Literature DB >> 28822017

Validation of a non-contact screening device for the combination of sleep-disordered breathing and periodic limb movements in sleep.

Gerhard Weinreich1, Sarah Terjung2,3, Yi Wang2, Stefanie Werther2, Alberto Zaffaroni4, Helmut Teschler2.   

Abstract

PURPOSE: Recent studies found that the non-contact screening device SleepMinder (ResMed Sensor Technologies, Dublin, Ireland) detects sleep-disordered breathing (SDB) with high diagnostic accuracy in cohorts suspected of this disorder. However, it was reported that in patients with periodic limb movement in sleep (PLMS), this non-contact device overestimates the apnea-hypopnea index (AHI). We aimed to overcome this limitation by introducing the novel sleep disorder index (SDI) which is sum of the AHI and the period limb movement index (PLMI).
METHODS: Between January 2011 and December 2013, we studied a mixed cohort of 57 patients (31 OSA, 19 PLMS). The easy-to-use non-contact device emits a very weak electromagnetic radiation and detects body movement by measuring the Doppler effect. We interpreted the device-generated movement index as the SDI and validated the diagnostic accuracy against simultaneous application of the gold-standard polysomnography (PSG).
RESULTS: We found that the SDI of the non-contact device correlated well with the sum of AHI and PLMI derived from PSG (r = 0.79, p = 0.01). For PSG-derived SDI cutoff ≥ 15/h, we obtained a sensitivity of 92.2% and a specificity of 95.8%. Positive likelihood ratio was 23.3 and negative likelihood ratio 0.03.
CONCLUSIONS: The studied non-contact screening device detects accurately the combination of the sleep disorders SDB and/or PLM. However, further testing is required in order to specify the nature of the underlying sleep disorder. At the current stage of algorithm development, the clinical strength is that the studied non-contact device can be used as a rule-out screening device for SDB and PLM.

Entities:  

Keywords:  Ambulatory monitoring; CSR; Cheyne-Stokes respiration; OSA; Obstructive sleep apnea; PLM; Periodic limb movement; Polysomnography

Mesh:

Year:  2017        PMID: 28822017     DOI: 10.1007/s11325-017-1546-x

Source DB:  PubMed          Journal:  Sleep Breath        ISSN: 1520-9512            Impact factor:   2.816


  21 in total

1.  Excess weight and sleep-disordered breathing.

Authors:  Terry Young; Paul E Peppard; Shahrad Taheri
Journal:  J Appl Physiol (1985)       Date:  2005-10

2.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

3.  Night-to-night variability in obstructive sleep apnea severity: relationship to overnight rostral fluid shift.

Authors:  Laura H White; Owen D Lyons; Azadeh Yadollahi; Clodagh M Ryan; T Douglas Bradley
Journal:  J Clin Sleep Med       Date:  2015-01-15       Impact factor: 4.062

4.  The association between sleep apnea and the risk of traffic accidents. Cooperative Group Burgos-Santander.

Authors:  J Terán-Santos; A Jiménez-Gómez; J Cordero-Guevara
Journal:  N Engl J Med       Date:  1999-03-18       Impact factor: 91.245

5.  Sleep/wake measurement using a non-contact biomotion sensor.

Authors:  Philip De Chazal; Niall Fox; Emer O'Hare; Conor Heneghan; Alberto Zaffaroni; Patricia Boyle; Stephanie Smith; Caroline O'Connell; Walter T McNicholas
Journal:  J Sleep Res       Date:  2010-08-12       Impact factor: 3.981

Review 6.  Periodic leg movements in sleep and periodic limb movement disorder: prevalence, clinical significance and treatment.

Authors:  Magdolna Hornyak; Bernd Feige; Dieter Riemann; Ulrich Voderholzer
Journal:  Sleep Med Rev       Date:  2006-06       Impact factor: 11.609

Review 7.  Cognitive deficits associated with sleep apnea syndrome: a proposed neuropsychological test battery.

Authors:  A Décary; I Rouleau; J Montplaisir
Journal:  Sleep       Date:  2000-05-01       Impact factor: 5.849

8.  The occurrence of sleep-disordered breathing among middle-aged adults.

Authors:  T Young; M Palta; J Dempsey; J Skatrud; S Weber; S Badr
Journal:  N Engl J Med       Date:  1993-04-29       Impact factor: 91.245

9.  Correlates of PLMs variability over multiple nights and impact upon RLS diagnosis.

Authors:  Lynn Marie Trotti; Donald L Bliwise; Sophia A Greer; Albert P Sigurdsson; Gudbjörg Birna Gudmundsdóttir; Thomas Wessel; Lisa M Organisak; Thor Sigthorsson; Kristleifur Kristjansson; Thordur Sigmundsson; David B Rye
Journal:  Sleep Med       Date:  2008-11-20       Impact factor: 3.492

10.  Rationale and design of the SERVE-HF study: treatment of sleep-disordered breathing with predominant central sleep apnoea with adaptive servo-ventilation in patients with chronic heart failure.

Authors:  Martin R Cowie; Holger Woehrle; Karl Wegscheider; Christiane Angermann; Marie-Pia d'Ortho; Erland Erdmann; Patrick Levy; Anita Simonds; Virend K Somers; Faiez Zannad; Helmut Teschler
Journal:  Eur J Heart Fail       Date:  2013-03-27       Impact factor: 15.534

View more
  3 in total

1.  Sleep Position Detection with a Wireless Audio-Motion Sensor-A Validation Study.

Authors:  Wojciech Kukwa; Tomasz Lis; Jonasz Łaba; Ron B Mitchell; Marcel Młyńczak
Journal:  Diagnostics (Basel)       Date:  2022-05-11

Review 2.  A Review of Scales to Evaluate Sleep Disturbances in Movement Disorders.

Authors:  Mónica M Kurtis; Roberta Balestrino; Carmen Rodriguez-Blazquez; Maria João Forjaz; Pablo Martinez-Martin
Journal:  Front Neurol       Date:  2018-05-29       Impact factor: 4.003

3.  Contactless detection of periodic leg movements during sleep: A 3D video pilot study.

Authors:  Stefan Seidel; Heinrich Garn; Markus Gall; Bernhard Kohn; Christoph Wiesmeyr; Markus Waser; Carmina Coronel; Andrijana Stefanic; Marion Böck; Markus Wimmer; Magdalena Mandl; Birgit Högl; Gerhard Klösch
Journal:  J Sleep Res       Date:  2020-02-04       Impact factor: 3.981

  3 in total