Literature DB >> 28900359

Effectiveness of ATP bioluminescence to assess hospital cleaning: a review.

N Nante1, E Ceriale2, G Messina1, D Lenzi3, P Manzi3.   

Abstract

INTRODUCTION: Contamination of hospital surfaces plays an important role in the transmission of several healthcare-associated microorganisms, therefore methods for evaluating hospital surfaces' cleaning gain particular importance. Among these, there are visual inspection, quantitative microbiology, fluorescent markers and adenosine triphosphate (ATP) bioluminescence. The latter seems to provide interesting features, detecting the presence of ATP on surface (as Relative Light Units, RLU), a proxy of organic matter and microbial contamination. Several studies have investigated the effectiveness of this technology; with this research, we aim to summarize the most significant results.
METHODS: A systematic review was conducted. The keywords (namely, "ATP", "bioluminescence", "hospital" and "surfaces") were searched in PubMed/MEDLINE and Scopus databases, in order to find relevant data, from January 2000 to October 2014. After the selection, we globally considered 27 articles.
RESULTS: Most of the studies were conducted in United Kingdom and in USA. Different threshold RLU benchmark values were identified by analyzed studies. Fourteen of these researches compared the ATP bioluminescence with microbiological methods, 11 identified a significant correlation between the two methods, although poor or not complete for 5. DISCUSSION: ATP bioluminescence is not a standardized methodology: each tool has different benchmark values, not always clearly defined. At the moment, we can say that the technique could be used to assess, in real time, hospital surfaces where cleanliness is required, but not sterility.

Entities:  

Keywords:  ATP Bioluminescence; Healthcare-Associated Infections; Hospital Surfaces

Mesh:

Substances:

Year:  2017        PMID: 28900359      PMCID: PMC5584088     

Source DB:  PubMed          Journal:  J Prev Med Hyg        ISSN: 1121-2233


Introduction

Healthcare-associated infections (HAIs) represent an important and widespread cause of morbidity and mortality among patients. Over the past decades, various scientific evidences have accumulated, indicating that contamination of hospital surfaces plays an important role in the transmission and diffusion of several healthcareassociated microorganisms [1, 2]. In particular, the hospital environment contributes to the transmission of several nosocomial pathogens, such as Clostridium difficile, methicillin-resistant Staphylococcus aureus (MRSA) and vancomycin-resistant Enterococcus (VRE) [2, 3]. These bacteria could survive in this setting for a variable period, from hours to days and, in some cases, even months, and could contaminate the surfaces or the medical devices [4]. Consequently, pathogens could infect patients or contaminate the hands of healthcare staff and, then, patients [2]. Within this perspective, methods to assess hospital environments cleaning can be considered an integral part of infections prevention and control programs. Among these, the most known and used are: visual inspection, microbial methods, fluorescent markers and adenosine triphosphate (ATP) bioluminescence. The latter measures the presence of ATP on surfaces. The ATP bioluminescence consists in a swab, used to sample a standardized area, which, subsequently, is placed in a tool that uses the firefly enzyme "luciferase" to catalyze the conversion of ATP in Adenosine Monophosphate (AMP): this reaction results into an emission of light which is detected by the bioluminometer and quantified in Relative Light Unit (RLU). The presence of ATP on surface, obviously, is a proxy of organic matter and, consequently, of microbial contamination. This method has been used in food industries since over 30 years. Its use in the health care environment is growing, but it is still controversial, in that different tools consider different threshold values, and, therefore, this technique seems not to be standardized. Several studies investigated the effectiveness and also evaluated the practical application of this technique in this setting. The aim of this study was to qualitatively synthesize and discuss the most significant results and implications of the applications of ATP bioluminescence in healthcare settings, reviewing the most recent scholarly literature.

