%) | 1 (7.7%) | 13 |
| High ICP group | 1 (14.3%) | 6 (85.7%) | 7 | |
| Total | 13 | 7 | 20 | |
Fig. 5The scan images of ONSD measurement with B-scan ultrasound. a is the picture of ultrasonic measurement of ONSD in a high intracranial pressure patient. In this picture, there is a more significant subarachnoid space around optic nerve. b is the picture of ultrasonic measurement of ONSD in a normal intracranial pressure patient