Teruko Fukuda1, Eri Asou2, Kimiko Nogi2, Kazuo Goto1. 1. Department of Clinical Laboratory Medicine, Teikyo University, 2-1-11 Kaga Itabashi, Tokyo 173-8605, Japan. 2. Teikyo University Hospital, 2-1-11 Kaga, Itabashi, Tokyo 173-0005, Japan.
Abstract
An evaluation of mouse red blood cell (RBC) and platelet (PLT) counting with an automated hematology analyzer was performed with three strains of mice, C57BL/6 (B6), BALB/c (BALB) and DBA/2 (D2). There were no significant differences in RBC and PLT counts between manual and automated optical methods in any of the samples, except for D2 mice. For D2, RBC counts obtained using the manual method were significantly lower than those obtained using the automated optical method (P<0.05), and PLT counts obtained using the manual method were higher than those obtained using the automated optical method (P<0.05). An automated hematology analyzer can be used for RBC and PLT counting; however, an appropriate method should be selected when D2 mice samples are used.
An evaluation of mouse red blood cell (RBC) and platelet (PLT) counting with an automated hematology analyzer was performed with three strains of mice, C57BL/6 (B6), BALB/c (BALB) and DBA/2 (D2). There were no significant differences in RBC and PLT counts between manual and automated optical methods in any of the samples, except for D2 mice. For D2, RBC counts obtained using the manual method were significantly lower than those obtained using the automated optical method (P<0.05), and PLT counts obtained using the manual method were higher than those obtained using the automated optical method (P<0.05). An automated hematology analyzer can be used for RBC and PLT counting; however, an appropriate method should be selected when D2 mice samples are used.
Entities:
Keywords:
automated hematology analyzer; mouse; platelet; red blood cell
Hematological analysis is an important test for the detection of disease and for monitoring
the health of humans and animals. Automated hematology analyzers have been routinely used as a
rapid, accurate and simple diagnostic method, not only for humans [5, 14], but also for animals [8, 15]. The three
major procedures performed in red blood cell (RBC) and platelet (PLT) counting are manual
phase contrast microscopy, impedance and optical light scatter/fluorescence [3]. Results produced from these three methods are not
necessarily identical due to differences in PLT particle size and aggregation, and red and/or
white blood cell fragmentation. Evaluations of the automated hematology analyzer using human
and animal samples have been reported [1, 12]. However, despite the fact that RBC and PLT in mice are
smaller in size and more numerous than in humans, evaluations of the automated hematology
analyzer using murine samples have not been reported, to the best of our knowledge.In this study, correlations between results from manual direct cell counting by microscopy
and an automated hematology analyzer, including optical and impedance methods, were evaluated
using commonly used three mouse strains.C57BL/6NCrl (B6), BALB/cAnNCrlCrlj (BALB) and DBA/2NCrl (D2) inbred mice were obtained from
Charles River Japan (Yokohama, Japan). Fifteen mice (five 3-week-old and ten 20-week-old male
mice) from each strain were included in the study. Approximately 500 µl blood
samples were collected from the axillary arteries and veins of mice under isoflurane
anesthesia and placed into sample tubes (Terumo Venoject II, TERUMO Corp., Tokyo, Japan)
containing ethylenediaminetetraacetic acid (EDTA-2K). Experiments were approved by the Animal
Care and Use Committee of the Teikyo University. The regulations were established in
compliance with the law for the humane treatment and management of animals.RBC and PLT were counted using an automated hematology analyzer (Sysmex XE-5000 hematology
analyzer, Sysmex, Kobe, Japan) in EDTA-anticoagulated whole blood samples. Measurements
included RBC and PLT counts by optical and impedance methods. Samples were thoroughly inverted
and then analyzed according to the manufacturer’s instructions.Whole blood samples in EDTA were diluted with isotonic solution (Gower’s solution), and then
the diluted blood was introduced into the counting chamber where the cell count was performed
microscopically (E200, Nikon, Tokyo, Japan).PLTs were counted in accordance with the brecher-chronkite method [2]. Whole blood samples were diluted with 1% ammonium oxalate solution. The
isotonic balance of the diluent was such that all erythrocytes are lysed while the platelets
remain intact. The dilution was mixed well and incubated at 25°C for 15 min to permit
erythrocyte lysis. Following the incubation period, the dilution was mounted on a
hematocytometer (ERMA Inc., Tokyo, Japan). The cells were allowed to settle and were then
counted in a specific area of the hemocytometer chamber under a microscope.Statistical analysis was performed using IBM SPSS® statistics software version 22
(Tokyo, Japan). Significant differences between the manual direct and automated counts were
determined using Student’s t-test, and the coefficient of variation (CV) of
the measurement results was calculated. P-values <0.05 were considered
statistically significant. Correlations between both test results were evaluated using the
Pearson’s correlation coefficient test.Mean± standard deviation (SD) RBC and PLT manual and automated counts for the different
methods, including optical and impedance methods, in BALB, B6 and D2mice (n=10, 20-week-old
mice each) are shown in Table 1A–C.