Methods

We conducted a systematic review according to the "Preferred Reporting Items for Systematic Reviews and Meta-analyses" (PRISMA) guidelines [5]. The search strategy consisted in a string of keywords such as ATP, bioluminescence, bioluminometer, hospital, healthcare setting, surfaces, connected by proper Boolean operators. For this scope, the keywords were searched in PubMed/MEDLINE and Scopus databases, in order to find relevant data. Time filter was applied and only articles from January 2000 to October 2014 were considered. Only articles written in English or for which an English translated text was available were included. Using Scopus we searched the keywords, selecting the option "all fields", whilst for PubMed/MEDLINE medical subject headings (MeSH) terms were used. All the searches as well as the screening were made by two researches EC and GM independently. Any disagreement was discussed until consensus was reached. We included all types of studies, except: i) not peerreviewed scholarly articles; ii) overview/review articles, which were excluded but scanned for including further potentially relevant studies, iii) articles not written in English language, iv) all laboratory studies, not conducted in health care settings, and v) articles lacking sufficient details or not pertinent with the aim of our review. Further, selected target journals were hand-searched for increasing the chance of getting relevant articles. For ensuring a high-quality of the included studies, we did not search in the grey literature. Table I reports our search strategy.
Tab. I.

Search strategy utilized in the current review.

strategy itemDetails
KeywordsAdenosine triphosphate, ATP, bioluminescence, bioluminometer, surfaces, hospital
DatabasesPubMed/MEDLINE, Scopus
Inclusion criteriaStudies investigating the applications and effectiveness of ATP bioluminescence Studies conducted in healthcare settings
Exclusion criteriaStudies not carried out in healthcare settings Studies lacking sufficient details Studies not pertinent with the aim of this review Study design: overview/review articles
Time filterJanuary 2000-October 2014
Language filterOnly articles written in English
Target journalsAmerican Journal of Infection Control; British Journal of Infection Control; Healthcare Infection; Infection control and hospital epidemiology: the official journal of the Society of Hospital Epidemiologists of America; International journal of hygiene and environmental health; Journal of Infection Control; Journal of Occupational and Environmental Hygiene
Search strategy utilized in the current review. General features of the studies. From each included study, we collected information about: i) the surname of the first author of the article and the year of publication; ii) the country where the research was carried out, iii) the setting in which the investigation was performed; iv) the samples used in the investigation; v) the RLU benchmark value discerning between clean and dirty surfaces (when investigated and reported), vi) the type of bioluminometers used, and vii) whether an eventual correlation between bioluminescence and microbiological methods existed (in terms of correlation coefficients, such as Pearson's coefficient, Spearman's rank coefficient, concordance κ coefficient or R2). The effect size of the correlation coefficient, when reported, was interpreted using the following rule of thumb: very high positive (or negative) if ranging from 0.90 to 1.00 (or ranging from -0.90 to -1.00): high positive (or negative) if ranging from 0.70 to 0.90 (or ranging from -0.70 to -0.90), moderate positive (or negative) if ranging from 0.50 to 0.70 (or ranging from -0.50 to -0.70), low/poor positive (or negative) if ranging from 0.30 to 0.50 (or ranging from -0.30 to -0.50), and little/no correlation if ranging from 0.00 to 0.30 (or ranging from 0.00 to -0.30). In the case the exact correlation coefficient was not indicated, we reported whether the correlation was statistically significant or not, on the basis of the p-value. Data extraction was carried out by two reviewers independently. In case of discrepancy, any disagreement was solved by discussion until consensus was reached or a third reviewer was involved.

Results

Using PubMed/MEDLINE we found 19 studies, 4 of which were judged not relevant for our investigation and one was a review, examining 12 papers, already found by our research. The final number of studies considered using PubMed/MEDLINE was 14. Therefore, in Scopus we found 315 papers, and only 27 were useful for the review. As such, the final number of studied included in the review was 27 (Fig. 1).
Fig. 1.

Preferred Reporting Items for Systematic Reviews and Meta-analyses (PRISMA) flowchart - Studies selection.

