Nosocomial legionellosis associated with use of oxygen bubble humidifiers and underwater chest drains.

A Moiraghi1, M Castellani Pastoris, C Barral, F Carle, A Sciacovelli, G Passarino, P Marforio.   

Abstract

In 1 year 12 of 48 patients who developed fatal pneumonia following admission with non-respiratory disorders to the Hospital Molinette, Torino, yielded Legionella pneumophila serogroup 1 from lung at autopsy. Patients were hospitalized on seven different wards for different conditions; only two of the wards had air conditioning but legionellas were not isolated from these. All patients were in poor health or immunocompromised. Some patients had inhaled humidified oxygen from piped supplies and three had undergone surgery. Legionella pneumophila serogroup 1 was detected in the water of oxygen bubble humidifiers and an underwater chest drain. The contaminated devices had been filled with tap or distilled water and the hospital water supply was found to be contaminated with L. pneumophila serogroup 1. Our findings suggest that filling bubble humidifiers or underwater chest drains with tap water is a potential hazard and should be avoided.

Entities:  

Mesh:

Year:  1987        PMID: 2888811     DOI: 10.1016/0195-6701(87)90031-4

Source DB:  PubMed          Journal:  J Hosp Infect        ISSN: 0195-6701            Impact factor:   3.926


  2 in total

1.  Legionella pneumophila in a hospital in Torino, Italy. A retrospective one-year study.

Authors:  A Moiraghi Ruggenini; M Castellani Pastoris; P J Dennis; C Barral; A Sciacovelli; F Carle; M Bolgiani; G Passarino; M G Mingrone; C Passi
Journal:  Epidemiol Infect       Date:  1989-02       Impact factor: 2.451

2.  Humidifiers for oxygen therapy: what risk for reusable and disposable devices?

Authors:  V La Fauci; G B Costa; A Facciolà; A Conti; R Riso; R Squeri
Journal:  J Prev Med Hyg       Date:  2017-06
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.