Literature DB >> 28956513

The Medina Embolic Device: Karolinska experience.

P Bhogal1,2, P A Brouwer1, L Yeo1,3, M Svensson4, M Söderman1.   

Abstract

Background The aim of this study was to report our single centre experience with the Medina Embolic Device (MED). Methods We performed a retrospective analysis of prospectively collected data to identify all patients treated with the MED. A total of 14 aneurysms (non-consecutive), in 13 patients, were treated including one ruptured and one partially thrombosed aneurysm. Fundus diameter was ≥5 mm in all cases. We evaluated the angiographic appearances, the clinical status, complications, and the need for adjunctive devices or repeat treatments. Results Aneurysm location was cavernous internal carotid artery (ICA; n = 1), supraclinoid ICA ( n = 1), terminal ICA ( n = 2), anterior communicating artery (AComA; n = 4), A2-3 ( n = 1), M1-2 junction ( n = 1), posterior communicating artery (PComA; n = 1), superior cerebellar artery (SCA; n = 1), and basilar tip ( n = 2). The average aneurysm fundus size was 8.6 mm (range 7-10 mm) and average neck size 3.75 mm (range 1.9-6.9 mm). Immediate angiographic results were modified Raymond-Roy occlusion classification (mRRC) I n = 2, mRRC II n = 1, mRRC IIIa n = 2, mRRC IIIb n = 2, the remaining 7 aneurysms showed complete opacification. At follow-up angiography (mean 5 months) mRRC I n = 5, mRRC II n = 5, mRRC IIIa n = 3, and persistent filling was seen in 1 aneurysm. Overall, four patients had repeat treatment and one is pending further treatment. Of the aneurysms treated with more than one MED, 75% showed complete occlusion at 6-month follow up whereas only one aneurysm treated with a single device showed complete occlusion. Overall, three patients had temporary complications and there were no deaths. Conclusions The MED is an intra-saccular flow-diverting device with satisfactory angiographic results and an acceptable safety profile. Use of a single MED cannot be recommended and further longer term studies are needed prior to widespread clinical use.

Entities:  

Keywords:  Aneurysm; Medina Embolic Device; flow diverter; intra-saccular

Mesh:

Year:  2017        PMID: 28956513      PMCID: PMC5772534          DOI: 10.1177/1591019917733125

Source DB:  PubMed          Journal:  Interv Neuroradiol        ISSN: 1591-0199            Impact factor:   1.610


  11 in total

1.  Treatment of an unraveled intracerebral coil.

Authors:  Chaohua Wang; Xiaodong Xie
Journal:  Catheter Cardiovasc Interv       Date:  2010-11-01       Impact factor: 2.692

2.  Medina® Embolization Device for the Treatment of Intracranial Aneurysms: Safety and Angiographic Effectiveness at 6 Months.

Authors:  Nader-Antoine Sourour; Saskia Vande Perre; Federico Di Maria; Chrysanthi Papagiannaki; Joseph Gabrieli; Silvia Pistocchi; Bruno Bartolini; Vincent Degos; Alexandre Carpentier; Jacques Chiras; Frédéric Clarençon
Journal:  Neurosurgery       Date:  2018-02-01       Impact factor: 4.654

Review 3.  The Past, Present and Future of Endovascular Aneurysm Treatment.

Authors:  H Henkes; W Weber
Journal:  Clin Neuroradiol       Date:  2015-06-02       Impact factor: 3.649

4.  One-Year Angiographic Follow-Up after WEB-SL Endovascular Treatment of Wide-Neck Bifurcation Intracranial Aneurysms.

Authors:  R Sivan-Hoffmann; B Gory; R Riva; P-E Labeyrie; F Signorelli; I Eldesouky; U Gonike-Sadeh; X Armoiry; F Turjman
Journal:  AJNR Am J Neuroradiol       Date:  2015-08-20       Impact factor: 3.825

5.  Remnants and recurrences after the use of the WEB intrasaccular device in large-neck bifurcation aneurysms.

Authors:  Christophe Cognard; Anne Christine Januel
Journal:  Neurosurgery       Date:  2015-05       Impact factor: 4.654

6.  Relation between aneurysm volume, packing, and compaction in 145 cerebral aneurysms treated with coils.

Authors:  Menno Sluzewski; Willem Jan van Rooij; Marian J Slob; Javier Oliván Bescós; Cornelis H Slump; Douwe Wijnalda
Journal:  Radiology       Date:  2004-04-29       Impact factor: 11.105

7.  Cerebral aneurysm regrowth and coil unraveling after incomplete Guglielmi detachable coil embolization: serial angiographical and histological findings.