Preferred Reporting Items for Systematic Reviews and Meta-analyses (PRISMA) flowchart - Studies selection. Concerning the data extracted, all studies were conducted in health care settings. Most of them were carried out in United Kingdom (UK) (10/27) [6- 15]; 9 of them were performed in USA [16-24], 1 in Italy [25], Turkey [26], Japan [27], Chile [28], Canada [3], Norway [29], Australia [30] and Brazil [31]. Eleven researches monitored the surfaces after cleaning [3, 7, 9-11, 15, 19, 20, 25, 27, 31], one before [23], 13 both pre and post cleaning [6, 8, 12-14, 16-18, 22, 24, 28- 30] and in the remaining two studies this information was not reported [21, 26]. Twenty-three studies (85.2%) identified a RLU benchmark value, discerning between clean and dirty surfaces. This value corresponded to 250 for 10 studies [6, 10, 12, 13, 16-18 21, 23, 24], to 500 for 7 researches [8, 9, 11, 12, 14, 29, 31], to 100 for 4 studies [7, 15, 25, 29], to 300 for 2 studies [15, 19]; 127 [27], 1000 [3] and to 45 [20] for one study, respectively. Moore et al. [12] identified two threshold values: 250 and 500 RLU. Andersen et al. [29] used two different tools and, so, considered two different values: 500 RLU for ATP Biotrace Cleantrace system and 100 for Hygiena System. Another paper considered as cutoff 100 RLU for several patient rooms surfaces and 300 RLU for floors [15]. Details about RLU threshold values for each type of bioluminometers are shown in Figure 2. Figure 3 shows the benchmark values according to geographical provenance of studies.
Fig. 2.

The benchmark Relative Light Units (RLU) values according to the different brands of bioluminometers used in the studies included in the current review.

Fig. 3.

The benchmark Relative Light Units (RLU) values according to geographical provenance of studies.

The benchmark Relative Light Units (RLU) values according to the different brands of bioluminometers used in the studies included in the current review. The benchmark Relative Light Units (RLU) values according to geographical provenance of studies. Fourteen papers (51.8%) [8, 10, 11, 13, 15, 17, 18, 22-27, 31] compared the effectiveness of ATP bioluminescence to assess hospital surfaces' cleaning with microbiological methods; in particular, these studies evaluated the correlation between RLU and Aerobic Colony Counts (ACC). Three of these studies have shown no correlation between the two compared methodologies [18, 22 31], whilst the remaining 11 have highlighted a correlation [8, 10, 11, 13, 15, 17, 23-27], although it is poor/ moderate according to 4 studies [15, 17, 25, 27] and one found only a pre-cleaning correlation [24].

Discussion

Contamination of hospital surfaces plays an important role in the transmission of several healthcare-associated microorganisms. In this perspective, methods for evaluating hospital surfaces' cleaning gain importance. Each of these methods show advantages and disadvantages, and directives indicating the most appropriate method to use in different health care settings do not exist. The ATP bioluminescence seems to provide interesting perspectives, detecting the presence of ATP on surfaces (as Relative Light Units, RLU), a proxy of organic matter and microbial contamination. The present review showed the ATP bioluminescence is not a standardized method for assessing cleanliness; each tool had different benchmark values, ranging from 45 RLU to 1000 RLU. The most used values were 250 and 500 RLU. It is also interesting to note that for the same brand of bioluminometer different threshold values were considered. The country where the studies were conducted may have influenced the choice of the RLU cutoff values, for example in the U.S. the most often used value corresponded to 500 RLU. The tool used in most studies was 3M Clean-Trace ATP System, following by Hygiena system and Biotrace Cleantrace System. These differences among the benchmark values make difficult the comparisons between measurements carried out with different tools [15]. A large majority of studies was conducted in UK and in USA, probably because in these Countries there is a growing interest about environmental hospital hygiene and methods to assess it. Furthermore, it should be noted that among 14 studies investigating the correlation between ATP bioluminescence and microbiological methods, 11 have found a significant correlation, although poor, or partial, for 5 papers. ATP bioluminescence would not seem to be a methodology very accurate in detecting bacteria. In addition, it provides a quantification of all organic material, including bacteria, but it also identifies others organic matters such as urine, milk and blood, which is a limiting aspect of the methods and is rather difficult to overcome [10]. Another limitation of this technique could be the residues of detergent or disinfectants on the surfaces [32], which may require rinsing of these surfaces before the use [33]. Despite of these considerations, some advantages of this technique can be listed, such as the possibility to provide real-time results (within 20 seconds of sampling), its simplicity of use (which makes possible the adoption of the method not only by trained healthcare staff), and the quantitative results. The latter allows comparisons between pre- and post-cleaning or between different surfaces. Our study has a number of shortcoming that should be properly recognized. Despite the broad and systematic search, the main limitations of this study could be found in the selection of papers written only in English language and, therefore, in the possibility of having missed some relevant papers, which, considering the novelty of the technique, could be found, instead, in the "grey literature". On the other hand, it should be expected that the quality of the latter papers would be inferior to those present in the scholarly peer-reviewed indexed literature. Concerning the main implications of the current study, we can conclude that the use of this technique could produce better results after a proper validation/standardization. To achieve this aim a multi-phase approach should be followed: i) to clarify the methods of sanitizing or disinfecting performed on each surfaces before the controls made with the bioluminometer; ii) to select the most appropriate surfaces where the analysis have to be conducted (for example "high risk objects" such as toilet seats, basins, door handles); iii) to choose the bioluminometer among the different brands available; iv) to calibrate the instruments studying the correlation between CFU and RLU, identifying the best threshold value (higher sensitivity and specificity) [33] which discriminates between clean and dirt surfaces.

