| Literature DB >> 28932800 |
Stephan Fellner1, Simone Hentze1, Uwe Kempin1, Evelyn Richter1, Jörg Rocktäschel1, Barbara Langer1.
Abstract
a)Entities:
Keywords: B-type natriuretic peptide; BNP; Immunoassay; Point of care
Year: 2015 PMID: 28932800 PMCID: PMC5597718 DOI: 10.1016/j.plabm.2015.04.002
Source DB: PubMed Journal: Pract Lab Med ISSN: 2352-5517
Fig. 1(A) responsIQ instrument and visualization of detection unit and (B) top view of cartridge without the cover part showing the microfluidic design.
Video A1Schematic demonstration of signal generation process during a responsIQ measurement on the example of an assay for the determination of three analytes. A laser scans the detection area and the increase of the fluorescence signal over time is monitored.
Fig. 2Typical calibration curve comprising two sections with different slopes according to the low and high concentration range. N=3. Mean values; whiskers represent standard deviation.
Limits of detection (LOD) and within-run coefficient of variation (CV) from 10 LOD experiments measured over a period of about one year.
| Experiment no. | Date of measurement | Cartridge lot | Instrument no. | LOD [pg/mL] | Within-run CV at | ||
|---|---|---|---|---|---|---|---|
| +30 pg/ml BNP equivalents ( | +60 pg/ml BNP equivalents ( | +90 pg/ml BNP equivalents ( | |||||
| 1 | 28 Jan 2013 | A | 1 | 1.9 | 7.6% | 3.4% | 5.7% |
| 2 | 07 Feb 2013 | B | 1 | 1.9 | 3.9% | 2.7% | 4.4% |
| 3 | 22 Mar 2013 | C | 1 | 1.9 | 4.4% | 4.5% | 4.2% |
| 4 | 02 Mai 2013 | D | 2 | 2.5 | 2.6% | 11.5% | 3.4% |
| 5 | 17 Jun 2013 | E | 3 | 4 | 6.1% | 5.2% | 5.9% |
| 6 | 18 Jun 2013 | F | 3 | 4.3 | 5.0% | 2.3% | 6.5% |
| 7 | 03 Sep 2013 | G | 2 | 1.8 | 6.0% | 4.4% | 3.6% |
| 8 | 30 Oct 2013 | H | 4 | 1.6 | 4.0% | 5.8% | 4.8% |
| 9 | 13 Feb 2014 | I | 1 | 1.4 | 4.3% | 2.9% | 4.5% |
| 10 | 13 Feb 2014 | I | 5 | 1.8 | 3.7% | 3.1% | 1.9% |
Fig. 3Shelf life of responsIQ BNP cartridges stored at 4–7 °C. Bars represent mean values of BNP signals generated on cartridges by goat serum containing 1000 pg/ml BNP equivalents after storage as indicated. Whiskers represent standard deviation. N=6–8.
Fig. 4Passing & Bablok fit of responsIQ BNP values from whole blood (A) versus Alere Triage BNP also from whole blood; both values measured on-site (single measurement) directly after blood draw. N=100; and (B) versus ADVIA Centaur BNP values from plasma. responsIQ double determination in pes diagnosesysteme lab and Centaur BNP single measurement in external lab. n=98.