The output factor correction as function of the photon beam field size - direct measurement and calculation from the lateral dose response functions of gas-filled and solid detectors.

Daniela Poppinga1, Björn Delfs2, Jutta Meyners3, Dietrich Harder4, Björn Poppe2, Hui Khee Looe2.   

Abstract

The first aim of this study has been to extend the systematic experimental study of the field size dependence of the output factor correction for three micro-ionization chambers (PTW 31014, PTW 31022 and IBA Razor chamber), two silicon diodes (PTW 60017 and IBA Razor Diode) and the synthetic diamond detector microDiamond (PTW 60019) in a 6 MV photon beam down to an effective field side length of 2.6mm, and to summarize the present knowledge of this factor by treating it as a function of the dosimetric field size. In order to vary the dosimetric field size over this large range, output factors measurements were performed at source-to-surface distances of 60cm and 90cm. Since the output factors obtained with the organic scintillation detector Exradin W1 (Standard Imaging, Middleton, USA) at all field sizes closely agreed with those measured by EBT3 radiochromic films (ISP Corp, Wayne, USA), the scintillation detector served as the reference detector. The measured output correction factors reflect the influences of the volume averaging and density effects upon the uncorrected output factor values. In case of the microDiamond detector these opposing influences result in output factor correction values less than 1 for moderately small field sizes and larger than 1 for very small field sizes. Our results agree with most of the published experimental as well as Monte-Carlo simulated data within detector-specific limits of uncertainty. The dosimetric field side length has been identified as a reliable determinant of the output factor correction, and typical functional curve shapes of the field-size dependent output factor correction vs. dosimetric field side length have been associated with gas-filled, silicon diode and synthetic diamond detectors. The second aim of this study has been a novel, semi-empirical approach to calculate the field-size dependent output correction factors of small photon detectors by convolving film measured true dose profile data with measured lateral response functions of the detectors. To achieve this, the set of previously published 2D lateral dose response functions was complemented by those of the novel detectors PTW PinPoint chamber 31022 (PTW Freiburg, Freiburg, Germany), Razor chamber and Razor Diode (IBA Dosimetry, Schwarzenbruck, Germany). The output correction factors calculated from the lateral dose response functions closely fit with the directly measured output correction factors, thus supporting the latter by an independent method.
Copyright © 2018. Published by Elsevier GmbH.

Entities:  

Keywords:  Gas-filled and solid detectors; Lateral dose response function; Output factor correction; Photon-beam dosimetry

Mesh:

Year:  2017        PMID: 28869164     DOI: 10.1016/j.zemedi.2017.07.006

Source DB:  PubMed          Journal:  Z Med Phys        ISSN: 0939-3889            Impact factor:   4.820


  8 in total

1.  A novel method for the determination of field output factors and output correction factors for small static fields for six diodes and a microdiamond detector in megavoltage photon beams.

Authors:  Božidar Casar; Eduard Gershkevitsh; Ignasi Mendez; Slaven Jurković; M Saiful Huq
Journal:  Med Phys       Date:  2018-12-24       Impact factor: 4.071

2.  The role of radiation-induced charge imbalance on the dose-response of a commercial synthetic diamond detector in small field dosimetry.

Authors:  Hui Khee Looe; Daniela Poppinga; Rafael Kranzer; Isabel Büsing; Tuba Tekin; Ann-Britt Ulrichs; Björn Delfs; Dennis Vogt; Jan Würfel; Björn Poppe
Journal:  Med Phys       Date:  2019-05-02       Impact factor: 4.071

3.  Technical Note: Characterization of the new microSilicon diode detector.

Authors:  Ann-Britt Schönfeld; Daniela Poppinga; Rafael Kranzer; Rudy Leon De Wilde; Kay Willborn; Björn Poppe; Hui Khee Looe
Journal:  Med Phys       Date:  2019-07-31       Impact factor: 4.071

4.  Evaluation of the PTW microDiamond in edge-on orientation for dosimetry in small fields.

Authors:  Owen J Brace; Sultan F Alhujaili; Jason R Paino; Duncan J Butler; Dean Wilkinson; Brad M Oborn; Anatoly B Rosenfeld; Michael L F Lerch; Marco Petasecca; Jeremy A Davis
Journal:  J Appl Clin Med Phys       Date:  2020-05-22       Impact factor: 2.102

5.  Corrections of photon beam profiles of small fields measured with ionization chambers using a three-layer neural network.

Authors:  Ann-Britt Schönfeld; Karl Mund; Guanghua Yan; Andreas Alexander Schönfeld; Hui Khee Looe; Björn Poppe
Journal:  J Appl Clin Med Phys       Date:  2021-10-11       Impact factor: 2.102

6.  Output correction factors for small static fields in megavoltage photon beams for seven ionization chambers in two orientations - perpendicular and parallel.

Authors:  Božidar Casar; Eduard Gershkevitsh; Ignasi Mendez; Slaven Jurković; M Saiful Huq
Journal:  Med Phys       Date:  2019-11-25       Impact factor: 4.071

7.  Evaluation of the RUBY modular QA phantom for planar and non-coplanar VMAT and stereotactic radiations.

Authors:  Daniela Poppinga; Jana Kretschmer; Leonie Brodbek; Jutta Meyners; Bjoern Poppe; Hui Khee Looe
Journal:  J Appl Clin Med Phys       Date:  2020-08-14       Impact factor: 2.102

8.  Small field output correction factors of the microSilicon detector and a deeper understanding of their origin by quantifying perturbation factors.

Authors:  Carolin Weber; Rafael Kranzer; Jan Weidner; Kevin Kröninger; Björn Poppe; Hui Khee Looe; Daniela Poppinga
Journal:  Med Phys       Date:  2020-04-13       Impact factor: 4.071

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.