Conclusions

In conclusion, the ATP bioluminescence could be considered a practical, useful method to assess hospital hygiene, performing better than visual inspection (namely, Bacharach scale, bassoumeter, and glossmeter), if properly adopted, also being aware of its possible limits.
Tab. II.

General features of the studies.

Author/yearCountrySettingSampleRLU benchmarkCorrelation RLU/CFUBrand bioluminometer
Ali et al., 2012 [6]UKHospital 60 samples pre and 60 post cleaning250Not investigatedNot reported
Amin et al., 2014 [16]U.S.Clinic ophtalmology ward396 samples pre and 396 post cleaning250Not investigated3M Clean-Trace ATP System
Amodio et al., 2013 [25]ItalyTeaching hospital193 surfaces postcleaning100Poor (R2=0.29)Luminocontrol II (PBI internarnational, Milano)
Andersen et al., 2008 [29]Norway Teaching hospital96 samples pre and 96 post cleaning100, 500 Not investigatedATP Biotrace Clean-Trace system; Hygiena System
Anderson et al., 2011 [7]UKHospital surgical ward 44 samples post-cleaning100Not investigatedSystemSure Plus system (Hygiena Int. Ltd)
Aycicek et al., 2015 [26]TurkeyHospital kitchen280 from 14 surfaces Not reportedSignificant (κ = 0.249)Pd-10 kikkoman Co, Japan
Boyce et al., 2009 [17]U.S.Teaching hospital510 samples pre and 503 post cleaning250From poor to moderate (r from 0.356 to 0.649)3M Clean-Trace ATP System
Boyce et al., 2011 [18]U.S.Teaching hospital500 samples pre and 500 post cleaning250Not significant3M Clean-Trace ATP System
Branch-Elliman et al., 2014 [19]U.S.Hospital820 samples post clearing300Not investigated3M Clean-Trace NG Luminometer
Cooper et al., 2007 [8]UK4 hospital 552 samples pre and 547 post cleaning 500SignificantBiotrace Cleantrace system
Ferreira et al., 2011 [31]BrazilHospital100 samples post clearing500Not significant3M Clean-Trace ATP System
Gillespie et al., 2012 [30]AustraliaHospital50 samples pre and 50 post cleaningNot reportedNot investigatedNot reported
Gold et al., 2013 [20]U.S.Intensive Care Unit (ICU)Postcleaning. Number of surfaces not specified45Not investigatedRuholf ATO Complete contamination Monitoring System
Gordon et al., 2014[3]CanadaTeaching hospital 15 HTOs in 36 patients rooms (first day) and in 37 patients room (second day) 1000Not investigatedNot reported
Griffith et al., 2008 [9]UKHospital 31 sites postcleaning500Not reportedBiotrace Clean-Trace system
Havill et al., 2011 [21]U.S.Teaching hospital 300 samples from 101 rolling blood pressure unit250Not investigated3M Clean-Trace ATP System
Lewis et al., 2008 [10]UKTeaching hospital180 samples post cleaning250SignificantBiotrace International, Ltd, Brigend, UK
Luick et al., 2013 [22]U.S.Teaching hospital250 surfaces pre and post cleaningNot reportedNot significantAccupoint HC (Neogen)
Malik et al., 2003 [11]UK4 hospitalsnon specified the samples number. Sampling done postcleaning500Significant Biotrace Cleantrace system
Mulvey et al., 2011 [13]UKTeaching hospital90 samples pre and post cleaning 250 SignificantHygiena system
Moore et al., 2010 [12]UKTeaching hospital90 samples pre and 90 post cleaning250, 500Not investigated3M Clean-Trace Clinical Hygiene Monitoring System
Sherlock et al., 2009 [14]UKTwo hospital wards120 sample pre and 120 post cleaning500Not investigated3M International Ltd, Brigend, UK
Smith et al., 2013 [23]U.S.Hospital10 samples in 10 rooms pre and post-cleaning250Significant only precleaning3M Clean-Trace ATP System
Smith et al., 2013 [24]U.S.Hospital 18 samples pre-cleaning in 10 rooms250Significant3M Clean-Trace ATP System
Watanabe et al., 2014 [27]JapanThree hospital of different sizes752 samples post cleaning127Poor (r = 0.287)3M Clean-Trace ATP System
Willis et al., 2007 [15]UKHospital108 samples post cleaning100(surfaces), 300 (floor) Poor (r = 0.15)Hygiena system
Zambrano et al., 2014 [28]ChileTeaching hospital198 samples pre and post cleaningNot reportedNot investigatedLightning MVPTM (Arquimed)
  28 in total