Authors:  Kentaro Mori; Yasuaki Nakao; Naoaki Horinaka; Ryo Wada; Akira Hirano; Minoru Maeda
Journal:  Neurol Med Chir (Tokyo)       Date:  2003-06       Impact factor: 1.742

8.  Characteristic features of unruptured intracranial aneurysms: predictive risk factors for aneurysm rupture.

Authors:  S-H You; D-S Kong; J-S Kim; P Jeon; K H Kim; H K Roh; G-M Kim; K-H Lee; S-C Hong
Journal:  J Neurol Neurosurg Psychiatry       Date:  2009-09-02       Impact factor: 10.154

9.  Periprocedural safety of aneurysm embolization with the Medina Coil System: the early human experience.

Authors:  Aquilla S Turk; Orlando Maia; Christian Candido Ferreira; Diogo Freitas; J Mocco; Ricardo Hanel
Journal:  J Neurointerv Surg       Date:  2015-01-27       Impact factor: 5.836

10.  The Medina Embolic Device: early clinical experience from a single center.

Authors:  Marta Aguilar Perez; Pervinder Bhogal; Rosa Martinez Moreno; Hansjörg Bäzner; Oliver Ganslandt; Hans Henkes
Journal:  J Neurointerv Surg       Date:  2016-08-02       Impact factor: 5.836

View more
  6 in total

1.  High frequency optical coherence tomography assessment of homogenous neck coverage by intrasaccular devices predicts successful aneurysm occlusion.

Authors:  Robert M King; Miklos Marosfoi; Jildaz Caroff; Giovanni J Ughi; Dale M Groth; Matthew J Gounis; Ajit S Puri
Journal:  J Neurointerv Surg       Date:  2019-04-29       Impact factor: 5.836

2.  The Combined Use of Intraluminal and Intrasaccular Flow Diversion for the Treatment of Intracranial Aneurysms: Report of 25 Cases.

Authors:  Pervinder Bhogal; Muhammad AlMatter; Victoria Hellstern; Oliver Ganslandt; Hansjörg Bäzner; Hans Henkes; Marta Aguilar-Pérez
Journal:  Neurointervention       Date:  2018-03-02

3.  Intra-aneurysmal flow disruption after implantation of the Medina® Embolization Device depends on aneurysm neck coverage.

Authors:  Andreas Maximilian Frölich; Marie Teresa Nawka; Marielle Ernst; Isabell Frischmuth; Jens Fiehler; Jan-Hendrik Buhk
Journal:  PLoS One       Date:  2018-02-06       Impact factor: 3.240

4.  Endosaccular Flow Disruption: A New Frontier in Endovascular Aneurysm Management.

Authors:  Adam A Dmytriw; Mohamed M Salem; Victor X D Yang; Timo Krings; Vitor M Pereira; Justin M Moore; Ajith J Thomas
Journal:  Neurosurgery       Date:  2020-02-01       Impact factor: 4.654

Review 5.  Endovascular Treatment of Intracranial Aneurysms.

Authors:  Antonis Adamou; Maria Alexandrou; Christian Roth; Achilles Chatziioannou; Panagiotis Papanagiotou
Journal:  Life (Basel)       Date:  2021-04-10

Review 6.  Treatment of unruptured middle cerebral artery aneurysms: Systematic review in an attempt to perform a network meta-analysis.

Authors:  Ignacio Arrese; Sergio García-García; Santiago Cepeda; Rosario Sarabia
Journal:  Front Surg       Date:  2022-09-28
  6 in total