| Literature DB >> 28921910 |
Ellen Husby1, Elisabeth D Svendsen2, Hilde K Andersen1, Anne Catrine T Martinsen1,3.
Abstract
Quality control (QC) of CT scanners is important to evaluate image quality and radiation dose. Different QC phantoms for testing image quality parameters on CT are commercially available, and Catphan phantoms are widely used for this purpose. More data from measured image quality parameters on CT are necessary to assess test methods, tolerance levels, and test frequencies. The aim of this study was to evaluate the stability of essential image quality parameters for axial and helical scans on one CT scanner over time. A Catphan 600 phantom was scanned on a Philips Ingenuity CT scanner for 100 days over a period of 6 months. At each day of testing, one helical scan covering the entire phantom and four axial scans covering four different modules in the phantom were performed. All images were uploaded into Image Owl for automatic analysis of CT numbers, modular transfer function (MTF), low-contrast resolution, noise, and uniformity. In general, the different image quality parameters for both scan techniques were stable over time compared to given tolerance levels. Average measured CT numbers differed between axial and helical scans, while MTF was almost identical for helical and axial scans. Axial scans had better low-contrast resolution and less noise than helical scans. The uniformity was relatively similar for axial and helical scans. Most standard deviations of measured values were larger for helical scans compared to axial scans. Test results in this study were stable over time for both scan techniques, but further studies on different CT scanners are required to confirm that this also holds true for other systems.Entities:
Keywords: Catphan; computed tomography; image quality; phantom; quality control
Mesh:
Year: 2017 PMID: 28921910 PMCID: PMC5689914 DOI: 10.1002/acm2.12186
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Scan parameters used for helical and axial scans of Catphan 600 phantom
| Tube voltage [kV] | Rotation time [s] | Tube current [mA] | Pitch | Collimation | Slice width [mm] | Kernel | Display field of view (DFOV) | Matrix | CTDIvol [mGy] | |
|---|---|---|---|---|---|---|---|---|---|---|
| Helical scan | 120 | 0.75 | 324 | 0.797 | 64 × 0.625 | 3 | Standard (B) | 210 | 512 | 20 |
| Axial scans | 120 | 0.75 | 407 | ‐ | 64 × 0.625 | 3 | Standard (B) | 210 | 512 | 20 |
Figure 1CT images from left to right of (a) the sensitometry module (CTP404), (b) the high‐resolution module (CTP528), (c) the low‐contrast module (CTP515), and (d) the uniformity module (CTP486).
Average values, absolute value of maximum deviation from average values, range, and standard deviations of CT numbers
| Scan type | Air | PMP | LDPE | Polystyrene | Acrylic | Delrin | Teflon | |
|---|---|---|---|---|---|---|---|---|
| Average CT number [HU] | Axial | −1005 | −174 | −81 | −23 | 138 | 368 | 985 |
| Average CT number [HU] | Helical | −988 | −175 | −85 | −27 | 130 | 354 | 953 |
| Max deviation [|HU|] | Axial | 8 | 9 | 7 | 3 | 3 | 5 | 23 |
| Max deviation [|HU|] | Helical | 5 | 4 | 5 | 6 | 6 | 8 | 14 |
| Range [HU] | Axial | 11 | 12 | 9 | 5 | 6 | 9 | 33 |
| Range [HU] | Helical | 8 | 8 | 9 | 10 | 11 | 15 | 26 |
|
| Axial | 1.9 | 1.8 | 1.1 | 0.9 | 1.1 | 1.9 | 5.2 |
|
| Helical | 1.8 | 2.3 | 2.3 | 2.5 | 2.7 | 3.9 | 6.4 |
Figure 2Measured CT numbers for (a) air, (b) PMP, (c) LDPE, (d) polystyrene, (e) acrylic, (f) delrin, and (g) teflon for axial and helical scans.
Average values, absolute value of maximum deviation from average values, range, and standard deviation of critical frequency values
| Axial 50% of MTF | Helical 50% of MTF | Axial 10% of MTF | Helical 10% of MTF | |
|---|---|---|---|---|
| Average [cycles/cm] | 3.84 | 3.82 | 6.59 | 6.58 |
| Max deviation [|cycles/cm|] and (|%|) | 0.13 (3.4%) | 0.18 (4.7%) | 0.17 (2.6%) | 0.24 (3.7%) |
| Range [cycles/cm] | 0.26 | 0.31 | 0.26 | 0.33 |
| ( | 0.04 | 0.05 | 0.04 | 0.04 |
Figure 3Critical frequency values [cycles/cm] for 50% and 10% of MTF for axial and helical scans.
Average values, range, and standard deviation of estimated diameter [mm] of smallest detectable targets
| Axial 1% contrast | Helical 1% contrast | Axial 0.5% contrast | Helical 0.5% contrast | Axial 0.3% contrast | Helical 0.3% contrast | |
|---|---|---|---|---|---|---|
| Average [mm] | 2.0 | 2.0 | 3.0 | 3.9 | 5.0 | 5.9 |
| Range [mm] | 0 | 0 | 0 | 1.0 | 0 | 1.0 |
| ( | 0 | 0 | 0 | 0.36 | 0 | 0.27 |
Average values, absolute value of maximum deviation from average values, range, and standard deviation of measured noise
| Axial noise | Helical noise | |
|---|---|---|
| Average [HU] | 4.7 | 5.3 |
| Max deviation [|HU|] and (|%|) | 0.2 (4.1%) | 0.2 (3.3%) |
| Range [HU] | 0.3 | 0.3 |
| ( | 0.07 | 0.07 |
Figure 4Measured noise for axial and helical scans.
Average values, absolute value of maximum deviation from average values, highest value, range, and standard deviation for uniformity
| Axial uniformity deviation [HU] | Helical uniformity deviation [HU] | |
|---|---|---|
| Average [HU] | 1.4 | 1.3 |
| Max deviation [|HU|] | 1.1 | 1.3 |
| Highest value [HU] | 2.2 | 2.6 |
| Range [HU] | 1.9 | 2.2 |
| ( | 0.3 | 0.4 |
Figure 5Uniformity for axial and helical scans.