Literature DB >> 28936890

Radiation dose reduction in plain radiography of the full-length lower extremity and full spine.

Mi Ran Jeon1, Hee Jin Park1, So Yeon Lee1, Kyung A Kang1, Eun Young Kim1, Hyun Pyo Hong1, Inyoung Youn1.   

Abstract

OBJECTIVE: To compare the diagnostic performance of standard- and low-dose radiographs of the full-length lower extremity and spine.
METHODS: This study included 223 patients who visited our hospital and received full-length lower extremity standing radiographs and full-spine radiographs. We determined the dose area product (DAP) of each image, and effective doses (ED, mSv) were calculated based on the DAP. Subjective evaluation of the full-length radiographs was based on image quality, which was assessed by bony cortex and trabecula evaluation, and on diagnostic performance, which was assessed by leg length measurement. Subjective evaluation of the full-spine radiographs was based on image quality, which was assessed by viewing the vertebral endplate, pedicle and lateral border of vertebral body, and on diagnostic performance from measurement of Cobb's angle.
RESULTS: For the full-length view and the full-spine view both the mean DAP and ED values of the standard-dose group were significantly higher than those of the low-dose group (p < 0.05). Mean scores for subjective values did not significantly differ based on the radiation dosage (p-values, 0.15-0.99). The subjective value scores for the full-length view were 2.94-2.98 in the standard-dose group and 2.91-3.00 in the low-dose group. Of note, both groups had very high scores. Additionally, the diagnostic performance scores between the two groups were also very high (range from 2.92 to 3.00).
CONCLUSION: Reducing mAs by 50% of the standard dose does not affect the radiograph image quality or its clinical validity. Advances in knowledge: Radiation dose reduction (50% of the standard dose of mAs) in plain radiography of the full-length lower extremity and full spine do not affect the clinical validity and the image quality.

Entities:  

Mesh:

Year:  2017        PMID: 28936890      PMCID: PMC6047650          DOI: 10.1259/bjr.20170483

Source DB:  PubMed          Journal:  Br J Radiol        ISSN: 0007-1285            Impact factor:   3.039


  15 in total

1.  The role of knee alignment in disease progression and functional decline in knee osteoarthritis.

Authors:  L Sharma; J Song; D T Felson; S Cahue; E Shamiyeh; D D Dunlop
Journal:  JAMA       Date:  2001-07-11       Impact factor: 56.272

2.  Understanding interobserver agreement: the kappa statistic.

Authors:  Anthony J Viera; Joanne M Garrett
Journal:  Fam Med       Date:  2005-05       Impact factor: 1.756

3.  Pelvic X-ray examinations in follow-up of hip arthroplasty or femoral osteosynthesis--dose reduction and quality criteria.

Authors:  Jost Karsten Kloth; Markus Rickert; Tobias Gotterbarm; Wolfram Stiller; Iris Burkholder; Hans-Ulrich Kauczor; Volker Ewerbeck; Marc-André Weber
Journal:  Eur J Radiol       Date:  2015-02-11       Impact factor: 3.528

4.  Computed radiographic measurement of limb-length discrepancy. Full-length standing anteroposterior radiograph compared with scanogram.

Authors:  Sanjeev Sabharwal; Caixia Zhao; John J McKeon; Emily McClemens; Michele Edgar; Fred Behrens
Journal:  J Bone Joint Surg Am       Date:  2006-10       Impact factor: 5.284

5.  Radiation dose reduction in scoliosis patients: low-dose full-spine radiography with digital flat panel detector and image stitching system.

Authors:  T Grieser; A Q Baldauf; K Ludwig
Journal:  Rofo       Date:  2011-05-25

6.  Roentgenographic assessment of the hip-knee-ankle axis in medial gonarthrosis. A study of reproducibility.

Authors:  S Odenbring; A M Berggren; L Peil
Journal:  Clin Orthop Relat Res       Date:  1993-04       Impact factor: 4.176

7.  Breast cancer mortality after diagnostic radiography: findings from the U.S. Scoliosis Cohort Study.

Authors:  M M Doody; J E Lonstein; M Stovall; D G Hacker; N Luckyanov; C E Land
Journal:  Spine (Phila Pa 1976)       Date:  2000-08-15       Impact factor: 3.468

8.  Quality-controlled dose reduction of full-leg radiography in patients with knee malalignment.

Authors:  Jost Karsten Kloth; Regina Neumann; Eva von Stillfried; Wolfram Stiller; Hans-Ulrich Kauczor; Volker Ewerbeck; Marc-André Weber
Journal:  Skeletal Radiol       Date:  2014-12-05       Impact factor: 2.199

9.  EOS low-dose radiography: a reliable and accurate upright assessment of lower-limb lengths.

Authors:  Benjamin G Escott; Bheeshma Ravi; Adam C Weathermon; Jay Acharya; Christopher L Gordon; Paul S Babyn; Simon P Kelley; Unni G Narayanan
Journal:  J Bone Joint Surg Am       Date:  2013-12-04       Impact factor: 5.284

10.  White paper on radiation protection by the European Society of Radiology.

Authors: 
Journal:  Insights Imaging       Date:  2011-06-03
View more
  2 in total

1.  Accuracy of biplanar linear radiography versus conventional radiographs when used for lower limb and implant measurements.

Authors:  Chen Xi Kasia Chua; Si Heng Sharon Tan; Andrew Kean Seng Lim; James Hoipo Hui
Journal:  Arch Orthop Trauma Surg       Date:  2021-01-02       Impact factor: 3.067

2.  Clinical utility of postprocessed low-dose radiographs in skeletal imaging.

Authors:  Johannes Kolck; Katharina Ziegeler; Thula Walter-Rittel; Kay Geert A Hermann; Bernd Hamm; Alexander Beck
Journal:  Br J Radiol       Date:  2022-01-05       Impact factor: 3.039

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.