Simple method for RF pulse measurement using gradient reversal.

Vanessa L Landes1, Krishna S Nayak2.   

Abstract

PURPOSE: To develop and evaluate a simple method for measuring the envelope of small-tip radiofrequency (RF) excitation waveforms in MRI, without extra hardware or synchronization. THEORY AND METHODS: Gradient reversal approach to evaluate RF (GRATER) involves RF excitation with a constant gradient and reversal of that gradient during signal reception to acquire the time-reversed version of an RF envelope. An outer-volume suppression prepulse is used optionally to preselect a uniform volume. GRATER was evaluated in phantom and in vivo experiments. It was compared with the programmed waveform and the traditional pick-up coil method.
RESULTS: In uniform phantom experiments, pick-up coil, GRATER, and outer-volume suppression + GRATER matched the programmed waveforms to less than 2.1%, less than 6.1%, and less than 2.4% normalized root mean square error, respectively, for real RF pulses with flip angle less than or equal to 30°, time-bandwidth product 2 to 8, and two to five excitation bands. For flip angles greater than 30°, GRATER measurement error increased as predicted by Bloch simulation. Fat-water phantom and in vivo experiments with outer-volume suppression + GRATER demonstrated less than 6.4% normalized root mean square error.
CONCLUSIONS: The GRATER sequence measures small-tip RF envelopes without extra hardware or synchronization in just over two times the RF duration. The sequence may be useful in prescan calibration and for measurement and precompensation of RF amplifier nonlinearity. Magn Reson Med 79:2642-2651, 2018.
© 2017 International Society for Magnetic Resonance in Medicine. © 2017 International Society for Magnetic Resonance in Medicine.

Entities:  

Keywords:  RF fidelity; RF nonlinearity; predistortion correction

Mesh:

Year:  2017        PMID: 28905516      PMCID: PMC5821550          DOI: 10.1002/mrm.26920

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   4.668


  9 in total

1.  Method for improved multiband excitation profiles using the Shinnar-Le Roux transform.

Authors:  C H Cunningham; M L Wood
Journal:  Magn Reson Med       Date:  1999-09       Impact factor: 4.668

Review 2.  Issues of spectral quality in clinical 1H-magnetic resonance spectroscopy and a gallery of artifacts.

Authors:  Roland Kreis
Journal:  NMR Biomed       Date:  2004-10       Impact factor: 4.044

3.  Effects of RF amplifier distortion on selective excitation and their correction by prewarping.

Authors:  F Chan; J Pauly; A Macovski
Journal:  Magn Reson Med       Date:  1992-02       Impact factor: 4.668

4.  Reduced field of view MRI with rapid, B1-robust outer volume suppression.

Authors:  Travis B Smith; Krishna S Nayak
Journal:  Magn Reson Med       Date:  2011-11-14       Impact factor: 4.668

5.  The NMR multi-transmit phased array: a Cartesian feedback approach.

Authors:  D I Hoult; G Kolansky; D Kripiakevich; S B King
Journal:  J Magn Reson       Date:  2004-11       Impact factor: 2.229

6.  Parameter relations for the Shinnar-Le Roux selective excitation pulse design algorithm [NMR imaging].

Authors:  J Pauly; P Le Roux; D Nishimura; A Macovski
Journal:  IEEE Trans Med Imaging       Date:  1991       Impact factor: 10.048

7.  Blipped-controlled aliasing in parallel imaging for simultaneous multislice echo planar imaging with reduced g-factor penalty.

Authors:  Kawin Setsompop; Borjan A Gagoski; Jonathan R Polimeni; Thomas Witzel; Van J Wedeen; Lawrence L Wald
Journal:  Magn Reson Med       Date:  2011-08-19       Impact factor: 4.668

8.  Frequency-offset Cartesian feedback for MRI power amplifier linearization.

Authors:  Marta G Zanchi; Pascal Stang; Adam Kerr; John M Pauly; Greig C Scott
Journal:  IEEE Trans Med Imaging       Date:  2010-10-18       Impact factor: 10.048

9.  Probing the regional distribution of pulmonary gas exchange through single-breath gas- and dissolved-phase 129Xe MR imaging.

Authors:  S Sivaram Kaushik; Matthew S Freeman; Zackary I Cleveland; John Davies; Jane Stiles; Rohan S Virgincar; Scott H Robertson; Mu He; Kevin T Kelly; W Michael Foster; H Page McAdams; Bastiaan Driehuys
Journal:  J Appl Physiol (1985)       Date:  2013-07-11
  9 in total
  2 in total

1.  Iterative correction of RF envelope distortion with GRATER-measured waveforms.

Authors:  Vanessa L Landes; Krishna S Nayak
Journal:  Magn Reson Med       Date:  2019-08-23       Impact factor: 4.668

2.  Time optimal control-based RF pulse design under gradient imperfections.

Authors:  Christoph S Aigner; Armin Rund; Samy Abo Seada; Anthony N Price; Joseph V Hajnal; Shaihan J Malik; Karl Kunisch; Rudolf Stollberger
Journal:  Magn Reson Med       Date:  2019-08-23       Impact factor: 4.668

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.