Literature DB >> 28900356

Humidifiers for oxygen therapy: what risk for reusable and disposable devices?

V La Fauci1, G B Costa1, A Facciolà2, A Conti2, R Riso2, R Squeri1.   

Abstract

INTRODUCTION: Nosocomial pneumonia accounts for the vast majority of healthcare-associated infections (HAI). Although numerous medical devices have been discussed as potential vehicles for microorganisms, very little is known about the role played by oxygen humidifiers as potential sources of nosocomial pathogens. The purpose of this research was to evaluate the safety of the reuse of humidifiers by analysing the rate of microbial contamination in reusable and disposable oxygen humidifiers used during therapy, and then discuss their potential role in the transmission of respiratory pathogens.
METHODS: Water samples from reusable and disposable oxygen humidifiers were collected from different wards of the University Hospital of Messina, Italy, where nosocomial pneumonia has a higher incidence rate due to the "critical" clinical conditions of inpatients. In particular, we monitored the Internal Medicine and Pulmonology wards for the medical area; the General Surgery and Thoracic and Cardiovascular Surgery wards for the surgical area and the Intensive Care Unit and Neonatal Intensive Care Unit for the emergency area. The samples were always collected after a period of 5 days from initial use for both types of humidifiers. Samples were processed using standard bacteriological techniques and microbial colonies were identified using manual and automated methods.
RESULTS: High rates of microbial contamination were observed in samples from reusable oxygen humidifiers employed in medical (83%), surgical (77%) and emergency (50%) areas. The most relevant pathogens were Pseudomonas aeruginosa, amongst the Gram-negative bacteria, and Staphylococcus aureus, amongst the Gram-positive bacteria. Other pathogens were detected in lower percentage. The disposable oxygen humidifier samples showed no contamination.
CONCLUSIONS: This research presents evidence of the high rate and type of microbial contamination of reusable humidifiers employed for oxygen therapy. These devices may thus be involved in the transmission of potential pathogens. It could be important, for the prevention of nosocomial pneumonia, to replace them with singleuse humidifiers for which the absence of microbial contamination has been confirmed.

Entities:  

Keywords:  Microbiological contamination; Nosocomial pneumonia; Oxygen humidifiers

Mesh:

Year:  2017        PMID: 28900356      PMCID: PMC5584085     

Source DB:  PubMed          Journal:  J Prev Med Hyg        ISSN: 1121-2233


  20 in total

1.  Association between implementation of CDC recommendations and ventilator-associated pneumonia at selected US hospitals.

Authors:  L P Manangan; S N Banerjee; W R Jarvis
Journal:  Am J Infect Control       Date:  2000-06       Impact factor: 2.918

2.  Pseudomonas aeruginosa infection potential of oxygen humidifier devices.

Authors:  E W Ahlgren; J F Chapel; G L Dorn
Journal:  Respir Care       Date:  1977-04       Impact factor: 2.258

Review 3.  The role of environmental cleaning in the control of hospital-acquired infection.

Authors:  S J Dancer
Journal:  J Hosp Infect       Date:  2009-09-01       Impact factor: 3.926

4.  Phenotypic and Genotypic Methods for Detection of Extended Spectrum β Lactamase Producing Escherichia coli and Klebsiella pneumoniae Isolated from Ventilator Associated Pneumonia.

Authors:  Veena Krishnamurthy; Vijaykumar G S; Sudeepa Kumar M; Prashanth H V; Prakash R; Nagaraj E R
Journal:  J Clin Diagn Res       Date:  2013-09-10

5.  Bubbling humidifiers produce microaerosols which can carry bacteria.

Authors:  F S Rhame; A Streifel; C McComb; M Boyle
Journal:  Infect Control       Date:  1986-08

6.  Extended use of prefilled humidifier reservoirs and the likelihood of contamination.

Authors:  D Seigel; B Romo
Journal:  Respir Care       Date:  1990-08       Impact factor: 2.258

7.  Pseudomonas aeruginosa: growth in distilled water from hospitals.

Authors:  M S Favero; L A Carson; W W Bond; N J Petersen
Journal:  Science       Date:  1971-08-27       Impact factor: 47.728

8.  Nosocomial legionellosis associated with use of oxygen bubble humidifiers and underwater chest drains.

Authors:  A Moiraghi; M Castellani Pastoris; C Barral; F Carle; A Sciacovelli; G Passarino; P Marforio
Journal:  J Hosp Infect       Date:  1987-07       Impact factor: 3.926

9.  Multistate point-prevalence survey of health care-associated infections.

Authors:  Shelley S Magill; Jonathan R Edwards; Wendy Bamberg; Zintars G Beldavs; Ghinwa Dumyati; Marion A Kainer; Ruth Lynfield; Meghan Maloney; Laura McAllister-Hollod; Joelle Nadle; Susan M Ray; Deborah L Thompson; Lucy E Wilson; Scott K Fridkin
Journal:  N Engl J Med       Date:  2014-03-27       Impact factor: 91.245

10.  Nosocomial and ventilator-associated pneumonia in a community hospital intensive care unit: a retrospective review and analysis.

Authors:  Mehrdad Behnia; Sharon C Logan; Linda Fallen; Philip Catalano
Journal:  BMC Res Notes       Date:  2014-04-11
View more
  6 in total

1.  Radiological spectrum of invasive mucormycosis in COVID-19.

Authors:  Nandini Passi; Anshu C Wadhwa; Swati Naik
Journal:  BJR Case Rep       Date:  2022-03-09

2.  Secondary Cutaneous Mucormycosis Post-COVID-19: Case Series from a Tertiary Center.

Authors:  B Abhirami; Prabhat Kumar; Dharmendra Kumar Mishra; Satyendra Prasad Yadav
Journal:  Indian J Dermatol       Date:  2022 Mar-Apr       Impact factor: 1.757

3.  Cold Bubble Humidification of Oxygen: Old habits die hard.

Authors:  Sugata Dasgupta; Shrestha Ghosh; Atanu Chandra
Journal:  Sultan Qaboos Univ Med J       Date:  2022-08-25

4.  Cold bubble humidification of low-flow oxygen does not prevent acute changes in inflammation and oxidative stress at nasal mucosa.

Authors:  Lauriana Alves Santana; Suellen Karoline Moreira Bezerra; Beatriz Mangueira Saraiva-Romanholo; Wellington Pereira Yamaguti; Iolanda de Fátima Lopes Calvo Tibério; Tabata Maruyama Dos Santos; Renato Fraga Righetti
Journal:  Sci Rep       Date:  2021-07-12       Impact factor: 4.379

5.  The nasal oxygen practice in intensive care units in China: A multi-centered survey.

Authors:  Zunjia Wen; Junyu Chen; Lanzheng Bian; Ailing Xie; Mingqi Peng; Mei Li; Li Wei
Journal:  PLoS One       Date:  2018-08-30       Impact factor: 3.240

6.  Young people, young adults and binge drinking.

Authors:  V LA Fauci; R Squeri; P Spataro; C Genovese; N Laudani; V Alessi
Journal:  J Prev Med Hyg       Date:  2019-12-20
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.