Table 1.
Mean (± standard deviation) manual and automated counts of RBC and PLT for the
different parameters, optical and impedance
For the manual counting method, mean± SD RBC counts in BALB mice (1,092.2 ± 95.4 ×
104 cells/µl) were significantly higher than those in C57BL/6
(989.6 ± 97.3 × 104 cells/µl) and D2 mice (1,003.2 ± 43.1 ×
104 cells/µl) (P<0.05). There was no
significant difference in RBC counts between B6 and D2mice. Mean ± SD PLT counts in BALB mice
(121.8 ± 13.3 × 104 cells/µl) were significantly higher than in B6
(104.4 ± 18.2 × 104 cells/µl) and D2 mice (107.8 ± 14.3 ×
104 cells/µl) (P<0.05). There was no
significant difference in PLT counts between B6 and D2mice.For RBC counts in BALB mice, the automated count using the impedance method generated
significantly lower values (998.2 ± 25.8 × 104 cells/µl) than the
manual method (1,092.2 ± 95.4 × 104 cells/µl)
(P<0.05). For RBC counts in D2 mice, automated counts using the optical
method (1,094.7 ± 44.1 × 104 cells/µl) were significantly higher
than those using the manual method (1,003.2 ± 43.1 × 104 cells/µl)
(P<0.05). There were no significant differences in counts between manual
and automated methods for the BALB and B6 strains of mice (P>0.05).For PLT in BALB B6 and D2mice, automated counts using the impedance method (BALB: 59.9 ± 6.1
× 104 cells/µl, B6: 55.9 ± 8.9 × 104
cells/µl and D2: 47.5 ± 9.7 × 104 cells/µl)
were significantly lower than those using the manual method (BALB: 121.8 ± 13.3 ×
104 cells/µl, B6: 104.4 ± 18.2 × 104
cells/µl and D2: 107.8 ± 14.3 × 104 cells/µl)
(P<0.05). Furthermore, PLT counts in D2 mice using the automated optical
method were also lower than those using the manual method (P=0.025). The CV
with the manual method was higher than that with the automated methods as shown in Table 1. Correlations between manual and automated
methods using 10-, 20-week-old and five 3-week-old mice of each strain are shown in Figs. 1 and 2.
Fig. 1.
Correlation between manual and automated impedance methods for RBC assessment in BALB/c
mice blood samples. Correlation coefficient: r=0.858, P<0.05.
Regression equation between manual (Y) and optical (X) methods: Y=1.52X+426.
Fig. 2.
Correlation between manual and automated optical methods for RBC assessment in DBA/2
mice blood samples. Correlation coefficient: r=0.841, P<0.05.
Regression equation between manual (Y) and optical (X) methods: Y=0.72X+203.
Correlation between manual and automated impedance methods for RBC assessment in BALB/c
mice blood samples. Correlation coefficient: r=0.858, P<0.05.
Regression equation between manual (Y) and optical (X) methods: Y=1.52X+426.Correlation between manual and automated optical methods for RBC assessment in DBA/2
mice blood samples. Correlation coefficient: r=0.841, P<0.05.
