Technical Note: The effect of 2D excitation profile on T1 measurement accuracy using the variable flip angle method with an average flip angle assumption.

Bryant T Svedin1, Dennis L Parker1.   

Abstract

PURPOSE: To study the accuracy and precision of T1 estimates using the Variable Flip Angle (VFA) method in 2D and 3D acquisitions.
METHODS: Excitation profiles were simulated using numerical implementation of the Bloch equations for Hamming-windowed sinc excitation pulses with different time-bandwidth products (TBP) of 2, 6, and 10 and for T1 values of 295 ms and 1045 ms. Experimental data were collected in 5° increments from 5° to 90° for the same T1 and TBP values. T1 was calculated for every combination of flip angle with and without a correction for B1 and slice profile variation. Calculations were also made for flat slice profile such as obtained in 3D acquisition. Monte Carlo simulations were performed to obtain T1 measurement uncertainty.
RESULTS: VFA T1 measurements in 2D without correction can result in a 40-80% underestimation of true T1 . Flip angle correction can reduce the underestimation, but results in accurate measurements of T1 only within a narrow band of flip angle combinations. The narrow band of accuracy increases with TBP, but remains too narrow for any practical range of T1 values or B1 variation. Simulated noisy VFA T1 measurements in 3D were accurate as long as the two angles chosen are on either side of the Ernst angle.
CONCLUSIONS: Accurate T1 estimates from VFA 2D acquisitions are possible, but only a narrow range of T1 values within a narrow range of flip angle combinations can be accurately calculated using a 2D slice. Unless a better flip angle correction method is used, these results demonstrate that accurate measurements of T1 in 2D cannot be obtained robustly enough for practical use and are more likely obtained by a thin slab 3D VFA acquisition than from multiple-slice 2D acquisitions. VFA T1 measurements in 3D are accurate for wide ranges of flip angle combinations and T1 values.
© 2017 American Association of Physicists in Medicine.

Entities:  

Keywords:  2D excitation profile; T1 measurement; slice profile; variable flip angle

Mesh:

Year:  2017        PMID: 28796308      PMCID: PMC5689072          DOI: 10.1002/mp.12513

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  22 in total

1.  An analysis of the accuracy of magnetic resonance flip angle measurement methods.

Authors:  Glen R Morrell; Matthias C Schabel
Journal:  Phys Med Biol       Date:  2010-09-29       Impact factor: 3.609

2.  Rapid high-resolution T(1) mapping by variable flip angles: accurate and precise measurements in the presence of radiofrequency field inhomogeneity.

Authors:  Hai-Ling Margaret Cheng; Graham A Wright
Journal:  Magn Reson Med       Date:  2006-03       Impact factor: 4.668

3.  Investigating the effect of exchange and multicomponent T(1) relaxation on the short repetition time spoiled steady-state signal and the DESPOT1 T(1) quantification method.

Authors:  Sean C L Deoni; Brian K Rutt; Derek K Jones
Journal:  J Magn Reson Imaging       Date:  2007-03       Impact factor: 4.813

4.  Optimized and combined T1 and B1 mapping technique for fast and accurate T1 quantification in contrast-enhanced abdominal MRI.

Authors:  Reto Treier; Andreas Steingoetter; Michael Fried; Werner Schwizer; Peter Boesiger
Journal:  Magn Reson Med       Date:  2007-03       Impact factor: 4.668

Review 5.  MR thermometry.

Authors:  Viola Rieke; Kim Butts Pauly
Journal:  J Magn Reson Imaging       Date:  2008-02       Impact factor: 4.813

6.  Accurate determination of spin-density and T1 in the presence of RF-field inhomogeneities and flip-angle miscalibration.

Authors:  R Venkatesan; W Lin; E M Haacke
Journal:  Magn Reson Med       Date:  1998-10       Impact factor: 4.668

7.  MR imaging: clinical use of the inversion recovery sequence.

Authors:  G M Bydder; I R Young
Journal:  J Comput Assist Tomogr       Date:  1985 Jul-Aug       Impact factor: 1.826

8.  Hybrid proton resonance frequency/T1 technique for simultaneous temperature monitoring in adipose and aqueous tissues.

Authors:  Nick Todd; Mahamadou Diakite; Allison Payne; Dennis L Parker
Journal:  Magn Reson Med       Date:  2012-03-05       Impact factor: 4.668

9.  Calculated T1 images derived from a partial saturation-inversion recovery pulse sequence with adiabatic fast passage.

Authors:  C J Hardy; W A Edelstein; D Vatis; R Harms; W J Adams
Journal:  Magn Reson Imaging       Date:  1985       Impact factor: 2.546

10.  High-resolution T1 and T2 mapping of the brain in a clinically acceptable time with DESPOT1 and DESPOT2.

Authors:  Sean C L Deoni; Terry M Peters; Brian K Rutt
Journal:  Magn Reson Med       Date:  2005-01       Impact factor: 4.668

View more
  2 in total

1.  Simultaneous proton resonance frequency shift thermometry and T1 measurements using a single reference variable flip angle T1 method.

Authors:  Bryant T Svedin; Allison Payne; Dennis L Parker
Journal:  Magn Reson Med       Date:  2019-01-16       Impact factor: 4.668

2.  Rapid 2D variable flip angle method for accurate and precise T1 measurements over a wide range of T1  values.

Authors:  Beatrice Lena; Clemens Bos; Cyril J Ferrer; Chrit T W Moonen; Max A Viergever; Lambertus W Bartels
Journal:  NMR Biomed       Date:  2021-05-24       Impact factor: 4.044

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.