1.  Is it really clean? An evaluation of the efficacy of four methods for determining hospital cleanliness.

Authors:  O Sherlock; N O'Connell; E Creamer; H Humphreys
Journal:  J Hosp Infect       Date:  2009-03-24       Impact factor: 3.926

2.  Variations in hospital daily cleaning practices.

Authors:  John M Boyce; Nancy L Havill; Abigail Lipka; Heather Havill; Ramo Rizvani
Journal:  Infect Control Hosp Epidemiol       Date:  2010-01       Impact factor: 3.254

3.  Cleanliness audit of clinical surfaces and equipment: who cleans what?

Authors:  R E Anderson; V Young; M Stewart; C Robertson; S J Dancer
Journal:  J Hosp Infect       Date:  2011-04-16       Impact factor: 3.926

4.  Environment cleaning without chemicals in clinical settings.

Authors:  Elizabeth Gillespie; Jill Wilson; Anita Lovegrove; Carmel Scott; Mardi Abernethy; Despina Kotsanas; Rhonda L Stuart
Journal:  Am J Infect Control       Date:  2012-11-22       Impact factor: 2.918

5.  Diagnostic assessment of different environmental cleaning monitoring methods.

Authors:  Laura Luick; Paul A Thompson; Mark H Loock; Sheree L Vetter; Joan Cook; Dubert M Guerrero
Journal:  Am J Infect Control       Date:  2013-02-04       Impact factor: 2.918

6.  Effect of surface coating and finish upon the cleanability of bed rails and the spread of Staphylococcus aureus.

Authors:  S Ali; G Moore; A P R Wilson
Journal:  J Hosp Infect       Date:  2012-01-20       Impact factor: 3.926

7.  Assessing the effectiveness of surface cleaning methods in intravitreal injection procedure rooms.

Authors:  Sejal R Amin; Carla M Folkert; Jay C Erie
Journal:  Ophthalmology       Date:  2013-10-07       Impact factor: 12.079

8.  Direct feedback with the ATP luminometer as a process improvement tool for terminal cleaning of patient rooms.

Authors:  Westyn Branch-Elliman; Ernie Robillard; Gary McCarthy; Kalpana Gupta
Journal:  Am J Infect Control       Date:  2014-02       Impact factor: 2.918

9.  The use of adenosine triphosphate bioluminescence to assess the efficacy of a modified cleaning program implemented within an intensive care setting.

Authors:  Ginny Moore; Debbie Smyth; Julie Singleton; Peter Wilson
Journal:  Am J Infect Control       Date:  2010-10       Impact factor: 2.918

10.  Monitoring the effectiveness of hospital cleaning practices by use of an adenosine triphosphate bioluminescence assay.