Regression equation between manual (Y) and optical (X) methods: Y=0.72X+203.For RBC counts, the regression equation between manual (Y) and impedance (X) in BALB mice was
Y=1.52X+426 (r=0.858, P<0.05), while the regression equation between
manual (Y) and optical (X) in D2 mice was Y=0.72X+203 (r=0.841,
P<0.05).PLT counts between manual and automated impedance methods in BALB and B6 mice, between manual
and automated optical methods in D2 mice, and between manual and automated impedance methods
in D2 mice are shown in Figs. 3, 4, 5, 6. Regression equations were Y=2.66X+39.76 (r=0.926, P<0.05),
Y=1.69X+7.45 (r=0.849, P<0.05), Y=1.12X+2.09 (r=0.95,
P<0.05) and Y=2.29X+4.81 (r=0.854, P<0.05),
respectively. Abnormal RBC morphology or PLT aggregation was not observed microscopically in
B6 and BALB mice. In contrast, PLT clotting was observed in all blood samples from D2 mice
including at 3 weeks and 20 weeks of age, as shown in Fig.
7.
Fig. 3.
Correlation between manual and automated impedance methods for PLT assessment in BALB/c
mice blood samples. Correlation coefficient: r=0.926, P<0.05.
Regression equation between manual (Y) and optical (X) methods: Y=2.66X+39.76.
Fig. 4.
Correlation between manual and automated impedance methods for PLT assessment in
C57BL/6 mice blood samples. Correlation coefficient: r=0.849,
P<0.05. Regression equation between manual (Y) and optical (X)
methods: Y=1.69X+7.45.
Fig. 5.
Correlation between manual and automated optical method for PLT assessment in DBA/2
mouse blood samples. Correlation coefficient: r=0.95, P<0.05.
Regression equation between manual (Y) and optical (X) methods: Y=1.12X+2.09.
Fig. 6.
Correlation between manual and automated impedance methods for PLT assessment in DBA/2
mouse blood samples. Correlation coefficient: r=0.854, P<0.05.
Regression equation between manual (Y) and optical (X) methods: Y=2.29X+4.81.
Fig. 7.
Platelet clotting in blood samples from DBA/2 mice aged 20 weeks (bar=100
µm).
Correlation between manual and automated impedance methods for PLT assessment in BALB/c
mice blood samples. Correlation coefficient: r=0.926, P<0.05.
Regression equation between manual (Y) and optical (X) methods: Y=2.66X+39.76.Correlation between manual and automated impedance methods for PLT assessment in
C57BL/6 mice blood samples. Correlation coefficient: r=0.849,
P<0.05. Regression equation between manual (Y) and optical (X)
methods: Y=1.69X+7.45.Correlation between manual and automated optical method for PLT assessment in DBA/2
mouse blood samples. Correlation coefficient: r=0.95, P<0.05.
Regression equation between manual (Y) and optical (X) methods: Y=1.12X+2.09.Correlation between manual and automated impedance methods for PLT assessment in DBA/2
mouse blood samples. Correlation coefficient: r=0.854, P<0.05.
Regression equation between manual (Y) and optical (X) methods: Y=2.29X+4.81.Platelet clotting in blood samples from DBA/2 mice aged 20 weeks (bar=100
µm).In this study, an evaluation of mouse red blood cell (RBC) and platelet (PLT) counting with
an automated hematology analyzer, including impedance and optical methods, was performed by
comparing data with that obtained by manual direct cell counting under a microscope. Fully
automated analyzers have been widely used to clarify hematological parameters in both humans
and animals. The automated hematology analyzer used in this study, the Sysmex XE-5000, has two
modes: the optical method and impedance method. Manual phase contrast microscopy count is
still currently recognized as the gold standard or reference method [3], because automated procedures are influenced by PLT clotting and cell
size [3, 12].
Impedance PLT counting can provide an accurate PLT count down to 20 × 103
cells/µl; however, a major disadvantage of this method is the difficulty in
distinguishing large PLT from extremely microcytic or fragmented RBC. In contrast, optical PLT
counting can distinguish between small and large PLT, RBC fragments and debris [3, 12]. In
particular, in the case of human blood samples, results from optical and impedance methods
have revealed differences in the presence of small-sized RBC, crushed erythrocytes, large PLT,
fragments of RBC and/or white blood cells, and aggregation of PLT and/or RBC [3]. It has been previously reported that the type of
anticoagulant affects PLT aggregation. Blood samples had significantly lower PLT counts and
more PLT aggregation in heparinized tubes than in EDTA tubes [13]; EDTA tubes were used in the present study. If blood specimens show an anomalous
particle-size distribution curve in the area where PLT counts are low (exceptionally low PLT
count samples), the counting method is automatically switched over to optical. In the present
study, even when the method was changed to optical, the PLT count of DBA2 mice was lower
compared with the manual method, suggesting that mouse strain should be taken into account
when PLT is measured using an automated hematology analyzer.Mouse blood samples, including cell size and volume, are relatively different compared with
human samples. The average size of RBC in humans and mice is 7.3 and 5.8 µm
in diameter [6], respectively, and the average RBC count
is 3.80–5.50 × 106 cells/µl and 10.2 × 106
cells/µl [7, 9], respectively. The average size of PLT in humans and mice is 1–2
µm and 0.5 µm [11],
respectively, and the average count is 150–400 × 103 cells/µl and
1,000–1,500 × 103 cells/µl, respectively [11]. Therefore, RBC and PLT sizes are larger in humans than in mice, while
counts are lower in humans than in mice.In this study, to evaluate the automated analyzer using samples from three strains of mice
(BALB, B6 and D2), the results of optical and impedance methods were compared with those of
the standard manual microscopic counting method. The results, including RBC and PLT counts,
suggested that the automated data from the impedance method were lower than those of the
manual microscopic method except with samples from B6 and D2mice. For B6 and D2mice samples,
significant differences between automated impedance and manual microscopic methods were not
observed. There were no significant differences between the automated data from the optical
and manual methods except for samples from D2 mice. The RBC counts of the automated optical
method were higher than those of manual method in D2 samples, while the PLT counts of the
automated optical method were lower than those of the manual method. PLT from D2 mice showed a
clumping tendency by microscopic analysis (data not shown) that may have caused a lower count
compared with the manual method; however, RBC condition, including form, did not differ
between mouse strains. Mean RBC count in D2 mice using the automated optical method was
1,094.7 ± 44.1 × 104 cells/µl, which was within the normal range,
suggesting the necessity for the definition of the normal range using automated methods with
D2 mice. D2 mice showed PLT clotting, the size of the clotting ranged between 50 and 100
µm (data not shown), suggesting that clotting PLT resulted in incorrect
counts with the impedance method. The analytical imprecision, CV, for PLT counting was higher
than that for RBC counting, which may have been caused by PLT clotting. It was reported that
PLT aggregation response in the whole blood was different depending on the animal species, for
example, the aggregation response against adenosine diphosphate; ADP was lower in baboons
comparing with in human [10]. The anticoagulants, such
as citrate, affect spontaneous aggregation of PLT in Wistar rat but not in Beagle dog [3]. These reports suggested that animal species should be
taken into account to count PLT using an automated hematology analyzer.Despite the fact that the data from manual and automated methods did not necessarily match, a
correlation between RBC and PLT results from both methods was recorded. This suggests that
automated data can be converted to manual data using a regression equation.It has been reported that blood counts including RBC and PLT vary depending on several
factors, such as anticoagulants, technique for obtaining a blood sample and sampling site
[7, 16]. All of
the samples used in the present study were collected in the same manner, using EDTA as a
coagulant; therefore, the differences between strains should be due to PLT function, such as
ease of clotting.In conclusion, an automated hematology analyzer, involving optical and impedance methods, can
be used as a rapid, accurate and simple diagnostic method. Sample conditions can influence the
accurate assessment of RBC and PLT counts. Our data showed that PLT counts using samples from
the D2 mouse strain were significantly lower compared with the manual method due to PLT
clotting. This suggests that the mouse strain should be taken into consideration when
performing hematological assessments.