Authors:  John M Boyce; Nancy L Havill; Diane G Dumigan; Michael Golebiewski; Ola Balogun; Ramo Rizvani
Journal:  Infect Control Hosp Epidemiol       Date:  2009-07       Impact factor: 3.254

View more
  14 in total

1.  Surface contamination in the operating room: use of adenosine triphosphate monitoring.

Authors:  Alex Ramirez; Sanjay Mohan; Rebecca Miller; Dmitry Tumin; Joshua C Uffman; Joseph D Tobias
Journal:  J Anesth       Date:  2018-12-11       Impact factor: 2.078

2.  Hygienic monitoring in long-term care facilities using ATP, crAssphage, and human noroviruses to direct environmental surface cleaning.

Authors:  Jennifer L Cannon; Geun Woo Park; Benjamin Anderson; Cortney Leone; Morgan Chao; Jan Vinjé; Angela M Fraser
Journal:  Am J Infect Control       Date:  2022-03       Impact factor: 2.918

3.  Factors Shaping Attitudes of Medical Staff towards Acceptance of the Standard Precautions.

Authors:  Jerzy Rosiński; Anna Różańska; Andrzej Jarynowski; Jadwiga Wójkowska-Mach; Polish Society Of Hospital Infections Team
Journal:  Int J Environ Res Public Health       Date:  2019-03-23       Impact factor: 3.390

4.  Evaluation of the total adenylate (ATP + ADP + AMP) test for cleaning verification in healthcare settings.

Authors:  M Bakke; S Suzuki; E Kirihara; S Mikami
Journal:  J Prev Med Hyg       Date:  2019-06-28

5.  Monitoring environmental contamination caused by SARS-CoV-2 in a healthcare facility by using adenosine triphosphate testing.

Authors:  Yu-Mi Lee; Dong Youn Kim; Ki-Ho Park; Mi Suk Lee; Young Jin Kim
Journal:  Am J Infect Control       Date:  2020-07-08       Impact factor: 2.918

6.  ATP bioluminescence assay for evaluating cleaning practices in operating theatres: applicability and limitations.

Authors:  Tiziana Sanna; Laura Dallolio; Alessandra Raggi; Magda Mazzetti; Giovanni Lorusso; Angela Zanni; Patrizia Farruggia; Erica Leoni
Journal:  BMC Infect Dis       Date:  2018-11-19       Impact factor: 3.090

7.  Hospital-Wide Surveillance of Healthcare-Associated Infections as a Source of Information about Specific Hospital Needs. A 5-Year Observation in a Multiprofile Provincial Hospital in the South of Poland.

Authors:  Małgorzata Kołpa; Marta Wałaszek; Anna Różańska; Zdzisław Wolak; Jadwiga Wójkowska-Mach
Journal:  Int J Environ Res Public Health       Date:  2018-09-07       Impact factor: 3.390

8.  Comparison between Two Types of Dental Unit Waterlines: How Evaluation of Microbiological Contamination Can Support Risk Containment.

Authors:  Jessica Lizzadro; Marta Mazzotta; Luna Girolamini; Ada Dormi; Tiziana Pellati; Sandra Cristino
Journal:  Int J Environ Res Public Health       Date:  2019-01-24       Impact factor: 3.390

9.  Evaluation of different detector types in measurement of ATP bioluminescence compared to colony counting method for measuring bacterial burden of hospital surfaces.

Authors:  Huiqiong Xu; Jiansheng Liang; Yimei Wang; Bin Wang; Tianbao Zhang; Xiaoli Liu; Lin Gong
Journal:  PLoS One       Date:  2019-09-06       Impact factor: 3.240

10.  ATP measurement as an objective method to measure environmental contamination in 9 hospitals in the Dutch/Belgian border area.

Authors:  Andreas van Arkel; Ina Willemsen; Linda Kilsdonk-Bode; Sindy Vlamings-Wagenaars; Anne van Oudheusden; Pascal De Waegemaeker; Isabel Leroux-Roels; Martine Verelst; Evelien Maas; Anita van Oosten; Patricia Willemse; Esther van Asselen; Ella Klomp-Berens; Karen Franssen; Elise Van Cauwenberg; Jan Kluytmans
Journal:  Antimicrob Resist Infect Control       Date:  2020-05-28       Impact factor: 4.887

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.